ToF Performance Evaluation of PET Modules With Digital Silicon Photomultiplier Technology During MR Operation

In 2012, we presented the Hyperion-II D preclinical PET insert which uses Philips Digital Photon Counting's digital SiPMs and is designed to be operated in a 3-T MRI. In this work we use the same platform equipped with scintillators having dimensions closer to a clinical application. This allows an investigation of the time of flight (ToF) performance of the platform and its behavior during simultaneous MR operation. We employ LYSO crystal arrays of 4×4 ×10 mm3 coupled to 4 ×4 PDPC DPC 3200-22 sensors (DPC) resulting in a one-to-one coupling of crystals to read-out channels. Six sensor stacks are mounted onto a singles processing unit in a 2 ×3 arrangement. Two modules are mounted horizontally facing each other on a gantry with a crystal-to-crystal spacing of 217.6 mm (gantry position). A second arrangement places the modules at the maximum distance of approximately 410 mm inside the MR bore (maximum distance position) which brings each module close to the gradient system. The DPCs are cooled down to approximately 5-10° C under operation. We disable 20% of the worst cells and use an overvoltage of Vov = 2.0 V and 2.5 V. To obtain the best time stamps, we use the trigger scheme 1 (first photon trigger), a narrow energy window of 511 ±50 keV and a minimum required light fraction of the main pixel of more than 65% to reject intercrystal scatter. By using a 22Na point source in the isocenter of the modules, the coincidence resolution time (CRT) of the two modules is evaluated inside the MRI system without MR activity and while using highly demanding gradient sequences. Inside the B0 field without any MR activity at an overvoltage of Vov = 2.0 V, the energy resolution is 11.45% (FWHM) and the CRT is 250 ps (FWHM). At an overvoltage of Vov = 2.5 V, the energy resolution is 11.15% (FWHM) and the CRT is 240 ps (FWHM). During a heavy z-gradient sequence (EPI factor: 49, gradient strength: 30 mT/m, slew rate: 192.3 mT/m/ms, TE/TR: 12/25 ms and switching duty cycle: 67%) at the gantry position and an overvoltage of Vov = 2.0 V, the energy resolution is degraded relatively by 4.1% and the CRT by 25%. Using the same sequence but at the maximum distance position and an overvoltage of Vov = 2.5 V, we measure a degradation of the energy resolution of 9.2% and a 52% degradation of the CRT. The Hyperion-IID platform proofs to deliver good timing performance and energy resolution inside the MRI system even under highly demanding gradient sequences.

[1]  H. Jadvar,et al.  Competitive advantage of PET/MRI. , 2014, European journal of radiology.

[2]  V. Schulz,et al.  PET/MRI insert using digital SiPMs: Investigation of MR-compatibility , 2014, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment.

[3]  Volkmar Schulz,et al.  Design concept of world's first preclinical PET/MR insert with fully digital silicon photomultiplier technology , 2012, 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC).

[4]  S. Cherry,et al.  Simultaneous acquisition of multislice PET and MR images: initial results with a MR-compatible PET scanner. , 2006, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[5]  S. Cherry,et al.  Simultaneous in vivo positron emission tomography and magnetic resonance imaging , 2008, Proceedings of the National Academy of Sciences of the United States of America.

[6]  H. Zaidi,et al.  An outlook on future design of hybrid PET/MRI systems. , 2011, Medical physics.

[7]  V. Schulz,et al.  First evaluations of the neighbor logic of the digital SiPM tile , 2012, 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC).

[8]  S. Cherry,et al.  Simultaneous PET and MR imaging , 1997, Physics in medicine and biology.

[9]  Yiping Shao,et al.  PET and NMR dual acquisition (PANDA): applications to isolated, perfused rat hearts , 1997, NMR in biomedicine.

[10]  Ciprian Catana,et al.  Performance test of an LSO-APD detector in a 7-T MRI scanner for simultaneous PET/MRI. , 2006, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[11]  H. Schlemmer,et al.  A look ahead: PET/MR versus PET/CT , 2009, European Journal of Nuclear Medicine and Molecular Imaging.

[12]  Gerald Antoch,et al.  Oncologic PET/MRI, Part 2: Bone Tumors, Soft-Tissue Tumors, Melanoma, and Lymphoma , 2012, The Journal of Nuclear Medicine.

[13]  Carsten Degenhardt,et al.  The digital silicon photomultiplier — System architecture and performance evaluation , 2010, IEEE Nuclear Science Symposuim & Medical Imaging Conference.

[14]  Giacomo Borghi,et al.  Probabilities of triggering and validation in a digital silicon photomultiplier , 2014 .

[15]  Volkmar Schulz,et al.  Gradient transparent RF housing for simultaneous PET/MRI using carbon fiber composites , 2012, 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC).

[16]  Martin S Judenhofer,et al.  Applications for preclinical PET/MRI. , 2013, Seminars in nuclear medicine.

[17]  D. Townsend Combined positron emission tomography-computed tomography: the historical perspective. , 2008, Seminars in ultrasound, CT, and MR.

[18]  Nadim Joni Shah,et al.  Analysis and Correction of Count Rate Reduction During Simultaneous MR-PET Measurements With the BrainPET Scanner , 2012, IEEE Transactions on Medical Imaging.

[19]  A. Drzezga,et al.  First Clinical Experience with Integrated Whole-Body PET/MR: Comparison to PET/CT in Patients with Oncologic Diagnoses , 2012, The Journal of Nuclear Medicine.

[20]  Ciprian Catana,et al.  Simultaneous PET-MRI: a new approach for functional and morphological imaging , 2008, Nature Medicine.

[21]  Bernd J Pichler,et al.  Latest Advances in Molecular Imaging Instrumentation , 2008, Journal of Nuclear Medicine.

[22]  Volkmar Schulz,et al.  An MR-compatible singles detection and processing unit for simultaneous preclinical PET/MR , 2012, 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC).

[23]  Volkmar Schulz,et al.  A preclinical PET/MR insert for a human 3T MR scanner , 2009, 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC).

[24]  Eiji Sugiyama,et al.  Simultaneous imaging using Si-PM-based PET and MRI for development of an integrated PET/MRI system , 2012, Physics in medicine and biology.

[25]  C. W. Lerche,et al.  Towards Software-Based Real-Time Singles and Coincidence Processing of Digital PET Detector Raw Data , 2013, IEEE Transactions on Nuclear Science.

[26]  G. Antoch,et al.  Oncologic PET/MRI, Part 1: Tumors of the Brain, Head and Neck, Chest, Abdomen, and Pelvis , 2012, The Journal of Nuclear Medicine.

[27]  Volkmar Schulz,et al.  MR compatibility aspects of a silicon photomultiplier-based PET/RF insert with integrated digitisation , 2014, Physics in medicine and biology.

[28]  Simon R. Cherry,et al.  Development of a PET detector system compatible with MRI/NMR systems , 1997 .

[29]  R. Dorscheid,et al.  Performance evaluation of a prototype Positron Emission Tomography scanner using Digital Photon Counters (DPC) , 2012, 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC).

[30]  Volkmar Schulz,et al.  Development of an MRI-compatible digital SiPM detector stack for simultaneous PET/MRI , 2016, Biomedical physics & engineering express.

[31]  Jae Sung Lee,et al.  Initial Results of Simultaneous PET/MRI Experiments with an MRI-Compatible Silicon Photomultiplier PET Scanner , 2012, The Journal of Nuclear Medicine.

[32]  Bernd J Pichler,et al.  Principles of PET/MR Imaging , 2014, The Journal of Nuclear Medicine.

[33]  Volkmar Schulz,et al.  FPGA-based singles and coincidences processing pipeline for integrated digital PET/MR detectors , 2012, 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC).

[34]  In Chan Song,et al.  An Investigation Into the Use of Geiger-Mode Solid-State Photomultipliers for Simultaneous PET and MRI Acquisition , 2008, IEEE Transactions on Nuclear Science.

[35]  Yong Choi,et al.  A feasibility study of photosensor charge signal transmission to preamplifier using long cable for development of hybrid PET-MRI. , 2010, Medical physics.

[36]  T. Frach,et al.  The digital silicon photomultiplier — Principle of operation and intrinsic detector performance , 2009, 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC).

[37]  Yong Choi,et al.  A small animal PET based on GAPDs and charge signal transmission approach for hybrid PET-MR imaging , 2011 .