Determination of Stent Frame Displacement After Endovascular Aneurysm Sealing

Purpose: To describe and validate a new methodology for visualizing and quantifying 3-dimensional (3D) displacement of the stent frames of the Nellix endosystem after endovascular aneurysm sealing (EVAS). Methods: The 3D positions of the stent frames were registered to 5 fixed anatomical landmarks on the post-EVAS computed tomography (CT) scans, facilitating comparison of the position and shape of the stent frames between consecutive follow-up scans. Displacement of the proximal and distal ends of the stent frames, the entire stent frame trajectories, as well as changes in distance between the stent frames were determined for 6 patients with >5-mm displacement and 6 patients with <5-mm displacement at 1-year follow-up. The measurements were performed by 2 independent observers; the intraclass correlation coefficient (ICC) was used to determine interobserver variability. Results: Three types of displacement were identified: displacement of the proximal and/or distal end of the stent frames, lateral displacement of one or both stent frames, and stent frame buckling. The ICC ranged from good (0.750) to excellent (0.958). No endoleak or migration was detected in the 12 patients on conventional CT angiography at 1 year. However, of the 6 patients with >5-mm displacement on the 1-year CT as determined by the new methodology, 2 went on to develop a type Ia endoleak in longer follow-up, and displacement progressed to >15 mm for 2 other patients. No endoleak or progressive displacement was appreciated for the patients with <5-mm displacement. Conclusion: The sac anchoring principle of the Nellix endosystem may result in several types of displacement that have not been observed during surveillance of regular endovascular aneurysm repairs. The presented methodology allows precise 3D determination of the Nellix endosystems and can detect subtle displacement better than standard CT angiography. Displacement >5 mm on the 1-year CT scans reconstructed with the new methodology may forecast impaired sealing and anchoring of the Nellix endosystem.

[1]  F. Torella,et al.  Clinical Outcome after Endovascular Sealing of Abdominal Aortic Aneurysms: A Retrospective Cohort Study. , 2017, Annals of vascular surgery.

[2]  C. Vahl,et al.  Three-Dimensional Analysis of Component Stability of the Nellix Endovascular Aneurysm Sealing System After Treatment of Infrarenal Abdominal Aortic Aneurysms , 2017, Journal of Endovascular Therapy.

[3]  D. Böckler,et al.  One‐year pivotal trial outcomes of the Nellix system for endovascular aneurysm sealing , 2017, Journal of vascular surgery.

[4]  M. Ferri,et al.  Midterm Outcomes of the Nellix Endovascular Aneurysm Sealing System , 2016, Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists.

[5]  D. Böckler,et al.  Imaging After Nellix Endovascular Aneurysm Sealing , 2016, Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists.

[6]  D. Böckler,et al.  Multicenter Nellix EndoVascular Aneurysm Sealing system experience in aneurysm sac sealing. , 2015, Journal of vascular surgery.

[7]  A. Karthikesalingam,et al.  Endovascular aneurysm sealing for infrarenal abdominal aortic aneurysms: 30-day outcomes of 105 patients in a single centre. , 2015, European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery.

[8]  Rodney A. White,et al.  Initial clinical experience with a sac-anchoring endoprosthesis for aortic aneurysm repair. , 2011, Journal of vascular surgery.

[9]  C. Zarins Stent-Graft Migration: How Do We Know When We Have it and What is its Significance? , 2004, Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists.

[10]  S. Haulon,et al.  Stent-Graft Migration: A Reappraisal of Analysis Methods and Proposed Revised Definition , 2004, Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists.

[11]  K. Ouriel,et al.  Aortoiliac morphologic correlations in aneurysms undergoing endovascular repair. , 2003, Journal of vascular surgery.

[12]  J. Matsumura,et al.  A multicenter controlled clinical trial of open versus endovascular treatment of abdominal aortic aneurysm. , 2003, Journal of vascular surgery.

[13]  C. Zarins,et al.  Reporting standards for endovascular aortic aneurysm repair. , 2002, Journal of vascular surgery.

[14]  D. Böckler,et al.  Results of the Nellix system investigational device exemption pivotal trial for endovascular aneurysm sealing. , 2016, Journal of vascular surgery.