A general method for cupping artifact correction of cone-beam breast computed tomography images

PurposeCone-beam breast computed tomography (CBBCT), a promising breast cancer diagnostic technique, has been under investigation for the past decade. However, owing to scattered radiation and beam hardening, CT numbers are not uniform on CBBCT images. This is known as cupping artifact, and it presents an obstacle for threshold-based volume segmentation. In this study, we proposed a general post-reconstruction method for cupping artifact correction.MethodsThere were four steps in the proposed method. First, three types of local region histogram peaks were calculated: adipose peaks with low CT numbers, glandular peaks with high CT numbers, and unidentified peaks. Second, a linear discriminant analysis classifier, which was trained by identified adipose and glandular peaks, was employed to identify the unidentified peaks as adipose or glandular peaks. Third, adipose background signal profile was fitted according to the adipose peaks using the least squares method. Finally, the adipose background signal profile was subtracted from original image to obtain cupping corrected imageResultsIn experimental study, standard deviation of adipose tissue CT numbers was obviously reduced and the CT numbers were more uniform after cupping correction by proposed method; in simulation study, root-mean-square errors were significantly reduced for both symmetric and asymmetric cupping artifacts, indicating that the proposed method was effective to both artifacts.ConclusionsA general method without a circularly symmetric assumption was proposed to correct cupping artifacts in CBBCT images for breast. It may be properly applied to images of real patient breasts with natural pendent geometry.

[1]  Biao Chen,et al.  Cone-beam volume CT breast imaging: feasibility study. , 2002, Medical physics.

[2]  Ruola Ning,et al.  Scatter correction for clinical cone beam CT breast imaging based on breast phantom studies. , 2011, Journal of X-ray science and technology.

[3]  J. Boone,et al.  Dedicated breast CT: radiation dose and image quality evaluation. , 2001, Radiology.

[4]  José V. Manjón,et al.  A nonparametric MRI inhomogeneity correction method , 2007, Medical Image Anal..

[5]  I. Sechopoulos,et al.  Cupping artifact correction and automated classification for high-resolution dedicated breast CT images. , 2012, Medical physics.

[6]  A. Rim,et al.  Trends in breast cancer screening and diagnosis. , 2008, Cleveland Clinic journal of medicine.

[7]  X Liu,et al.  A post-reconstruction method to correct cupping artifacts in cone beam breast computed tomography. , 2007, Medical physics.

[8]  Jean-Marc Dinten,et al.  Evaluation of a physical based approach of scattered radiation correction in cone beam CT with an anthropomorphic thorax phantom , 2006, SPIE Medical Imaging.

[9]  J Michael O'Connor,et al.  Characterization of scatter in cone-beam CT breast imaging: comparison of experimental measurements and Monte Carlo simulation. , 2009, Medical physics.

[10]  Lingyun Chen,et al.  Feasibility of volume-of-interest (VOI) scanning technique in cone beam breast CT--a preliminary study. , 2008, Medical physics.

[11]  N. Otsu A threshold selection method from gray level histograms , 1979 .

[12]  Aruna A. Vedula,et al.  Microcalcification detection using cone-beam CT mammography with a flat-panel imager. , 2004, Physics in medicine and biology.

[13]  Tianpeng Wang,et al.  High resolution dual detector volume-of-interest cone beam breast CT--Demonstration with a bench top system. , 2011, Medical physics.

[14]  Frank Verhaegen,et al.  Scatter correction for kilovoltage cone-beam computed tomography (CBCT) images using Monte Carlo simulations , 2006, SPIE Medical Imaging.

[15]  Tao Han Image segmentation, modeling, and simulation in 3D breast x-ray imaging , 2010 .

[16]  W. P. Evans,et al.  Breast cancer screening: Successes and challenges , 2012, CA: a cancer journal for clinicians.

[17]  J Michael O'Connor,et al.  Generation of voxelized breast phantoms from surgical mastectomy specimens. , 2013, Medical physics.

[18]  M E Read,et al.  Breast skin thickness: normal range and causes of thickening shown on film-screen mammography. , 1984, Journal of the Canadian Association of Radiologists.

[19]  Lingyun Chen,et al.  Feasibility of volume-of-interest (VOI) scanning technique in cone beam breast CT-a preliminary study. , 2008, Medical physics.

[20]  L. Feldkamp,et al.  Practical cone-beam algorithm , 1984 .

[21]  D. Jaffray,et al.  Cone-beam computed tomography with a flat-panel imager: magnitude and effects of x-ray scatter. , 2001, Medical physics.

[22]  Timothy D. Solberg,et al.  Scatter correction for cone-beam computed tomography using moving blocker strips: a preliminary study. , 2010 .

[23]  J. Affeldt,et al.  The feasibility study , 2019, The Information System Consultant’s Handbook.

[24]  D. Jaffray,et al.  The influence of antiscatter grids on soft-tissue detectability in cone-beam computed tomography with flat-panel detectors. , 2004, Medical physics.

[25]  John M. Boone,et al.  Simulation of Mechanical Compression of Breast Tissue , 2007, IEEE Transactions on Biomedical Engineering.

[26]  J. Boone,et al.  Evaluation of x-ray scatter properties in a dedicated cone-beam breast CT scanner. , 2005, Medical physics.

[27]  J. Boone,et al.  Evaluation of x-ray scatter properties in a dedicated cone-beam breast CT scanner. , 2005, Medical physics.

[28]  S. Richard,et al.  A simple, direct method for x-ray scatter estimation and correction in digital radiography and cone-beam CT. , 2005, Medical physics.

[29]  Kai Yang,et al.  A breast-specific, negligible-dose scatter correction technique for dedicated cone-beam breast CT: a physics-based approach to improve Hounsfield Unit accuracy , 2014, Physics in medicine and biology.

[30]  Selin Carkaci,et al.  Dedicated cone-beam breast CT: feasibility study with surgical mastectomy specimens. , 2007, AJR. American journal of roentgenology.

[31]  Avinash C. Kak,et al.  PCA versus LDA , 2001, IEEE Trans. Pattern Anal. Mach. Intell..

[32]  Tianpeng Wang,et al.  Radiation doses in cone-beam breast computed tomography: a Monte Carlo simulation study. , 2011, Medical physics.