Design and validation of a novel Cartesian biomechanical testing system with coordinated 6DOF real-time load control: application to the lumbar spine (L1-S, L4-L5).

Robotic methods applied to in-vitro biomechanical testing potentially offer more comprehensive evaluations however, standard position control algorithms make real-time load control problematic. This paper describes and evaluates a novel custom developed Cartesian force controlled biomechanical testing system with coordinated 6 degree of freedom (DOF) real-time load control. A custom developed 6-DOF serial manipulator with cascaded force over position control algorithms was designed, assembled, and programmed. Dial gauge tests assessed accuracy of custom linear axes. Standard test input and tuning procedures refined control performance. Two single motion segment units (L4-L5) and lumbar (L1-S) spine segments were tested under continuous pure moment application in flexion-extension, left-right lateral bending and axial rotation to 8Nm under full 6-DOF load control. Mean load control tracking errors between commanded and experimental loads were computed. Global spinal ranges of motion were compared to previously published values for standard non-robotic protocols. Individual linear and rotational axis position control accuracies were equal to or less than 6.35μm and 0.0167° respectively. Pilot pure bending tests demonstrated stable load control performance, as well as load rates, rotational velocities, and ranges of motion comparable to those for standard non-robotic in-vitro tests. Tracking errors for zero commanded forces and all moment controlled axes were less than 0.81±0.68N and 0.18±0.19Nm over all tests, respectively. The Cartesian based system simplified control application and demonstrated robust position and load control that was not limited to single axis or zero commanded loads. In addition to emulating standard biomechanical tests, the novel Cartesian force controlled testing system developed is a promising tool for biomechanical assessments with coordinated dynamic load application and coupled motion response in 6DOF.

[1]  S Arai,et al.  The use of robotics technology to study human joint kinematics: a new methodology. , 1993, Journal of biomechanical engineering.

[2]  D. M. Gorinevskiĭ,et al.  Force control of robotics systems , 1997 .

[3]  H. Wilke,et al.  Biomechanical effect of different lumbar interspinous implants on flexibility and intradiscal pressure , 2008, European Spine Journal.

[4]  Uvo Hölscher,et al.  Evaluation of a robot-assisted testing system for multisegmental spine specimens. , 2012, Journal of biomechanics.

[5]  Kathleen Denis,et al.  Machining and Accuracy Studies for a Tibial Knee Implant Using a Force-Controlled Robot , 1998 .

[6]  Glen A. Livesay,et al.  Improvement of Accuracy in a High-Capacity, Six Degree-of-freedom Load Cell: Application to Robotic Testing of Musculoskeletal Joints , 2004, Annals of Biomedical Engineering.

[7]  G A Livesay,et al.  A combined robotic/universal force sensor approach to determine in situ forces of knee ligaments. , 1996, Journal of biomechanics.

[8]  M. Panjabi,et al.  Multidirectional Testing of One- and Two-Level ProDisc-L Versus Simulated Fusions , 2007, Spine.

[9]  L. Claes,et al.  A universal spine tester for in vitro experiments with muscle force simulation , 2005, European Spine Journal.

[10]  M M Panjabi,et al.  Mechanical behavior of the human lumbar and lumbosacral spine as shown by three-dimensional load-displacement curves. , 1994, The Journal of bone and joint surgery. American volume.

[11]  Rosemary E Thompson,et al.  Defining the Neutral Zone of sheep intervertebral joints during dynamic motions: an in vitro study. , 2003, Clinical biomechanics.

[12]  David L Butler,et al.  Investigating the effects of anterior tibial translation on anterior knee force in the porcine model: Is the porcine knee ACL dependent? , 2011, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[13]  Li-Qun Zhang,et al.  Developing a 6-DOF robot to investigate multi-axis ACL injuries under valgus loading coupled with tibia internal rotation , 2010, 2010 Annual International Conference of the IEEE Engineering in Medicine and Biology.

[14]  Freddie H. Fu,et al.  Varying Femoral Tunnels between the Anatomical Footprint and Isometric Positions , 2005, The American journal of sports medicine.

[15]  Harry E Rubash,et al.  Kinematics of the knee at high flexion angles: An in vitro investigation , 2004, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[16]  Takeshi Sekito,et al.  A novel robotic system for joint biomechanical tests: application to the human knee joint. , 2004, Journal of biomechanical engineering.

[17]  T. Oxland,et al.  Neutral zone and range of motion in the spine are greater with stepwise loading than with a continuous loading protocol. An in vitro porcine investigation. , 2004, Journal of biomechanics.

[18]  Lars G. Gilbertson,et al.  New methods to study lumbar spine biomechanics: Delineation of in vitro load-displacement characteristics by using a robotic/UFS testing system with hybrid control , 2000 .

[19]  Avinash G Patwardhan,et al.  Effect of compressive follower preload on the flexion–extension response of the human lumbar spine , 2003, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[20]  I. Stokes,et al.  Structural behavior of human lumbar spinal motion segments. , 2004, Journal of biomechanics.

[21]  Darrell J Goertzen,et al.  A novel application of velocity-based force control for use in robotic biomechanical testing. , 2009, Journal of biomechanics.

[22]  James P. Dickey,et al.  New methodology for multi-dimensional spinal joint testing with a parallel robot , 2006, Medical & Biological Engineering & Computing.

[23]  Janet L Ronsky,et al.  Reproduction of in vivo motion using a parallel robot. , 2007, Journal of biomechanical engineering.

[24]  Dimitry Gorinevsky,et al.  Force Control of Robotics Systems , 1998 .

[25]  James P Dickey,et al.  Representation of passive spinal element contributions to in vitro flexion-extension using a polynomial model: illustration using the porcine lumbar spine. , 2003, Journal of biomechanics.

[26]  Herbert F. Voigt,et al.  IEEE Engineering in Medicine and Biology Society , 2019, IEEE Transactions on Biomedical Engineering.

[27]  S L Woo,et al.  Use of robotic technology for diathrodial joint research. , 1999, Journal of science and medicine in sport.

[28]  Denis J. DiAngelo,et al.  A Multiaxis Programmable Robot for the Study of Multibody Spine Biomechanics Using a Real-Time Trajectory Path Modification Force and Displacement Control Strategy , 2013 .