A constant-flow calorimeter for the measurement of acoustic power at megahertz frequencies.
暂无分享,去创建一个
The calorimeter absorbs, in a flow of castor oil, a fraction of the acoustic power output of an ultrasound transducer. It has been used to measure powere of between 1 mW and 10 W at frequencies from 1-5 to 3 MHz. Examples are given of results obtained for an ultrasound therapy unit and an ultrasonic pulse--echo scanner. Sources of inaccuracy are defined. It appears that constant-flow calorimetry offers a potentially very sensitive method for the absolute measurement of acoustic power.