Projection Reconstruction Magnetic Particle Imaging

We acquire the first experimental 3-D tomographic images with magnetic particle imaging (MPI) using projection reconstruction methodology, which is similar to algorithms employed in X-ray computed tomography. The primary advantage of projection reconstruction methods is an order of magnitude increase in signal-to-noise ratio (SNR) due to averaging. We first derive the point spread function, resolution, number of projections required, and the SNR gain in projection reconstruction MPI. We then design and construct the first scanner capable of gathering the necessary data for nonaliased projection reconstruction and experimentally verify our mathematical predictions. We demonstrate that filtered backprojection in MPI is experimentally feasible and illustrate the SNR and resolution improvements with projection reconstruction. Finally, we show that MPI is capable of producing three dimensional imaging volumes in both phantoms and postmortem mice.

[1]  B. Logan,et al.  Reconstructing Interior Head Tissue from X-Ray Transmissions , 1974 .

[2]  Patrick W. Goodwill,et al.  The impact of filtering direct-feedthrough on the x-space theory of magnetic particle imaging , 2011, Medical Imaging.

[3]  Thorsten M. Buzug,et al.  Generation of a static magnetic field-free line using two Maxwell coil pairs , 2010 .

[4]  Thorsten M. Buzug,et al.  Field-free line formation in a magnetic field , 2010 .

[5]  Q. Pankhurst,et al.  Progress in applications of magnetic nanoparticles in biomedicine , 2009 .

[6]  P. Turski,et al.  Radiation dose reduction in time-resolved CT angiography using highly constrained back projection reconstruction. , 2009, Physics in medicine and biology.

[7]  D. Holdsworth,et al.  Micro-CT in small animal and specimen imaging , 2002 .

[8]  Lutz Trahms,et al.  Determination of the binding reaction between avidin and biotin by relaxation measurements of magnetic nanoparticles , 1999 .

[9]  R. Bracewell Strip Integration in Radio Astronomy , 1956 .

[10]  D. Donoho,et al.  Sparse MRI: The application of compressed sensing for rapid MR imaging , 2007, Magnetic resonance in medicine.

[11]  William V. Houston,et al.  The Fine Structure and the Wave-Length of the Balmer Lines , 1926 .

[12]  Thorsten M Buzug,et al.  Experimental generation of an arbitrarily rotated field-free line for the use in magnetic particle imaging. , 2011, Medical physics.

[13]  Jie Tang,et al.  Prior image constrained compressed sensing (PICCS): a method to accurately reconstruct dynamic CT images from highly undersampled projection data sets. , 2008, Medical physics.

[14]  Patrick W. Goodwill,et al.  Experimental demonstration of x-space magnetic particle imaging , 2011, Medical Imaging.

[15]  J. Connolly,et al.  Proposed biosensors based on time-dependent properties of magnetic fluids , 2001 .

[16]  Jeffrey A. Fessler,et al.  Fourier-based forward and back-projectors in iterative fan-beam tomographic image reconstruction , 2006, 2004 2nd IEEE International Symposium on Biomedical Imaging: Nano to Macro (IEEE Cat No. 04EX821).

[17]  Rafael C. González,et al.  Digital image processing, 3rd Edition , 2008 .

[18]  Thorsten M Buzug,et al.  Efficient generation of a magnetic field-free line. , 2010, Medical physics.

[19]  G. N. Ramachandran,et al.  Three-dimensional reconstruction from radiographs and electron micrographs: application of convolutions instead of Fourier transforms. , 1971, Proceedings of the National Academy of Sciences of the United States of America.

[20]  Bernhard Gleich,et al.  Tomographic imaging using the nonlinear response of magnetic particles , 2005, Nature.

[21]  Patrick W. Goodwill,et al.  Multidimensional X-Space Magnetic Particle Imaging , 2011, IEEE Transactions on Medical Imaging.

[22]  Patrick W. Goodwill,et al.  Linear and Shift Invariance of Magnetic Particle Imaging , 2012 .

[23]  Bo Zheng,et al.  Projection X-Space Magnetic Particle Imaging , 2012, IEEE Transactions on Medical Imaging.

[24]  A. C. Riddle,et al.  Inversion of Fan-Beam Scans in Radio Astronomy , 1967 .

[25]  Bernhard Gleich,et al.  Magnetic particle imaging using a field free line , 2008 .

[26]  Bo Zheng,et al.  An x-space magnetic particle imaging scanner. , 2012, The Review of scientific instruments.

[27]  Bernhard Gleich,et al.  Analysis of a 3-D System Function Measured for Magnetic Particle Imaging , 2012, IEEE Transactions on Medical Imaging.

[28]  S M Jorgensen,et al.  Three-dimensional imaging of vasculature and parenchyma in intact rodent organs with X-ray micro-CT. , 1998, The American journal of physiology.

[29]  R. Pazdur,et al.  FDA report: Ferumoxytol for intravenous iron therapy in adult patients with chronic kidney disease , 2010, American journal of hematology.

[30]  Patrick W. Goodwill,et al.  The X-Space Formulation of the Magnetic Particle Imaging Process: 1-D Signal, Resolution, Bandwidth, SNR, SAR, and Magnetostimulation , 2010, IEEE Transactions on Medical Imaging.

[31]  B Gleich,et al.  Three-dimensional real-time in vivo magnetic particle imaging , 2009, Physics in medicine and biology.

[32]  Thorsten M. Buzug,et al.  A Fourier slice theorem for magnetic particle imaging using a field-free line , 2011 .