Accuracy evaluation of a new glucometer with automated hematocrit measurement and correction.

BACKGROUND Blood glucose meters are widely used in point of care testing, however, many studies have shown inaccuracies in the glucose measurement due to a number of factors. The present study evaluated the accuracy of a new glucometer capable of simultaneous measurement of patient's hematocrit with algorithmic adjustment of glucose result. This meter was compared with a reference method and 2 other existing meters widely used in the market. METHODS Venous whole blood samples from healthy volunteers were pooled and reconstituted to produce 5 different hematocrit (30-60%) concentrations. Each hematocrit specimen was spiked to produce 4 different glucose (50-500 mg/dl) concentrations. RESULTS Hematocrit measured by the new meter correlated well with the reference method. Mean percentage error differences, compared to the reference method, showed obvious differences between existing meters across the wide hematocrit range at various glucose concentrations. The new meter showed a steady and consistent glucose concentrations compared to the reference method. CONCLUSION The new glucometer, which simultaneously measures hematocrit and performs automated correction for the hematocrit effect, provides a glucose result with improved accuracy. Its measurement of hematocrit from the same blood sample will eliminate the need for additional collection of blood or measurement using another method.

[1]  F. Wirth,et al.  Incidence of neonatal hyperviscosity at sea level. , 1980, The Journal of pediatrics.

[2]  D. Benaron,et al.  Neonatal Hypoglycemia, Part I: Background and Definition , 1997, Clinical pediatrics.

[3]  K. Wiener The Effect of Haematocrit on Reagent Strip Tests for Glucose , 1991, Diabetic medicine : a journal of the British Diabetic Association.

[4]  N. Sharief,et al.  The inaccuracy of venous and capillary blood glucose measurement using reagent strips in the newborn period and the effect of haematocrit. , 2000, Early human development.

[5]  V T Innanen,et al.  Point-of-care glucose testing: cost savings and ease of use with the Ames Glucometer Elite. , 1995, Clinical chemistry.

[6]  M. Hyneck,et al.  Accuracy of four methods of home blood glucose monitoring in hemodialysis patients. , 1984, Clinical pharmacy.

[7]  A. Aynsley-Green Glucose: A fuel for thought! , 1991, Journal of paediatrics and child health.

[8]  R A Gatti,et al.  Hematocrit values of capillary blood in the newborn infant. , 1967, The Journal of pediatrics.

[9]  G J Kost,et al.  Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing. , 2009, Archives of pathology & laboratory medicine.

[10]  E. Kilpatrick,et al.  The Effect of Variations in Haematocrit, Mean Cell Volume and Red Blood Cell Count on Reagent Strip Tests for Glucose , 1993, Annals of clinical biochemistry.

[11]  A. Eidelman,et al.  Screening for hypoglycemia with plasma in neonatal blood of high hematocrit value. , 1989, Critical care medicine.

[12]  J. D. Clark,et al.  Are Blood Glucose Reagent Strips Reliable in Renal Failure? , 1991, Diabetic medicine : a journal of the British Diabetic Association.

[13]  G J Kost,et al.  Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing , 2001, Critical care medicine.

[14]  D. Chan,et al.  Laboratory and bedside evaluation of portable glucose meters. , 1995, American journal of clinical pathology.

[15]  E. Kilpatrick,et al.  Intra‐operative blood glucose measurements , 1994, Anaesthesia.

[16]  G J Kost,et al.  Point-of-care glucose testing: effects of critical care variables, influence of reference instruments, and a modular glucose meter design. , 2000, Archives of pathology & laboratory medicine.

[17]  James Stephen Krinsley,et al.  Association between hyperglycemia and increased hospital mortality in a heterogeneous population of critically ill patients. , 2003, Mayo Clinic proceedings.