Application of an image-guided navigation system in breast cancer localization

Image-guided navigation on the basis of pre-therapy images in a deformable organ, such as the breast, requires a survey of the factors that cause uncertainties. A deformable breast-tissue-mimicking phantom with simulated tumors was employed to investigate the accuracy of lesion localization with a needle instrument coupled to an optical measurement system. The RMS deviation was 1.1 mm with errors ≤ 2.0 mm in 96% of the procedures. Ultrasonography data acquired during needle localization of breast tumors were analyzed in 20 patients (23 tumors; 12 benign, 11 malignant) to investigate the deformation due to presence of instruments. The overall RMS tumor shift was 2.3 mm after release of pressure on the needle. To establish an optimal strategy to correct for breast motion due to breathing experiments with a volunteer were performed. Tracking a single centre marker was found to be most effective to improve registration accuracy. Average deviations of 8.2 mm were reduced to 1.1 mm. The combined impact of these different uncertainties resulted in distributions defined by: μ = 2.5 mm, σ = 1.4 mm (benign and malignant), μ = 3.1 mm, σ = 1.8 mm (benign), μ = 1.7 mm, σ = 0.9 mm (malignant).