Generation of virtual monochromatic CBCT from dual kV∕MV beam projections.

PURPOSE To develop a novel on-board imaging technique which allows generation of virtual monochromatic (VM) cone-beam CT (CBCT) with a selected energy from combined kilovoltage (kV)∕megavoltage (MV) beam projections. METHODS With the current orthogonal kV∕MV imaging hardware equipped in modern linear accelerators, both MV projections (from gantry angle of 0°-100°) and kV projections (90°-200°) were acquired as gantry rotated a total of 110°. A selected range of overlap projections between 90° to 100° were then decomposed into two material projections using experimentally determined parameters from orthogonally stacked aluminum and acrylic step-wedges. Given attenuation coefficients of aluminum and acrylic at a predetermined energy, one set of VM projections could be synthesized from two corresponding sets of decomposed projections. Two linear functions were generated using projection information at overlap angles to convert kV and MV projections at nonoverlap angles to approximate VM projections for CBCT reconstruction. The contrast-to-noise ratios (CNRs) were calculated for different inserts in VM CBCTs of a CatPhan phantom with various selected energies and compared with those in kV and MV CBCTs. The effect of overlap projection number on CNR was evaluated. Additionally, the effect of beam orientation was studied by scanning the CatPhan sandwiched with two 5 cm solid-water phantoms on both lateral sides and an electronic density phantom with two metal bolt inserts. RESULTS Proper selection of VM energy [30 and 40 keV for low-density polyethylene (LDPE), polymethylpentene, 2 MeV for Delrin] provided comparable or even better CNR results as compared with kV or MV CBCT. An increased number of overlap kV and MV projection demonstrated only marginal improvements of CNR for different inserts (with the exception of LDPE) and therefore one projection overlap was found to be sufficient for the CatPhan study. It was also evident that the optimal CBCT image quality was achieved when MV beams penetrated through the heavy attenuation direction of the object. CONCLUSIONS A novel technique was developed to generate VM CBCTs from kV∕MV projections. This technique has the potential to improve CNR at selected VM energies and to suppress artifacts at appropriate beam orientations.

[1]  P. Munro,et al.  Four-dimensional cone-beam computed tomography using an on-board imager. , 2006, Medical physics.

[2]  A. Macovski,et al.  Energy-selective reconstructions in X-ray computerised tomography , 1976, Physics in medicine and biology.

[3]  Patrick A Kupelian,et al.  Performance characterization of megavoltage computed tomography imaging on a helical tomotherapy unit. , 2005, Medical physics.

[4]  Jürgen Hesser,et al.  Breath-hold target localization with simultaneous kilovoltage/megavoltage cone-beam computed tomography and fast reconstruction. , 2010, International journal of radiation oncology, biology, physics.

[5]  J. Pouliot,et al.  Megavoltage cone-beam CT: system description and clinical applications. , 2006, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[6]  M. Jiang,et al.  Ordered-subset simultaneous algebraic reconstruction techniques (OS-SART) , 2004 .

[7]  J A Rowlands,et al.  Imaging of 1.0-mm-diameter radiopaque markers with megavoltage X-rays: an improved online imaging system. , 2002, International journal of radiation oncology, biology, physics.

[8]  D W O Rogers,et al.  Monte Carlo calculation of nine megavoltage photon beam spectra using the BEAM code. , 2002, Medical physics.

[9]  R. Alvarez,et al.  A Comparison of Noise and Dose in Conventional and Energy Selective Computed Tomography , 1979, IEEE Transactions on Nuclear Science.

[10]  A. Macovski,et al.  Generalized image combinations in dual KVP digital radiography. , 1981, Medical physics.

[11]  Jan-Jakob Sonke,et al.  Frameless stereotactic body radiotherapy for lung cancer using four-dimensional cone beam CT guidance. , 2009, International journal of radiation oncology, biology, physics.

[12]  Frederik Wenz,et al.  Fast kilovoltage/megavoltage (kVMV) breathhold cone-beam CT for image-guided radiotherapy of lung cancer , 2010, Physics in medicine and biology.

[13]  Fang-Fang Yin,et al.  A technique for on-board CT reconstruction using both kilovoltage and megavoltage beam projections for 3D treatment verification. , 2005, Medical physics.

[14]  F. Yin,et al.  Minimizing image noise in on-board CT reconstruction using both kilovoltage and megavoltage beam projections. , 2007, Medical physics.

[15]  A Fenster,et al.  An accurate method for direct dual-energy calibration and decomposition. , 1990, Medical physics.

[16]  L. Feldkamp,et al.  Practical cone-beam algorithm , 1984 .

[17]  Yu Zou,et al.  Analysis of fast kV-switching in dual energy CT using a pre-reconstruction decomposition technique , 2008, SPIE Medical Imaging.