Three-month validation of a turbuhaler electronic monitoring device: implications for asthma clinical trial use

Background Electronic monitoring of inhaled asthma therapy is suggested as the ‘gold standard’ for measuring patterns of medication use in clinical trials. The SmartTurbo (Adherium (NZ) Ltd, Auckland, New Zealand) is an electronic monitor for use with a turbuhaler device (AstraZeneca, UK). The aim of this study was to determine the accuracy of the SmartTurbo in recording Symbicort actuations over a 12-week period of use. Methods Twenty SmartTurbo monitors were attached to the base of 20 Symbicort turbuhalers. Bench testing in a research facility was undertaken on days 0, 5, 6, 7, 8, 9, 14, 21, 28, 56 and 84. Patterns of ‘low-use’ (2 sets of 2 actuations on the same day) and ‘high-use’ (2 sets of 8 actuations on the same day) were performed. The date and time of actuations were recorded in a paper diary and compared with data uploaded from the SmartTurbo monitors. Results 2800 actuations were performed. Monitor sensitivity was 99.9% with a lower 97.5% confidence bound of 99.6%. The positive predictive value was 99.9% with a 97.5% lower confidence bound of 99.7%. Accuracy was not affected by whether the pattern of inhaler use was low or high, or whether there was a delay in uploading the actuation data. Conclusions The SmartTurbo monitor is highly accurate in recording and retaining electronic data in this 12-week bench study. It can be recommended for use in clinical trial settings, in which quality control systems are incorporated into study protocols to ensure accurate data acquisition.

[1]  D. Shaw,et al.  Predictors of severe exacerbations, poor asthma control, and β-agonist overuse for patients with asthma. , 2014, The journal of allergy and clinical immunology. In practice.

[2]  M. Weatherall,et al.  Combination budesonide/formoterol inhaler as maintenance and reliever therapy in Māori with asthma , 2014, Respirology.

[3]  Richard B. Reilly,et al.  A Method to Assess Adherence in Inhaler Use through Analysis of Acoustic Recordings of Inhaler Events , 2014, PloS one.

[4]  R. Reilly,et al.  Acoustic Analysis of Inhaler Sounds From Community-Dwelling Asthmatic Patients for Automatic Assessment of Adherence , 2014, IEEE Journal of Translational Engineering in Health and Medicine.

[5]  R. Hancox,et al.  Combination corticosteroid/β-agonist inhaler as reliever therapy: a solution for intermittent and mild asthma? , 2014, The Journal of allergy and clinical immunology.

[6]  D. Shaw,et al.  Metrics of salbutamol use as predictors of future adverse outcomes in asthma , 2013, Clinical and experimental allergy : journal of the British Society for Allergy and Clinical Immunology.

[7]  Short-acting β-agonist use as a marker of current asthma control. , 2013, The journal of allergy and clinical immunology. In practice.

[8]  D. Shaw,et al.  Efficacy and safety of maintenance and reliever combination budesonide-formoterol inhaler in patients with asthma at risk of severe exacerbations: a randomised controlled trial. , 2013, The Lancet. Respiratory medicine.

[9]  D. Shaw,et al.  Use of metered-dose inhaler electronic monitoring in a real-world asthma randomized controlled trial. , 2013, The journal of allergy and clinical immunology. In practice.

[10]  B. Bender Advancing the science of adherence measurement: implications for the clinician. , 2013, The journal of allergy and clinical immunology. In practice.

[11]  R. Beasley,et al.  Six-month in vitro validation of a metered-dose inhaler electronic monitoring device: implications for asthma clinical trial use. , 2012, Journal of Allergy and Clinical Immunology.

[12]  C. Rand,et al.  The Reliability and Patient Acceptability of the SmartTrack Device: A New Electronic Monitor and Reminder Device for Metered Dose Inhalers , 2012, The Journal of asthma : official journal of the Association for the Care of Asthma.

[13]  T. Haahtela,et al.  True adherence with the Turbuhaler in young children with asthma , 2010, Archives of Disease in Childhood.

[14]  C. Rand,et al.  Electronic Monitoring of Medication Adherence: When Is High-Tech Best? , 2002, Journal of Clinical Psychology in Medical Settings.

[15]  L. Hendeles,et al.  Accuracy of three electronic monitors for metered-dose inhalers. , 2002, Chest.