Performance assessment of a new dynamic scan mode for perfusion computed tomography using a biological phantom

Perfusion computed tomography is increasingly being used for stroke and tumor assessment. Using continuous periodic table movement the spatial coverage can be increased beyond the detector width with a new adaptive spiral scanning technique (A4DS). The purpose of this study was to compare perfusion values acquired with the A4DS technique with results from standard dynamic scans at different temporal sampling rates. A biological perfusion phantom (preserved porcine kidney) was scanned with both techniques. In standard mode three scans were performed at adjacent overlapping positions (detector width 38.4 mm) covering the whole phantom. Data were reconstructed with temporal resolutions of 0.5, 1 and 1.5 s. The A4DS scan was performed with a cycle time of 1.5 s and scan ranges of 100 and 148 mm respectively. The phantom was not repositioned between scans in order to assure that identical image slices showed identical phantom slices. Tissue flow was calculated with a deconvolution type algorithm. Regions of interest were drawn in strongly and moderately enhancing areas and around the whole cortex in three slices in the upper, central and lower portion of the phantom. In the flow range of 40 to 100 ml/100ml/min values did not differ by more than 5 ml/100ml/min between any of the scan protocols used. The correlation between the continuous table movement modes and the 0.5 s standard mode was excellent (r2>0.98) indicating that the new mode is well suited for perfusion measurements and allows increasing the coverage by almost a factor of four.

[1]  Ernst Klotz,et al.  Perfusion CT: noninvasive surrogate marker for stratification of pancreatic cancer response to concurrent chemo- and radiation therapy. , 2009, Radiology.

[2]  Martin Mory,et al.  [An experimental organ model for magnetic resonance imaging]. , 2007, Zeitschrift fur medizinische Physik.

[3]  Steve Halligan,et al.  Colorectal tumor vascularity: quantitative assessment with multidetector CT--do tumor perfusion measurements reflect angiogenesis? , 2008, Radiology.

[4]  Jürgen Biederer,et al.  Artificial thorax for MR imaging studies in porcine heart-lung preparations. , 2003, Radiology.

[5]  Ashok Srinivasan,et al.  State-of-the-Art Imag- ing of Acute Stroke 1 , 2006 .

[6]  Thomas Lehnert,et al.  Quantitative Measurements of Perfusion and Permeability of Oropharyngeal and Oral Cavity Cancer, Recurrent Disease, and Associated Lymph Nodes Using First-Pass Contrast-Enhanced Computed Tomography Studies , 2007, Investigative radiology.

[7]  E. Klotz,et al.  Comprehensive imaging of ischemic stroke with multisection CT. , 2003, Radiographics : a review publication of the Radiological Society of North America, Inc.

[8]  Behzad Ebrahimi,et al.  A perfusion phantom for quantitative medical imaging , 2008, SPIE Medical Imaging.

[9]  Kazuhiro Katada,et al.  Whole-brain perfusion CT performed with a prototype 256-detector row CT system: initial experience. , 2009, Radiology.

[10]  J. Babb,et al.  Measuring Elevated Microvascular Permeability and Predicting Hemorrhagic Transformation in Acute Ischemic Stroke Using First-Pass Dynamic Perfusion CT Imaging , 2007, American Journal of Neuroradiology.

[11]  D R Pickens,et al.  Magnetic resonance perfusion/diffusion imaging of the excised dog kidney. , 1992, Investigative radiology.

[12]  Sung Hyun Kim,et al.  Perfusion CT of the brain using 40-mm-wide detector and toggling table technique for initial imaging of acute stroke. , 2008, AJR. American journal of roentgenology.

[13]  Vicky Goh,et al.  Acute tumor vascular effects following fractionated radiotherapy in human lung cancer: In vivo whole tumor assessment using volumetric perfusion computed tomography. , 2007, International journal of radiation oncology, biology, physics.

[14]  Ernst Klotz,et al.  A Biological Phantom for Contrast-Media-Based Perfusion Studies With CT , 2009, Investigative radiology.