An automatic method for determining the centre of rotation of a mechanically scanned UCT system
暂无分享,去创建一个
A method is described for determining the centre of rotation of a mechanically scanned reflection ultrasound computed tomography system. It is based on the principle of obtaining opposing images of a test object containing many point targets. The method is automatic in the sense that the centre of rotation is calculated by a computer without the need for an operator to make direct measurements on the mechanical system. For the particular reflection UCT system described here, the centre of rotation is obtained in 3-5 min with a repeatability (+or-2 SD) of +or-0.3 mm. Ways in which even higher accuracy can be obtained are discussed. The basic principle of the method is applicable to any concentric imaging system for which a good approximation to an ideal point target can be produced.