Analysis and correction of motion artifacts in diffusion weighted imaging
暂无分享,去创建一个
For diffusion‐weighted magnetic resonance imaging and under circumstances where patient movement can be modeled as rigid body motion, it is shown both theoretically and experimentally that translations and rotations produce phase errors which are zero‐ and first‐order, respectively, in position. Whlile a navigator echo can be used to correct the imaging data for arbitrary translations, only when the diffusion gradient is applied in the phase encode direction is there sufficient information to correct for rotations around all axes, and therefore for general rigid body motion. Experiments in test objects and human brain imaging confirm theoretical predictions and demonstrate that appropriate corrections dramatically improve image quality in vivo.
[1] J. Kucharczyk,et al. Early detection of regional cerebral ischemia in cats: Comparison of diffusion‐ and T2‐weighted MRI and spectroscopy , 1990, Magnetic resonance in medicine.
[2] P. Grenier,et al. MR imaging of intravoxel incoherent motions: application to diffusion and perfusion in neurologic disorders. , 1986, Radiology.
[3] R. Ordidge,et al. Correction of motional artifacts in diffusion-weighted MR images using navigator echoes. , 1994, Magnetic resonance imaging.