Delayed gadolinium-enhanced MR imaging of articular cartilage: three-dimensional T1 mapping with variable flip angles and B1 correction.

PURPOSE To develop and verify the accuracy of a rapid imaging protocol for delayed gadolinium-enhanced magnetic resonance (MR) imaging of cartilage that was based on three-dimensional (3D) spoiled gradient-recalled acquisition in the steady state (SPGR) sequences with variable flip angles (FAs) (VFAs) and where a correction method for B(1) field inhomogeneities was applied. MATERIALS AND METHODS The institutional research ethics board approved this study. Written informed consent was obtained from all subjects. A B(1) field inhomogeneity correction method was applied to a 3D SPGR pulse sequence with VFAs (repetition time msec/echo time msec, 7.1/3.3; FAs, 2 degrees , 5 degrees , 10 degrees , and 20 degrees ) and was used to perform delayed gadolinium-enhanced MR imaging of cartilage 3D T1 measurements at 1.5 T. The 3D T1 measurements were validated with the reference standard (the results of T1 mapping by using a single-section two-dimensional [2D] inversion-recovery [IR] fast spin-echo [SE] pulse sequence in vitro and in vivo) in six healthy volunteers. RESULTS T1 values calculated from 3D T1 maps were not significantly different from reference T1 values in vitro (P = .195) and in vivo (P = .52) when a B(1) field inhomogeneity correction method was applied. In vivo T1 mapping of the articular surface of the whole femoropatellar joint, including data acquisition, was performed in approximately 8 minutes of acquisition time at a spatial resolution of 0.55 x 0.55 x 3.00 mm. CONCLUSION Rapid T1 mapping by using 3D SPGR acquisitions with a VFA approach and a correction for B(1) field inhomogeneities can be used for delayed gadolinium-enhanced MR imaging of cartilage. T1 measurements performed in vitro and in vivo by using this approach are highly accurate when compared with those performed by using standard 2D IR fast SE T1 mapping as a reference.

[1]  B K Rutt,et al.  A fast 3D look-locker method for volumetric T1 mapping. , 1999, Magnetic resonance imaging.

[2]  P. E. Crewson,et al.  Hypothesis testing II: means. , 2003, Radiology.

[3]  D. Burstein,et al.  Gd‐DTPA2− as a measure of cartilage degradation , 1996, Magnetic resonance in medicine.

[4]  Fernando Boada,et al.  MRI of articular cartilage in OA: novel pulse sequences and compositional/functional markers. , 2006, Osteoarthritis and cartilage.

[5]  W E Reddick,et al.  Correction of errors caused by imperfect inversion pulses in MR imaging measurement of T1 relaxation times. , 1998, Magnetic resonance imaging.

[6]  Carl Johan Tiderius,et al.  Delayed gadolinium‐enhanced MRI of cartilage (dGEMRIC) in early knee osteoarthritis , 2003, Magnetic resonance in medicine.

[7]  D. Burstein,et al.  Three‐dimensional delayed gadolinium‐enhanced MRI of cartilage (dGEMRIC) at 1.5T and 3.0T , 2006, Journal of magnetic resonance imaging : JMRI.

[8]  D. Burstein,et al.  Glycosaminoglycan in articular cartilage: in vivo assessment with delayed Gd(DTPA)(2-)-enhanced MR imaging. , 1997, Radiology.

[9]  Ravinder R Regatte,et al.  Rapid 3D‐T1 mapping of cartilage with variable flip angle and parallel imaging at 3.0T , 2008, Journal of magnetic resonance imaging : JMRI.

[10]  Colin Studholme,et al.  Automated 3-D registration of MR and CT images of the head , 1996, Medical Image Anal..

[11]  G. Wright,et al.  Rapid high‐resolution T1 mapping by variable flip angles: Accurate and precise measurements in the presence of radiofrequency field inhomogeneity , 2006, Magnetic resonance in medicine.

[12]  D. Burstein,et al.  Nondestructive imaging of human cartilage glycosaminoglycan concentration by MRI , 1999, Magnetic resonance in medicine.

[13]  Ying Wu,et al.  Accuracy of T1 measurement with 3‐D Look‐Locker technique for dGEMRIC , 2008, Journal of Magnetic Resonance Imaging.

[14]  Martha L Gray,et al.  Assessment of early osteoarthritis in hip dysplasia with delayed gadolinium-enhanced magnetic resonance imaging of cartilage. , 2003, The Journal of bone and joint surgery. American volume.

[15]  S. Schoenberg,et al.  Measurement of signal‐to‐noise ratios in MR images: Influence of multichannel coils, parallel imaging, and reconstruction filters , 2007, Journal of magnetic resonance imaging : JMRI.

[16]  K. T. Scott,et al.  Protocol issues for delayed Gd(DTPA)2–‐enhanced MRI (dGEMRIC) for clinical evaluation of articular cartilage , 2001, Magnetic resonance in medicine.

[17]  P Leander,et al.  Gd‐DTPA2–‐enhanced MRI of femoral knee cartilage: A dose‐response study in healthy volunteers , 2001, Magnetic resonance in medicine.

[18]  Bruce C. Po,et al.  Glycosaminoglycan distribution in cartilage as determined by delayed gadolinium-enhanced MRI of cartilage (dGEMRIC): potential clinical applications. , 2004, AJR. American journal of roentgenology.

[19]  Michael B. Smith,et al.  SAR and B1 field distributions in a heterogeneous human head model within a birdcage coil , 1998, Magnetic resonance in medicine.

[20]  Pottumarthi Prasad,et al.  Three-Dimensional T1 Mapping for dGEMRIC at 3.0 T Using the Look Locker Method , 2006, Investigative radiology.

[21]  Thomas M. Link,et al.  Cartilage imaging: motivation, techniques, current and future significance , 2007, European Radiology.

[22]  T W Redpath,et al.  Accuracy of T1 measurement in dynamic contrast‐enhanced breast MRI using two‐ and three‐dimensional variable flip angle fast low‐angle shot , 1999, Journal of magnetic resonance imaging : JMRI.