Non-contact, ultrasound-based indentation method for measuring elastic properties of biological tissues using Harmonic Motion Imaging (HMI)

Noninvasive measurement of mechanical properties of biological tissues in vivo could play a significant role in improving the current understanding of tissue biomechanics. In this study, we propose a method for measuring elastic properties non-invasively by using internal indentation as generated by harmonic motion imaging (HMI). In HMI, an oscillating acoustic radiation force is produced by a focused ultrasound transducer at the focal region, and the resulting displacements are estimated by tracking radiofrequency signals acquired by an imaging transducer. In this study, the focal spot region was modeled as a rigid cylindrical piston that exerts an oscillatory, uniform internal force to the underlying tissue. The HMI elastic modulus EHMI was defined as the ratio of the applied force to the axial strain measured by 1D ultrasound imaging. The accuracy and the precision of the EHMI estimate were assessed both numerically and experimentally in polyacrylamide tissue-mimicking phantoms. Initial feasibility of this method in soft tissues was also shown in canine liver specimens in vitro. Very good correlation and agreement was found between the measured Young's modulus and the HMI modulus in the numerical study (r(2) > 0.99, relative error <10%) and on polyacrylamide gels (r(2) = 0.95, relative error <24%). The average HMI modulus on five liver samples was found to EHMI = 2.62  ±  0.41 kPa, compared to EMechTesting = 4.2  ±  2.58 kPa measured by rheometry. This study has demonstrated for the first time the initial feasibility of a non-invasive, model-independent method to estimate local elastic properties of biological tissues at a submillimeter scale using an internal indentation-like approach. Ongoing studies include in vitro experiments in a larger number of samples and feasibility testing in in vivo models as well as pathological human specimens.

[1]  Philip V Bayly,et al.  Measurement of the dynamic shear modulus of mouse brain tissue in vivo by magnetic resonance elastography. , 2008, Journal of biomechanical engineering.

[2]  Y. Fung Elasticity of soft tissues in simple elongation. , 1967, The American journal of physiology.

[3]  Elisa E. Konofagou,et al.  Single-Element Focused Ultrasound Transducer Method for Harmonic Motion Imaging , 2006, Ultrasonic imaging.

[4]  Richard L Ehman,et al.  Noninvasive muscle tension measurement using the novel technique of magnetic resonance elastography (MRE). , 2003, Journal of biomechanics.

[5]  I. N. Sneddon The relation between load and penetration in the axisymmetric boussinesq problem for a punch of arbitrary profile , 1965 .

[6]  Roger W Nightingale,et al.  An Integrated Indenter-ARFI Imaging System for Tissue Stiffness Quantification , 2008, Ultrasonic imaging.

[7]  K. Nightingale,et al.  Quantifying hepatic shear modulus in vivo using acoustic radiation force. , 2008, Ultrasound in medicine & biology.

[8]  Thomas Deffieux,et al.  Shear Wave Spectroscopy for In Vivo Quantification of Human Soft Tissues Visco-Elasticity , 2009, IEEE Transactions on Medical Imaging.

[9]  S. Emelianov,et al.  Shear wave elasticity imaging: a new ultrasonic technology of medical diagnostics. , 1998, Ultrasound in medicine & biology.

[10]  G. Pharr,et al.  An improved technique for determining hardness and elastic modulus using load and displacement sensing indentation experiments , 1992 .

[11]  William F. Walker,et al.  Sonorheometry: A Noncontact Method for the Dynamic Assessment of Thrombosis , 2004, Annals of Biomedical Engineering.

[12]  Mathias Fink,et al.  Human muscle hardness assessment during incremental isometric contraction using transient elastography. , 2005, Journal of biomechanics.

[13]  Ralph Sinkus,et al.  Magnetic resonance elastography , 2013 .

[14]  Mostafa Fatemi,et al.  Quantifying elasticity and viscosity from measurement of shear wave speed dispersion. , 2004, The Journal of the Acoustical Society of America.

[15]  Gregg Trahey,et al.  Observations of Tissue Response to Acoustic Radiation Force: Opportunities for Imaging , 2002, Ultrasonic imaging.

[16]  Guy Cloutier,et al.  Characterization of blood clot viscoelasticity by dynamic ultrasound elastography and modeling of the rheological behavior. , 2011, Journal of biomechanics.

[17]  Ichiro Sakuma,et al.  In vitro Measurement of Mechanical Properties of Liver Tissue under Compression and Elongation Using a New Test Piece Holding Method with Surgical Glue , 2003, IS4TH.

[18]  K D Paulsen,et al.  Time-harmonic magnetic resonance elastography of the normal feline brain. , 2010, Journal of biomechanics.

[19]  D Farina,et al.  Blind source identification from the multichannel surface electromyogram , 2014, Physiological measurement.

[20]  Jonathan Vappou Magnetic resonance- and ultrasound imaging-based elasticity imaging methods: a review. , 2012, Critical reviews in biomedical engineering.

[21]  Zachary J. Domire,et al.  Feasibility of using magnetic resonance elastography to study the effect of aging on shear modulus of skeletal muscle. , 2009, Journal of applied biomechanics.

[22]  Shigao Chen,et al.  Shearwave dispersion ultrasound vibrometry (SDUV) for measuring tissue elasticity and viscosity , 2009, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[23]  Caroline Maleke,et al.  Quantitative viscoelastic parameters measured by harmonic motion imaging , 2009, Physics in medicine and biology.

[24]  Ralph Sinkus,et al.  In vivo brain viscoelastic properties measured by magnetic resonance elastography , 2008, NMR in biomedicine.

[25]  H. Haslach,et al.  Nonlinear viscoelastic, thermodynamically consistent, models for biological soft tissue , 2005, Biomechanics and modeling in mechanobiology.

[26]  Rémy Willinger,et al.  Assessment of in vivo and post-mortem mechanical behavior of brain tissue using magnetic resonance elastography. , 2008, Journal of biomechanics.

[27]  Jianwen Luo,et al.  Performance assessment of HIFU lesion detection by harmonic motion imaging for focused ultrasound (HMIFU): a 3-D finite-element-based framework with experimental validation. , 2011, Ultrasound in medicine & biology.

[28]  E E Konofagou,et al.  Harmonic motion imaging for focused ultrasound (HMIFU): a fully integrated technique for sonication and monitoring of thermal ablation in tissues , 2008, Physics in medicine and biology.

[29]  P. Asbach,et al.  Noninvasive assessment of the rheological behavior of human organs using multifrequency MR elastography: a study of brain and liver viscoelasticity , 2007, Physics in medicine and biology.

[30]  R. Ogden,et al.  A New Constitutive Framework for Arterial Wall Mechanics and a Comparative Study of Material Models , 2000 .

[31]  Jianwen Luo,et al.  Simulation Study of Amplitude-Modulated (AM) Harmonic Motion Imaging (HMI) for Stiffness Contrast Quantification with Experimental Validation , 2010, Ultrasonic imaging.