Influence of some instrumental parameters on the determination of quantitative data using the PET scanner PC4096-15WB

In order to evaluate the influence of physical and instrumental limitations of the PET technique on metabolic data, various parameters have been tested. The measurements were done with a whole body PET scanner PC4096-15WB. The dependence on the transaxial resolution of reconstruction filter cut off frequency αc was investigated. The high resolution (4.8 mm atr=0) in the field of view) and the low scatter fraction (<5%) of this system yields a higher grey/white matter ratio (3.8:1) than that found with older scanners (2:1). The use of a rotating pin source allows transmission scans to be performed after tracer administration with an error of less than 5%.

[1]  L. Feinendegen,et al.  A preliminary evaluation of a positron camera system using weighted decoding of individual crystals , 1988 .

[2]  M. Daube-Witherspoon,et al.  A method for postinjection PET transmission measurements with a rotating source. , 1988, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[3]  K. Herholz,et al.  Regional Kinetic Constants and Cerebral Metabolic Rate for Glucose in Normal Human Volunteers Determined by Dynamic Positron Emission Tomography of [18F]-2-Fluoro-2-Deoxy-D-Glucose , 1984, Journal of cerebral blood flow and metabolism : official journal of the International Society of Cerebral Blood Flow and Metabolism.