The geometric accuracy of frameless stereotactic radiosurgery using a 6D robotic couch system

The aim of this paper is to assess the overall geometric accuracy of the Novalis system using the Robotic Tilt Module in terms of the uncertainty in frameless stereotactic radiotherapy. We analyzed the following three metrics: (1) the correction accuracy of the robotic couch, (2) the uncertainty of the isocenter position with gantry and couch rotation, and (3) the shift in position between the isocenter and central point detected with the ExacTrac x-ray system. Based on the concept of uncertainty, the overall accuracy was calculated from these values. The accuracy in positional correction with the robotic couch was 0.07 +/- 0.22 mm, the positional shift of the isocenter associated with gantry rotation was 0.35 mm, the positional shift of the isocenter associated with couch rotation was 0.38 mm and the difference in position between the isocenter and the ExacTrac x-ray system was 0.30 mm. The accuracy of intracranial stereotactic radiosurgery with the Novalis system in our clinic was 0.31 +/- 0.77 mm. The overall geometric accuracy based on the concept of uncertainty was 0.31 +/- 0.77 mm, which is within the tolerance given in the American Association of Physicists in Medicine report no. 54.

[1]  R. Onimaru,et al.  Hypofractionated stereotactic radiotherapy alone without whole-brain irradiation for patients with solitary and oligo brain metastasis using noninvasive fixation of the skull. , 2003, International journal of radiation oncology, biology, physics.

[2]  A Ertl,et al.  Quality assurance for the Leksell gamma unit: considering magnetic resonance image-distortion and delineation failure in the targeting of the internal auditory canal. , 1999, Medical physics.

[3]  Matthias Guckenberger,et al.  Precision of Image-Guided Radiotherapy (IGRT) in Six Degrees of Freedom and Limitations in Clinical Practice , 2007, Strahlentherapie und Onkologie.

[4]  Stanley H Benedict,et al.  Intracranial stereotactic positioning systems: Report of the American Association of Physicists in Medicine Radiation Therapy Committee Task Group No. 68. , 2005, Medical physics.

[5]  F. Yin,et al.  Clinical assessment and characterization of a dual‐tube kilovoltage X‐ray localization system in the radiotherapy treatment room , 2008, Journal of applied clinical medical physics.

[6]  Fang-Fang Yin,et al.  Use of the BrainLAB ExacTrac X-Ray 6D system in image-guided radiotherapy. , 2008, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[7]  Kenneth R Hogstrom,et al.  Accuracy of cranial coplanar beam therapy using an oblique, stereoscopic x-ray image guidance system. , 2008, Medical physics.

[8]  Jenghwa Chang,et al.  Accuracy and feasibility of cone-beam computed tomography for stereotactic radiosurgery setup. , 2007, Medical physics.

[9]  M. Brock,et al.  NOVALIS FRAMELESS IMAGE‐GUIDED NONINVASIVE RADIOSURGERY: INITIAL EXPERIENCE , 2008, Neurosurgery.

[10]  K. Wallner,et al.  Radiographic findings and morbidity in patients treated with stereotactic radiosurgery. , 1998, International journal of radiation oncology, biology, physics.

[11]  Matthias Guckenberger,et al.  Positioning accuracy of cone-beam computed tomography in combination with a HexaPOD robot treatment table. , 2007, International journal of radiation oncology, biology, physics.

[12]  K. Winston,et al.  A system for stereotactic radiosurgery with a linear accelerator. , 1986, International journal of radiation oncology, biology, physics.

[13]  R Bale,et al.  Repositioning accuracy: comparison of a noninvasive head holder with thermoplastic mask for fractionated radiotherapy and a case report. , 1998, International journal of radiation oncology, biology, physics.

[14]  M. Mehta,et al.  A multi-institutional review of radiosurgery alone vs. radiosurgery with whole brain radiotherapy as the initial management of brain metastases. , 2002, International journal of radiation oncology, biology, physics.

[15]  Cristina Garibaldi,et al.  3D optoelectronic analysis of interfractional patient setup variability in frameless extracranial stereotactic radiotherapy. , 2006, International journal of radiation oncology, biology, physics.

[16]  Carlo Cavedon,et al.  Cone beam CT image guidance for intracranial stereotactic treatments: comparison with a frame guided set-up. , 2007, International journal of radiation oncology, biology, physics.

[17]  L. Court,et al.  Evaluation of the precision of portal-image-guided head-and-neck localization: an intra- and interobserver study. , 2007, Medical physics.

[18]  R E Albright,et al.  Stereotactic radiosurgery for the treatment of brain metastases , 1997, Cancer.

[19]  Dirk Verellen,et al.  Quality assurance of a system for improved target localization and patient set-up that combines real-time infrared tracking and stereoscopic X-ray imaging. , 2003, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[20]  J. Wong,et al.  Flat-panel cone-beam computed tomography for image-guided radiation therapy. , 2002, International journal of radiation oncology, biology, physics.

[21]  M Bellerive,et al.  Adaptation and verification of the relocatable Gill-Thomas-Cosman frame in stereotactic radiotherapy. , 1994, International journal of radiation oncology, biology, physics.

[22]  F. Lohr,et al.  Radiosurgery alone or in combination with whole-brain radiotherapy for brain metastases. , 1998, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.

[23]  Stanley J. Rosenthal,et al.  A precision cranial immobilization system for conformal stereotactic fractionated radiation therapy. , 1995, International journal of radiation oncology, biology, physics.

[24]  Stanley H Benedict,et al.  Intracranial stereotactic positioning systems: Report of the American Association of Physicists in Medicine Radiation Therapy Committee Task Group no. 68. , 2005, Medical physics.

[25]  R. Laing,et al.  Fractionated stereotactic external beam radiotherapy in the management of brain metastases. , 1993, European journal of cancer.

[26]  Hui Yan,et al.  A phantom study on the positioning accuracy of the Novalis Body system. , 2003, Medical physics.

[27]  H. Abe,et al.  Stereotactic irradiation without whole-brain irradiation for single brain metastasis. , 1997, International journal of radiation oncology, biology, physics.

[28]  E. Shaw,et al.  Current radiosurgery practice: results of an astro survey , 1994 .

[29]  Michael J. Miller,et al.  Geometrical accuracy of the Novalis stereotactic radiosurgery system for trigeminal neuralgia. , 2004, Journal of neurosurgery.

[30]  J. Buatti,et al.  The University of Florida frameless high-precision stereotactic radiotherapy system. , 1997, International journal of radiation oncology, biology, physics.

[31]  Jun-Sang Kim,et al.  Isocenter accuracy in frameless stereotactic radiotherapy using implanted fiducials. , 2003, International journal of radiation oncology, biology, physics.