Abstract
We have fabricated and measured the thermal and DC electrical properties of transition-edge sensing (TES) bolometers to be employed on the SPIDER experiment, a balloon-based observatory for studying the polarization of the cosmic microwave background (CMB). The bolometers consist of Al and Ti thermistors in series and a termination resistor which couples to an on-chip, polarization sensitive, 150 GHz slot-array antenna through a superconducting microstrip transmission line. Several important parameters were measured. Transition temperature measurements were performed by measuring the Johnson noise in the Ti thermistor. Current-voltage characteristic measurements were performed at various temperatures allowing for the deduction of the thermal conductance and the temperature coefficient of resistance. Electrical noise equivalent power was measured to sub-Hertz frequencies. Finally, the time constant of the bolometers was measured within the Al and Ti transitions where electrothermal feedback speeds up the bolometer response compared with the natural time constant measured just above the Ti transition temperature. The results of these measurements are within the design specifications for SPIDER.