A Monte Carlo investigation of the dosimetric characteristics of the VariSource 192Ir high dose rate brachytherapy source.

An analytical Monte Carlo simulation code, incorporating in detail source construction and dimensions, was used to investigate the dosimetric characteristics of the VariSource 192Ir high dose rate brachytherapy source. Dose-rate constant, radial dose functions, and anisotropy functions, utilized in the AAPM Task Group 43 dose estimation formalism, were calculated with the source centered in a spherical water phantom of 30 cm in diameter. The results, which are in agreement with the corresponding data available in the literature, are compared to those obtained in our previous study for the widely used microSelectron 192Ir high dose-rate brachytherapy source. The dose-rate constant was found to be equal to 1.043 +/- 0.005 cGy h(-1) U(-1) for the VariSource, compared to a value of 1.116 +/- 0.006 cGy h(-1) U(-1) calculated for the microSelectron. Significant differences in the anisotropy of the two sources are observed only for polar angles close to their long axis and are due to their different dimensions.

[1]  Ruqing Wang,et al.  Monte Carlo dosimetry of the VariSource high dose rate 192Ir source. , 1998, Medical physics.

[2]  L. Sakelliou,et al.  Dose rate distributions around 60Co, 137Cs, 198Au, 192Ir, 241Am, 125I (models 6702 and 6711) brachytherapy sources and the nuclide 99Tcm. , 1992, Physics in medicine and biology.

[3]  F. Waterman,et al.  Anisotropy of an 192iridium high dose rate source measured with a miniature ionization chamber. , 1997, Medical physics.

[4]  L. Sakelliou,et al.  Monte Carlo and TLD dosimetry of an 192Ir high dose-rate brachytherapy source. , 1998, Medical physics.

[5]  A. Nahum Perturbation effects in dosimetry: Part I. Kilovoltage x-rays and electrons. , 1996, Physics in medicine and biology.

[6]  L. Anderson,et al.  Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43 , 1995 .

[7]  G Zarris,et al.  Accurate Monte Carlo calculations of the combined attenuation and build-up factors, for energies (20-1500 keV) and distances (0-10 cm) relevant in brachytherapy. , 1991, Physics in medicine and biology.

[8]  J F Williamson,et al.  Monte Carlo aided dosimetry of the microselectron pulsed and high dose-rate 192Ir sources. , 1995, Medical physics.

[9]  S. H. Cho,et al.  Anisotropy measurements of a high dose rate Ir-192 source in air and in polystyrene. , 1994, Medical physics.

[10]  C Kolotas,et al.  Application of the Monte Carlo integration (MCI) method for calculation of the anisotropy of 192Ir brachytherapy sources. , 1998, Physics in medicine and biology.

[11]  J F Williamson,et al.  Comparison of measured and calculated dose rates in water near I-125 and Ir-192 seeds. , 1991, Medical physics.

[12]  Jose Perez-Calatayud,et al.  Monte Carlo calculation of dose rate distributions around Ir192 wires , 1997 .