Evaluation of a Cerebral Oximeter as a Monitor of Cerebral Ischemia during Carotid Endarterectomy

BackgroundStroke is an important contributor to perioperative morbidity and mortality associated with carotid endarterectomy (CEA). This investigation was designed to compare the performance of the INVOS-3100 cerebral oximeter to neurologic function, as a means of detecting cerebral ischemia induced by carotid cross-clamping, in patients undergoing carotid endarterectomy with cervical plexus block. MethodsNinety-nine patients undergoing 100 CEAs with regional anesthesia (deep or superficial cervical plexus block) were studied. Bilateral regional cerebrovascular oxygen saturation (rSO2) was monitored using the INVOS-3100 cerebral oximeter. Patients were retrospectively assigned to one of two groups: those in whom a change in mental status or contralateral motor deficit was noted after internal carotid clamping (neurologic symptoms; n = 10) and those who did not show any neurologic change (no neurologic symptoms; n = 90). Data from 94 operations (neurologic symptoms = 10 and no neurologic symptoms = 84) were adequate for statistical analyses for group comparisons. A relative decrease in ipsilateral rSO2 after carotid occlusion (calculated as a percentage of preocclusion value) during all operations (n = 100) was also calculated to determine the critical level of rSO2 decrease associated with a change in neurologic function. ResultsThe mean (± SD) decrease in rSO2 after carotid occlusion in the neurologic symptoms group (from 63.2 ± 8.4% to 51.0 ± 11.6%) was significantly greater (P = 0.0002) than in the no neurologic symptoms group (from 65.8 ± 8.5% to 61.0 ± 9.3%). Logistic regression analysis used to determine if a change in rSO2, calculated as a percentage of preclamp value, could be used to predict change in neurologic function was highly significant (likelihood ratio chi-square = 13.7;P = 0.0002). A 20% decrease in rSO2 reading from the preclamp baseline, as a predictor of neurologic compromise, resulted in a sensitivity of 80% and specificity of 82.2%. The false-positive rate using this cutoff point was 66.7%, and the false-negative rate was 2.6%, providing a positive predictive value of 33.3% and a negative predictive value of 97.4%. ConclusionMonitoring rSO2 with INVOS-3100 to detect cerebral ischemia during CEA has a high negative predictive value, but the positive predictive value is low.

[1]  D. Sackett,et al.  Benefit of carotid endarterectomy in patients with symptomatic moderate or severe stenosis. North American Symptomatic Carotid Endarterectomy Trial Collaborators. , 1998, The New England journal of medicine.

[2]  R. Sclabassi,et al.  Cerebral monitoring by means of oximetry and somatosensory evoked potentials during carotid endarterectomy. , 1998, Journal of neurosurgery.

[3]  K. Roberts,et al.  NEAR INFRARED SPECTROSCOPY DETECTS CRITICAL CEREBRAL HYPOXIA DURING CAROTID ENDARTERECTOMY IN AWAKE PATIENTS , 1998 .

[4]  D. Ward,et al.  Accuracy of a Cerebral Oximeter in Healthy Volunteers under Conditions of Isocapnic Hypoxia , 1998, Anesthesiology.

[5]  W. Trubel,et al.  The contribution of extracranial blood oxygenation on near‐infrared spectroscopy during carotid thrombendarterectomy , 1997, Anaesthesia.

[6]  G. Mead,et al.  Cerebral oximetry: a useful monitor during carotid artery surgery , 1996, Anaesthesia.

[7]  D. Prough,et al.  Validation in Volunteers of a Near-Infrared Spectroscope for Monitoring Brain Oxygenation In Vivo , 1996, Anesthesia and analgesia.

[8]  J. Wildsmith,et al.  Cerebral oximetry: a useful monitor during carotid artery surgery , 1995, Anaesthesia.

[9]  K. Hongo,et al.  Noninvasive cerebral optical spectroscopy: depth-resolved measurements of cerebral haemodynamics using indocyanine green. , 1995, Neurological research.

[10]  R. Nelson,et al.  Extracerebral absorption of near infrared light influences the detection of increased cerebral oxygenation monitored by near infrared spectroscopy. , 1995, Journal of neurology, neurosurgery, and psychiatry.

[11]  L. Erdoes,et al.  Critical carotid artery stenosis: diagnosis, timing of surgery, and outcome. , 1994, Journal of vascular surgery.

[12]  N. Kane,et al.  Near-infrared spectroscopy in adults: effects of extracranial ischaemia and intracranial hypoxia on estimation of cerebral oxygenation. , 1994, British journal of anaesthesia.

[13]  A. Mortimer,et al.  Light‐reflective cerebral oximetry and jugular bulb venous oxygen saturation during carotid endarterectomy , 1994, The British journal of surgery.

[14]  D. Sackett,et al.  Beneficial effect of carotid endarterectomy in symptomatic patients with high-grade carotid stenosis. , 1991, The New England journal of medicine.

[15]  A. Lam,et al.  Monitoring electrophysiologic function during carotid endarterectomy: a comparison of somatosensory evoked potentials and conventional electroencephalogram. , 1991, Anesthesiology.

[16]  K. Ktenidis,et al.  Intraoperative assessment of cerebral ischemia during carotid surgery. , 1990, The Journal of cardiovascular surgery.

[17]  B. Gewertz,et al.  Intraoperative monitoring during carotid endarterectomy. , 1987, Current problems in surgery.

[18]  T. Sundt,et al.  Correlation of Regional Cerebral Blood Flow (rCBF) with EEG Changes During Isoflurane Anesthesia for Carotid Endarterectomy: Critical rCBF , 1987, Anesthesiology.

[19]  W. Blume,et al.  Significance of EEG changes at carotid endarterectomy. , 1986, Stroke.

[20]  F W Sharbrough,et al.  Correlation of cerebral blood flow and electroencephalographic changes during carotid endarterectomy: with results of surgery and hemodynamics of cerebral ischemia. , 1981, Mayo Clinic proceedings.

[21]  L. Symon,et al.  Central conduction time as an index of ischaemia in subarachnoid haemorrhage , 1979, Journal of the Neurological Sciences.

[22]  F. Jöbsis Noninvasive, infrared monitoring of cerebral and myocardial oxygen sufficiency and circulatory parameters. , 1977, Science.

[23]  J. Pniewski,et al.  Comparison of near-infrared spectroscopy and somatosensory evoked potentials for the detection of cerebral ischemia during carotid endarterectomy. , 1999, Stroke.

[24]  J. Kirkpatrick,et al.  Benefit of carotid endarterectomy in patients with symptomatic moderate or severe stenosis. , 1998, Journal of insurance medicine.

[25]  A. Algra,et al.  Near-infrared reflected spectroscopy and electroencephalography during carotid endarterectomy--in search of a new shunt criterion. , 1998, Neurological research.

[26]  J. Stanley,et al.  Cerebral oximetry in patients undergoing carotid endarterectomy under regional anesthesia. , 1996, Stroke.

[27]  M. Goldstein,et al.  Endarterectomy for asymptomatic carotid artery stenosis. , 1995, JAMA.