Rapid 3-D cone-beam reconstruction with the simultaneous algebraic reconstruction technique (SART) using 2-D texture mapping hardware

Algebraic reconstruction methods, such as the algebraic reconstruction technique (ART) and the related simultaneous ART (SART), reconstruct a two-dimensional (2-D) or three-dimensional (3-D) object from its X-ray projections. The algebraic methods have, in certain scenarios, many advantages over the more popular Filtered Backprojection approaches and have also recently been shown to perform well for 3-D cone-beam reconstruction. However, so far the slow speed of these iterative methods have prohibited their routine use in clinical applications. Here, the authors address this shortcoming and investigate the utility of widely available 2-D texture mapping graphics hardware for the purpose of accelerating the 3-D algebraic reconstruction. They find that this hardware allows 3-D cone-beam reconstructions to be obtained at almost interactive speeds, with speed-ups of over 50 with respect to implementations that only use general-purpose CPUs. However the authors also find that the reconstruction quality is rather sensitive to the resolution of the framebuffer, and to address this critical issue they propose a scheme that extends the precision of a given framebuffer by 4 bits, using the color channels. With this extension, a 12-bit framebuffer delivers useful reconstructions for 0.5% tissue contrast, while an 8-bit framebuffer requires 4%. Since graphics hardware generates an entire image for each volume projection, it is most appropriately used with an algebraic reconstruction method that performs volume correction at that granularity as well, such as SART or SIRT. The authors chose SART for its faster convergence properties.

[1]  G. Herman,et al.  Algebraic reconstruction techniques (ART) for three-dimensional electron microscopy and x-ray photography. , 1970, Journal of theoretical biology.

[2]  P. Gilbert Iterative methods for the three-dimensional reconstruction of an object from projections. , 1972, Journal of theoretical biology.

[3]  B. F. Logan,et al.  The Fourier reconstruction of a head section , 1974 .

[4]  P. Joseph An Improved Algorithm for Reprojecting Rays through Pixel Images , 1983, IEEE Transactions on Medical Imaging.

[5]  A. Kak,et al.  Simultaneous Algebraic Reconstruction Technique (SART): A Superior Implementation of the Art Algorithm , 1984, Ultrasonic imaging.

[6]  A. H. Andersen Algebraic reconstruction in CT from limited views. , 1989, IEEE transactions on medical imaging.

[7]  Bruce D. Smith Cone-beam tomography: recent advances and a tutorial review , 1990 .

[8]  The accumulation buffer: hardware support for high-quality rendering , 1990, SIGGRAPH.

[9]  Lee Westover,et al.  Footprint evaluation for volume rendering , 1990, SIGGRAPH.

[10]  Philippe Rizo,et al.  Comparison of two three-dimensional x-ray cone-beam-reconstruction algorithms with circular source trajectories , 1991 .

[11]  Mark Segal,et al.  Fast shadows and lighting effects using texture mapping , 1992, SIGGRAPH.

[12]  R. Lewitt Alternatives to voxels for image representation in iterative reconstruction algorithms , 1992, Physics in medicine and biology.

[13]  J O Deasy,et al.  Tomotherapy: a new concept for the delivery of dynamic conformal radiotherapy. , 1993, Medical physics.

[14]  T. K. Narayan,et al.  Evaluation of task-oriented performance of several fully 3D PET reconstruction algorithms. , 1994, Physics in medicine and biology.

[15]  Brian Cabral,et al.  Accelerated volume rendering and tomographic reconstruction using texture mapping hardware , 1994, VVS '94.

[16]  M. Carter Computer graphics: Principles and practice , 1997 .

[17]  Thomas Ertl,et al.  Computer Graphics - Principles and Practice, 3rd Edition , 2014 .

[18]  M. Meißner,et al.  Vizard II, a PCI-card for real-time volume rendering , 1998, SIGGRAPH 1998.

[19]  Klaus Mueller,et al.  Anti-Aliased 3D Cone-Beam Reconstruction of Low-Contrast Objects with Algebraic Methods , 1999, IEEE Trans. Medical Imaging.

[20]  Klaus Mueller,et al.  Fast Implementation of Algebraic Methods for 3D Reconstruction from Cone-Beam Data , 1999, IEEE Trans. Medical Imaging.

[21]  Klaus Mueller,et al.  Use of graphics hardware to accelerate algebraic reconstruction methods , 1999, Medical Imaging.

[22]  K. Mueller,et al.  Cone-beam computed tomography (CT) for a megavoltage linear accelerator (LINAC) using an electronic portal imaging device (EPID) and the algebraic reconstruction technique (ART) , 2000, Proceedings of the 22nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society (Cat. No.00CH37143).

[23]  M. Glas,et al.  Principles of Computerized Tomographic Imaging , 2000 .