Investigation of oesophageal photoplethysmographic signals and blood oxygen saturation measurements in cardiothoracic surgery patients.

Pulse oximeter probes attached to the finger may fail to estimate blood oxygen saturation (SpO2) in patients with compromised peripheral perfusion (e.g. hypothermic cardiopulmonary bypass surgery). The measurement of SpO2 from a central organ such as the oesophagus is suggested as an alternative to overcome this problem. A reflectance oesophageal pulse oximeter probe and a processing system implemented in LabVIEW were developed. The system was evaluated in clinical measurements on 50 cardiothoracic surgery patients. Oesophageal photoplethysmographic (PPG) signals with large amplitudes and high signal-to-noise ratios were measured from various depths within the oesophagus from all the cardiothoracic patients. The oesophageal PPG amplitudes from these patients were in good agreement with previous oesophageal PPG amplitude measurements from healthy anaesthetized patients. The oesophageal pulse oximeter SpO2 results agreed well with the estimated arterial oxygen saturation (SaO2) values inferred from the oxygen tension obtained by blood gas analysis. The mean (+/- SD) of the differences between the oesophageal pulse oximeter SpO2 readings and those from blood gas analysis was 0.02 +/- 0.88%. Also, the oesophageal pulse oximeter was found to be reliable and accurate in five cases of poor peripheral perfusion when a commercial finger pulse oximeter probe failed to estimate oxygen saturation values for at least 10 min. These results suggest that the arterial blood circulation to the oesophagus is less subject to vasoconstriction and decreased PPG amplitudes than are the peripheral sites used for pulse oximetry such as the finger. It is concluded that oesophageal SPO2 monitoring may be of clinical value.

[1]  J Rosenberg,et al.  Nasal pulse oximetry overestimates oxygen saturation , 1990, Anaesthesia.

[2]  Douglas G. Altman,et al.  Measurement in Medicine: The Analysis of Method Comparison Studies , 1983 .

[3]  Y. Mendelson,et al.  Noninvasive pulse oximetry utilizing skin reflectance photoplethysmography , 1988, IEEE Transactions on Biomedical Engineering.

[4]  P A Kyriacou,et al.  Investigation of the human oesophagus as a new monitoring site for blood oxygen saturation. , 2001, Physiological measurement.

[5]  J L Higgins,et al.  Photoplethysmographic evaluation of the relationship between skin reflectance and skin blood volume. , 1986, Journal of biomedical engineering.

[6]  P A Oberg,et al.  Photoplethysmography. Part 2. Influence of light source wavelength. , 1991, Medical & biological engineering & computing.

[7]  V. C. Roberts,et al.  Photoplethysmography- fundamental aspects of the optical properties of blood in motion , 1982 .

[8]  J. C. Dorlas,et al.  Photo-electric plethysmography as a monitoring device in anaesthesia. Application and interpretation. , 1985, British journal of anaesthesia.

[9]  R. K. Webb,et al.  A comparison of the performance of 20 pulse oximeters under conditions of poor perfusion , 1991, Anaesthesia.

[10]  D. Reich,et al.  Predictors of Pulse Oximetry Data Failure , 1996, Anesthesiology.

[11]  H F Mahieu,et al.  Photoelectric plethysmography--some fundamental aspects of the reflection and transmission method. , 1981, Clinical physics and physiological measurement : an official journal of the Hospital Physicists' Association, Deutsche Gesellschaft fur Medizinische Physik and the European Federation of Organisations for Medical Physics.