PURPOSE
An electronic portal imaging device (EPID) was recently installed on our dual-energy linear accelerator. Commissioning and quality assurance techniques were developed for the EPID.
METHODS AND MATERIALS
A commissioning procedure was developed consisting of five parts: (a) physical operation and safety; (b) image acquisition, resolution, and sensitivity calibration; (c) image storage, analysis, and handling; (d) reference image acquisition; and (e) clinical operations.
RESULTS
The physical operation and safety tests relate to the motions of the unit, stability of the unit supports, safety interlocks, and interlock overrides. Imager contrast and spatial resolutions are monitored by imaging a contrast-detail phantom. The imager calibration procedure consists of a no-radiation image to compensate for signal offsets, as well as a "flat-field image." The flat-field image is taken with 5.0 cm of homogeneous phantom material placed at isocenter to provide some photon scatter and to approximate the presence of a patient. Daily quality assurance procedures consists of safety tests and the acquisition and inspection of images of the contrast-detail phantom. After 1 year, the frequency of the daily procedure was reduced to weekly. Quarterly QA procedures are conducted by the physicist and consist of the same procedures conducted in the weekly test. The annual QA procedure consists of a duplication of the commissioning procedure.
CONCLUSION
The procedures discussed in this article were applied to an ionization-chamber device. They have been useful in identifying difficulties with the EPID operation, including the need for recalibrating and monitoring the accelerator output stability.
[1]
L Dong,et al.
An objective method for evaluating electronic portal imaging devices.
,
1994,
Medical physics.
[2]
P. Munro,et al.
A review of electronic portal imaging devices (EPIDs).
,
1992,
Medical physics.
[3]
M. van Herk,et al.
Physical aspects of a liquid-filled ionization chamber with pulsed polarizing voltage.
,
1991
.
[4]
J. W. Epstein,et al.
On-line radiotherapy imaging with an array of fiber-optic image reducers.
,
1990,
International journal of radiation oncology, biology, physics.
[5]
M van Herk,et al.
First clinical experience with a newly developed electronic portal imaging device.
,
1990,
International journal of radiation oncology, biology, physics.
[6]
A Fenster,et al.
Therapy imaging: a signal-to-noise analysis of a fluoroscopic imaging system for radiotherapy localization.
,
1990,
Medical physics.
[7]
Michael G. Herman,et al.
CLINICAL USE OF ON-LINE PORTAL IMAGING FOR DAILY PATIENT TREATMENT VERIFICATION
,
1994
.
[8]
J. W. Epstein,et al.
Development of a second-generation fiber-optic on-line image verification system.
,
1993,
International journal of radiation oncology, biology, physics.