A microcontroller‐based implantable neuromuscular stimulation system with wireless power and data transmission for animal experiments

Abstract The aim of this study is to develop an implantable microcontroller‐based stimulation system with external wireless controller for data and power transmission. The system coupling efficiency can reach 35% in a reasonable range of distance via a self‐oscillating class E transmitter with power transmission efficiency up to 90%. The implanted micro‐stimulator is built on a double‐layer printed circuit board, which is 3 cm in diameter, by using SMD components mostly. Its overall power consumption is around 30 mW. Our results indicate the regulated voltage can be maintained at a tractable range when the coupling distance is within 30 mm. Within this distance, lateral displacement of the implant is allowed as long as it is inside the area of the transmitting coil. Meanwhile, a wide range of stimulation patterns can be generated for nerve stimulation and blocking for animal experiments. All the functional blocks of a prototype implantable neuromuscular system have been successfully designed, implanted, and tested which is valuable for future ASIC design for a miniature implantable microstimulator.

[1]  A. Barriskill,et al.  Development of functional electrical stimulators utilizing cochlear implant technology. , 2001, Medical engineering & physics.

[2]  K. Najafi,et al.  A single-channel implantable microstimulator for functional neuromuscular stimulation , 1997, IEEE Transactions on Biomedical Engineering.

[3]  D J Hitchings,et al.  ASIC or PIC? Implantable stimulators based on semi-custom CMOS technology or low-power microcontroller architecture. , 2001, Medical engineering & physics.

[4]  Harvey W. Ko,et al.  In-vivo measurement of brain edema with the magnetic bio-impedance method , 1996, Proceedings of 18th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[5]  M W Johnson,et al.  Implantable transducer for two-degree of freedom joint angle sensing. , 1994, IEEE transactions on rehabilitation engineering : a publication of the IEEE Engineering in Medicine and Biology Society.

[6]  J J Chen,et al.  The validity of stimulus-evoked EMG for studying muscle fatigue characteristics of paraplegic subjects during dynamic cycling movement. , 1997, IEEE transactions on rehabilitation engineering : a publication of the IEEE Engineering in Medicine and Biology Society.

[7]  W.J. Heetderks,et al.  RF powering of millimeter- and submillimeter-sized neural prosthetic implants , 1988, IEEE Transactions on Biomedical Engineering.

[8]  M. Elhilali,et al.  Implantable selective stimulator to improve bladder voiding: design and chronic experiments in dogs. , 2000, IEEE transactions on rehabilitation engineering : a publication of the IEEE Engineering in Medicine and Biology Society.

[9]  C.M. Zierhofer,et al.  High-efficiency coupling-insensitive transcutaneous power and data transmission via an inductive link , 1990, IEEE Transactions on Biomedical Engineering.

[10]  P Atanasov,et al.  Short-term canine implantation of a glucose monitoring-telemetry device. , 1996, Medical engineering & physics.

[11]  Mohamad Sawan,et al.  Programmable high-amplitude balanced stimulus current-source for implantable microstimulators , 1997, Proceedings of the 19th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. 'Magnificent Milestones and Emerging Opportunities in Medical Engineering' (Cat. No.97CH36136).

[12]  W. Liu,et al.  A neuro-stimulus chip with telemetry unit for retinal prosthetic device , 2000, IEEE Journal of Solid-State Circuits.

[13]  G E Loeb,et al.  BION system for distributed neural prosthetic interfaces. , 2001, Medical engineering & physics.

[14]  P. Hunter Peckham,et al.  Multichannel Implantable Stimulator for Control of Paralyzed Muscle , 1981, IEEE Transactions on Biomedical Engineering.

[15]  K Arabi,et al.  Electronic design of a multichannel programmable implant for neuromuscular electrical stimulation. , 1999, IEEE transactions on rehabilitation engineering : a publication of the IEEE Engineering in Medicine and Biology Society.

[16]  Babak Ziaie,et al.  A self-oscillating detuning-insensitive class-E transmitter for implantable microsystems , 2001, IEEE Transactions on Biomedical Engineering.

[17]  N. Hoshimiya,et al.  Development of percutaneous intramuscular electrode for multichannel FES system , 1989, IEEE Transactions on Biomedical Engineering.

[18]  P H Peckham,et al.  Implanted stimulators for restoration of function in spinal cord injury. , 2001, Medical engineering & physics.

[19]  N. O. Sokal Class-E switching-mode high-efficiency tuned RF/microwave power amplifier: improved design equations , 2000, 2000 IEEE MTT-S International Microwave Symposium Digest (Cat. No.00CH37017).

[20]  C.M. Zierhofer,et al.  Geometric approach for coupling enhancement of magnetically coupled coils , 1996, IEEE Transactions on Biomedical Engineering.

[21]  P.R. Troyk,et al.  Closed-loop class E transcutaneous power and data link for MicroImplants , 1992, IEEE Transactions on Biomedical Engineering.