Cone beam volume CT image artifacts caused by defective cells in x-ray flat panel imagers and the artifact removal using a wavelet-analysis-based algorithm.

The application of x-ray flat panel imagers (FPIs) in cone beam volume CT (CBVCT) has attracted increasing attention. However, due to a deficient semiconductor array manufacturing process, defective cells unavoidably exist in x-ray FPIs. These defective cells cause their corresponding image pixels in a projection image to behave abnormally in signal gray level, and result in severe streak and ring artifacts in a CBVCT image reconstructed from the projection images. Since a three-dimensional (3-D) back-projection is involved in CBVCT, the formation of the streak and ring artifacts is different from that in the two-dimensional (2-D) fan beam CT. In this paper, a geometric analysis of the abnormality propagation in the 3D back-projection is presented, and the morphology of the streak and ring artifacts caused by the abnormality propagation is investigated through both computer simulation and phantom studies. In order to calibrate those artifacts, a 2D wavelet-analysis-based statistical approach to correct the abnormal pixels is proposed. The approach consists of three steps: (1) the location-invariant defective cells in an x-ray FPI are recognized by applying 2-D wavelet analysis on flat-field images, and a comprehensive defective cell template is acquired; (2) based upon the template, the abnormal signal gray level of the projection image pixels corresponding to the location-invariant defective cells is replaced with the interpolation of that of their normal neighbor pixels; (3) that corresponding to the isolated location-variant defective cells are corrected using a narrow-windowed median filter. The CBVCT images of a CT low-contrast phantom are employed to evaluate this proposed approach, showing that the streak and ring artifacts can be reliably eliminated. The novelty and merit of the approach are the incorporation of the wavelet analysis whose intrinsic multi-resolution analysis and localizability make the recognition algorithm robust under variable x-ray exposure levels between 30% and 70% of the dynamic range of an x-ray FPI.

[1]  J Yorkston,et al.  Empirical and theoretical investigation of the noise performance of indirect detection, active matrix flat-panel imagers (AMFPIs) for diagnostic radiology. , 1997, Medical physics.

[2]  Brian G. Rodricks,et al.  Filtered gain calibration and its effect on DQE and image quality in digital imaging systems , 2000, Medical Imaging.

[3]  Richard L. Weisfield,et al.  Improved page-size 127-um-pixel amorphous-silicon image sensor for x-ray diagnostic medical imaging applications , 1997, Medical Imaging.

[4]  J H Siewerdsen,et al.  Cone-beam computed tomography with a flat-panel imager: initial performance characterization. , 2000, Medical physics.

[5]  Isaias D. Job,et al.  Flat-panel imaging system for fluoroscopy applications , 1998, Medical Imaging.

[6]  John A. Rowlands,et al.  Detective quantum efficiency of direct flat-panel x-ray imaging detectors for fluoroscopy , 1998, Medical Imaging.

[7]  D. Jaffray,et al.  A ghost story: spatio-temporal response characteristics of an indirect-detection flat-panel imager. , 1999, Medical physics.

[8]  Norbert Conrads,et al.  Dynamic x-ray imaging system based on an amorphous silicon thin-film array , 1998, Medical Imaging.

[9]  G. Kowalski,et al.  The Influence of Fixed Errors of a Detector Array on the Reconstruction of Objects from Their Projections , 1977, IEEE Transactions on Nuclear Science.

[10]  Wei Zhao,et al.  Flat panel detector for digital radiology using active matrix readout of amorphous selenium , 1997, Medical Imaging.

[11]  J Yorkston,et al.  Signal, noise power spectrum, and detective quantum efficiency of indirect-detection flat-panel imagers for diagnostic radiology. , 1998, Medical physics.

[12]  Stéphane Mallat,et al.  Singularity detection and processing with wavelets , 1992, IEEE Trans. Inf. Theory.

[13]  John M. Boone,et al.  Flat-field correction technique for digital detectors , 1998, Medical Imaging.

[14]  Jean-Pierre Moy,et al.  How does real offset and gain correction affect the DQE in images from x-ray flat detectors? , 1999, Medical Imaging.

[15]  J H Siewerdsen,et al.  Cone-beam computed tomography with a flat-panel imager: effects of image lag. , 1999, Medical physics.

[16]  Mark A. Hartney,et al.  New amorphous-silicon image sensor for x-ray diagnostic medical imaging applications , 1998, Medical Imaging.

[17]  Tsuneo Saito,et al.  Derivation and implementation of a cone-beam reconstruction algorithm for nonplanar orbits , 1994, IEEE Trans. Medical Imaging.

[18]  Jeffrey H. Siewerdsen,et al.  Cone-beam CT with a flat-panel imager: noise considerations for fully 3D computed tomography , 2000, Medical Imaging.

[19]  Ruola Ning,et al.  Accurate and efficient calibration method for a selenium flat-panel-detector-based volume tomographic angiography imaging system , 1999, Medical Imaging.

[20]  Ruola Ning,et al.  Flat panel detector-based cone-beam volume CT angiography imaging: system evaluation , 2000, IEEE Transactions on Medical Imaging.

[21]  G. Kowalski Suppression of Ring Artefacts in CT Fan-Beam Scanners , 1978 .

[22]  L. Feldkamp,et al.  Practical cone-beam algorithm , 1984 .

[23]  Ruola Ning,et al.  2D wavelet-analysis-based calibration technique for flat-panel imaging detectors: application in cone beam volume CT , 1999, Medical Imaging.

[24]  Maxwell J. Allen,et al.  Characterization of an amorphous-silicon fluoroscopic imager , 1997, Medical Imaging.

[25]  Ruola Ning,et al.  Selenium flat-panel detector-based volume tomographic angiography imaging: phantom studies , 1998, Medical Imaging.

[26]  J Yorkston,et al.  Empirical investigation of the signal performance of a high-resolution, indirect detection, active matrix flat-panel imager (AMFPI) for fluoroscopic and radiographic operation. , 1997, Medical physics.

[27]  Norbert Jung,et al.  Technical and clinical results of an experimental flat dynamic (digital) x-ray image detector (FDXD) system with real-time corrections , 1998, Medical Imaging.

[28]  Manabu Tanaka,et al.  Development of a selenium-based flat-panel detector for real-time radiography and fluoroscopy , 1998, Medical Imaging.

[29]  Rolf Clackdoyle,et al.  Direct reconstruction of cone-beam data acquired with a vertex path containing a circle , 1998, IEEE Trans. Image Process..

[30]  F Noo,et al.  The dual-ellipse cross vertex path for exact reconstruction of long objects in cone-beam tomography. , 1998, Physics in medicine and biology.

[31]  Rolf Clackdoyle,et al.  A cone-beam reconstruction algorithm using shift-variant filtering and cone-beam backprojection , 1994, IEEE Trans. Medical Imaging.

[32]  Yong Yu,et al.  Flat-panel detector-based cone beam volume CT breast imaging: detector evaluation , 2003, SPIE Medical Imaging.

[33]  Robert G. Gould,et al.  Specification, acceptance testing and quality control of diagnostic x-ray imaging equipment , 1994 .

[34]  Brian G. Rodricks,et al.  Radiographic imaging characteristics of a direct conversion detector using selenium and thin film transistor array , 1997, Medical Imaging.