Precision of cerebral oxygenation and hemoglobin concentration measurements in neonates measured by near-infrared spectroscopy.

BACKGROUND AND AIM One source of error with near-infrared spectroscopy (NIRS) is the assumption that the measured tissue is optically homogeneous. This is not always the case. Our aim is to assess the impact of tissue homogeneity (TH) on the precision of NIRS measurements in neonates. METHODS On 36 term and 27 preterm neonates at least five 1-min measurements are performed on each subject using the OxiplexTS. The sensor position is slightly changed before each measurement while assessing TH. The precision for cerebral tissue oxygenation saturation (StO(2)) and total hemoglobin concentration (tHb) are calculated by repeated measures analysis of variance. RESULTS The mean StO(2) is not significantly different between term and preterm infants. The mean tHb is significantly lower in preterm infants (p < 0.01). With increasing TH, the precision of StO(2) increase from 5.6 to 4.6% for preterm and from 11.0 to 2.0% for term infants; the precision of tHb increases from 10.1 to 7.5μM for preterm and from 16.4 to 3.5 μM for term infants. The precision for StO(2) is higher in term than in preterm infants. The precision for tHb shows no significant difference between the two groups. CONCLUSIONS The precision of NIRS measurements correlates with tissue homogeneity.

[1]  Robert D. Retherford,et al.  Multiple Classification Analysis , 2011 .

[2]  A. Blasi,et al.  Illuminating the developing brain: The past, present and future of functional near infrared spectroscopy , 2010, Neuroscience & Biobehavioral Reviews.

[3]  Ali Alaraj,et al.  Near infrared technology in neuroscience: past, present and future , 2009, Neurological research.

[4]  R. Ohls,et al.  Current controversies in the management of the anemia of prematurity. , 2009, Seminars in perinatology.

[5]  M. Toet,et al.  Brain monitoring in neonates. , 2009, Early human development.

[6]  F. Northington,et al.  A systematic review of the role of intrapartum hypoxia-ischemia in the causation of neonatal encephalopathy. , 2008, American journal of obstetrics and gynecology.

[7]  Janet S. Soul,et al.  Cerebral Hemodynamic Changes During Intensive Care of Preterm Infants , 2008, Pediatrics.

[8]  Gunnar Naulaers,et al.  Monitoring Neonatal Regional Cerebral Oxygen Saturation in Clinical Practice: Value and Pitfalls , 2008, Neonatology.

[9]  N. Luban Management of anemia in the newborn. , 2008, Early human development.

[10]  Martin Wolf,et al.  Hemodynamic response to visual stimulation in newborn infants using functional near‐infrared spectroscopy , 2008, Human brain mapping.

[11]  Sanjay Aher,et al.  Neonatal anemia. , 2008, Seminars in fetal & neonatal medicine.

[12]  Martin Wolf,et al.  Progress of near-infrared spectroscopy and topography for brain and muscle clinical applications. , 2007, Journal of biomedical optics.

[13]  C. Hagmann,et al.  Outcome after intrapartum hypoxic ischaemia at term. , 2007, Seminars in fetal & neonatal medicine.

[14]  David A Boas,et al.  Assessment of Infant Brain Development With Frequency-Domain Near-Infrared Spectroscopy , 2007, Pediatric Research.

[15]  Gorm Greisen,et al.  Precision of measurement of cerebral tissue oxygenation index using near-infrared spectroscopy in preterm neonates. , 2006, Journal of biomedical optics.

[16]  D. Delpy,et al.  System for long-term measurement of cerebral blood and tissue oxygenation on newborn infants by near infra-red transillumination , 1988, Medical and Biological Engineering and Computing.

[17]  A. Raj,et al.  Assessment of Cerebral Tissue Oxygenation in Patients With Sickle Cell Disease: Effect of Transfusion Therapy , 2004, Journal of pediatric hematology/oncology.

[18]  Jan Menke,et al.  Reproducibility of Cerebral Near Infrared Spectroscopy in Neonates , 2003, Neonatology.

[19]  Sabine Van Huffel,et al.  Measurement of tissue oxygenation index during the first three days in premature born infants. , 2003, Advances in experimental medicine and biology.

[20]  V Dietz,et al.  The influence of a clear layer on near-infrared spectrophotometry measurements using a liquid neonatal head phantom. , 1999, Physics in medicine and biology.

[21]  G. Greisen,et al.  Near-infrared monitoring of cerebral tissue oxygen saturation and blood volume in newborn piglets. , 1997, The American journal of physiology.

[22]  P. Andrews,et al.  A comparison of the Invos 3100 and the Critikon 2020 near‐infrared spectrophotometers as monitors of cerebral oxygenation , 1997, Anaesthesia.

[23]  V Dietz,et al.  Measurement of absolute cerebral haemoglobin concentration in adults and neonates. , 1997, Advances in experimental medicine and biology.

[24]  E. Gratton,et al.  Frequency-domain multichannel optical detector for noninvasive tissue spectroscopy and oximetry , 1995 .

[25]  E Gratton,et al.  Propagation of photon-density waves in strongly scattering media containing an absorbing semi-infinite plane bounded by a straight edge. , 1993, Journal of the Optical Society of America. A, Optics and image science.