The Effect of Kinetic Properties on Statistical Variations of Ultrasound Signals Backscattered from Flowing Blood

The statistical distributions of ultrasonic signals backscattered from blood have recently been used to characterize hemodynamic properties, such as red blood cell (RBC) aggregation and blood coagulation. However, a thorough understanding of the relationship between blood properties and the statistical behavior of signals backscattered from flowing blood is still lacking. This prompted us to use the statistical parameter to characterize signals backscattered from both whole blood and RBC suspensions at different flow velocities (from 10 to 60 cm/s) and hematocrits (from 20% to 50%) under a steady laminar flow condition. The Nakagami parameter, scaling parameter, backscatter amplitude profile and flow velocity profile across a flow tube were acquired using a 10 MHz focused ultrasonic transducer. The backscattered signal peaked approximately at the centerline of the flow tube due to the effects of RBC aggregation, with the peak value increasing as the flow velocity of whole blood decreased. The Nakagami parameter increased from 0.45 to 0.78 as the flow velocity increased from 10 to 60 cm/s. The probability density function (PDF) of signals backscattered from flowing whole blood conformed with a pre-Rayleigh distribution. The Nakagami parameter was close to 1 for signals backscattered from RBC suspensions at all the flow velocities and hematocrits tested, for which the PDF was Rayleigh distributed. These differences in the statistical distributions of backscattered signals between whole blood and RBC suspensions suggest that variations in the size of dynamic scatterers in the flow affect the shape of the backscattered signal envelope, which should be considered in future statistical models used to characterize blood properties.

[1]  Shyh-Hau Wang,et al.  The effect of logarithmic compression on estimation of the Nakagami parameter for ultrasonic tissue characterization: a simulation study. , 2005, Physics in medicine and biology.

[2]  W. Nichols McDonald's Blood Flow in Arteries , 1996 .

[3]  K. Shung,et al.  Ultrasonic backscatter from flowing whole blood. II: Dependence on frequency and fibrinogen concentration. , 1988, The Journal of the Acoustical Society of America.

[4]  Shyh-Hau Wang,et al.  Detection of coagulating blood under steady flow by statistical analysis of backscattered signals , 2007, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[5]  J. S. Lee,et al.  A linear relation between the compressibility and density of blood. , 2001, The Journal of the Acoustical Society of America.

[6]  W. A. Verhoef,et al.  Texture of B-Mode Echograms: 3-D Simulations and Experiments of the Effects of Diffraction and Scatterer Density , 1985 .

[7]  K. Parker,et al.  Deviations from Rayleigh Statistics in Ultrasonic Speckle , 1988, Ultrasonic imaging.

[8]  K. Shung,et al.  On the Ultrasound Scattering from Blood as a Function of Hematocrit , 1982, IEEE Transactions on Sonics and Ultrasonics.

[9]  Shyh-Hau Wang,et al.  Characterization of Blood Properties from Coagulating Blood of Different Hematocrits Using Ultrasonic Backscatter and Attenuation , 2006 .

[10]  B. Goldberg,et al.  Classification of ultrasonic B-mode images of breast masses using Nakagami distribution , 2001, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[11]  D. S. Luciano,et al.  Human Physiology: The Mechanism of Body Function , 1975 .

[12]  K. Shung,et al.  An approach for measuring ultrasonic backscattering from biological tissues with focused transducers , 1997, IEEE Transactions on Biomedical Engineering.

[13]  G. Cloutier,et al.  The effects of hematocrit, shear rate, and turbulence on ultrasonic Doppler spectrum from blood , 1992, IEEE Transactions on Biomedical Engineering.

[14]  P. Shankar A general statistical model for ultrasonic backscattering from tissues , 2000 .

[15]  Shyh-Hau Wang,et al.  In Vitro Study on Assessment of Blood Coagulation and Clot Formation Using Doppler Ultrasound , 2005 .

[16]  K.K. Shung,et al.  An in vitro study of the effects of Doppler angle, fibrinogen, and hematocrit on ultrasonic Doppler power , 1999, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[17]  F. Foster,et al.  Non-Gaussian statistics and temporal variations of the ultrasound signal backscattered by blood at frequencies between 10 and 58 MHz. , 2004, Journal of the Acoustical Society of America.

[18]  J. Tarbell,et al.  Effect of flow disturbance on ultrasonic backscatter from blood. , 1984, The Journal of the Acoustical Society of America.

[19]  K. Kirk Shung,et al.  Ultrasound: An Unexplored Tool for Blood Flow Visualization and Hemodynamic Measurements , 2003 .

[20]  F. Dunn,et al.  Ultrasonic Scattering in Biological Tissues , 1992 .

[21]  G Cloutier,et al.  Ultrasound backscattering from non-aggregating and aggregating erythrocytes--a review. , 1997, Biorheology.

[22]  G. Cloutier,et al.  Power Doppler ultrasound evaluation of the shear rate and shear stress dependences of red blood cell aggregation , 1996, IEEE Transactions on Biomedical Engineering.

[23]  K. Shung,et al.  The "black hole" phenomenon in ultrasonic backscattering measurement under pulsatile flow with porcine whole blood in a rigid tube. , 2001, Biorheology.

[24]  K. Shung,et al.  Ultrasonic backscatter from flowing whole blood. I: Dependence on shear rate and hematocrit. , 1988, The Journal of the Acoustical Society of America.

[25]  P. Tsui,et al.  The effect of transducer characteristics on the estimation of Nakagami paramater as a function of scatterer concentration. , 2004, Ultrasound in medicine & biology.

[26]  T. Wilson,et al.  Intervening attenuation affects first-order statistical properties of ultrasound echo signals , 1999, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[27]  K.K. Shung,et al.  In vivo measurements of ultrasonic backscattering in blood , 2001, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[28]  P. Tsui,et al.  Detection of blood coagulation and clot formation using quantitative ultrasonic parameters. , 2005, Ultrasound in medicine & biology.

[29]  P M Shankar Comments on 'The effect of logarithmic compression on the estimation of the Nakagami parameter for ultrasonic tissue characterization'. , 2006, Physics in medicine and biology.

[30]  John M. Reid,et al.  Scattering of Ultrasound by Blood , 1976, IEEE Transactions on Biomedical Engineering.

[31]  K. Shung,et al.  Ultrasonic absorption in deer blood , 1981 .