A helical PTFE arteriovenous access graft to swirl flow across the distal anastomosis: results of a preliminary clinical study.

Intimal hyperplasia develops preferentially in regions where the blood flow is stagnant and wall shear stress low. The small amplitude helical geometry of the SwirlGraft was designed to ensure physiological-type swirling flow, and thus suppress the triggers. We report the first conceptual testing of the SwirlGraft. Primary, assisted primary and secondary patency rates at 6 months in 20 patients were 57.9+/-11.4%, 84.4+/-8.3% and 100+/-0.0%. There was angiographic evidence of reduction of helical geometry in a proportion of the grafts. The helical graft is associated with high assisted primary and secondary patency. Elaboration of the surgical implantation techniques and an improved SwirlGraft design can be expected to exploit the advantages of the helical concept.

[1]  M Ojha,et al.  Spatial and temporal variations of wall shear stress within an end-to-side arterial anastomosis model. , 1993, Journal of biomechanics.

[2]  S. Sherwin,et al.  The influence of out-of-plane geometry on the flow within a distal end-to-side anastomosis. , 2000, Journal of biomechanical engineering.

[3]  S Glagov,et al.  Anastomotic intimal hyperplasia: mechanical injury or flow induced. , 1992, Journal of vascular surgery.

[4]  Third Nkf-K Doqi III. NKF-K/DOQI Clinical Practice Guidelines for Vascular Access: update 2000. , 2001, American journal of kidney diseases : the official journal of the National Kidney Foundation.

[5]  T. Pilgram,et al.  Dialysis access grafts: anatomic location of venous stenosis and results of angioplasty. , 1995, Radiology.

[6]  P. Davies,et al.  Flow-mediated endothelial mechanotransduction. , 1995, Physiological reviews.

[7]  N Levin,et al.  NKF-K/DOQI Clinical Practice Guidelines: Update 2000. Foreword. , 2001, American journal of kidney diseases : the official journal of the National Kidney Foundation.

[8]  N. Cheshire,et al.  Preliminary comparative study of small amplitude helical and conventional ePTFE arteriovenous shunts in pigs , 2005, Journal of The Royal Society Interface.

[9]  K. T. Scott,et al.  Non-planar curvature and branching of arteries and non-planar-type flow , 1996, Proceedings of the Royal Society of London. Series A: Mathematical, Physical and Engineering Sciences.

[10]  Colin G. Caro,et al.  Implications of 3D vascular geometry , 2006 .

[11]  Randy L Carter,et al.  Patency of autogenous and polytetrafluoroethylene upper extremity arteriovenous hemodialysis accesses: a systematic review. , 2003, Journal of vascular surgery.

[12]  R. Schroter,et al.  Atheroma and arterial wall shear - Observation, correlation and proposal of a shear dependent mass transfer mechanism for atherogenesis , 1971, Proceedings of the Royal Society of London. Series B. Biological Sciences.