Physical performance of a positron tomograph for brain imaging with retractable septa.

Performance characteristics of a new design of positron tomograph with automatically retractable septa for brain imaging have been studied. The device, consisting of block BGO detectors (8 x 8 elements per block), has a ring diameter of 76 cm and an axial FOV of 106.5 mm. The in-plane resolution is on average 5.8 mm and 5.0 mm (FWHM) for stationary and wobble sampling, respectively, over the central 18 cm of the transaxial FOV. Its unique feature is the capability of data acquisition both in the 'conventional' 2D mode (with septa) or 3D mode (septa retracted) where coincidences between any of the 16 detector rings are acquired. When scattered events are subtracted, the efficiency for a 20 cm diameter uniform cylinder increases overall by a factor of 4.8 between 2D (septa extended) and 3D modes. For a 20 cm phantom the trues/singles ratio is higher for 3D than for 2D but for a given unscattered trues rate, the randoms rate in 3D is higher. At 380 keV the scatter fraction within a 20 cm cylinder is 10% (septa extended) and 36% (retracted). In spite of the increase in scatter when septa are retracted, the increased efficiency in the 3D mode of acquisition yields distinct advantages, particularly in the many studies where tracer concentration is low and consequently where dead time and random rates are less important.

[1]  S. E. Derenzo,et al.  The Electronics for the Donner 600-Crystal Positron Tomograph , 1986, IEEE Transactions on Nuclear Science.

[2]  R. Nutt,et al.  A Multicrystal Two Dimensional BGO Detector System for Positron Emission Tomography , 1986, IEEE Transactions on Nuclear Science.

[3]  W. F. Jones,et al.  A Real Time Digital Coincidence Processor for Positron Emission Tomography , 1986, IEEE Transactions on Nuclear Science.

[4]  J. C. Moyers A High Performance Detector Electronics System for Positron Tomography , 1990 .

[5]  Joel S. Karp,et al.  Special Papers Section-NPSS/Medical Imaging Conference Cross-Plane Scattering Correction-Point Source Deconvolution in PET , 1991 .

[6]  Spinks Tj Experience with the EEC phantom and test procedures in two block detector scanners. , 1991 .

[7]  M. Gilardi,et al.  Physical performance of the latest generation of commercial positron scanner , 1988 .

[8]  Paul E. Kinahan,et al.  Three-dimensional image reconstruction in object space , 1988 .

[9]  S R Cherry,et al.  3D PET using a conventional multislice tomograph without septa. , 1991, Journal of computer assisted tomography.

[10]  W. W. Moses,et al.  A positron tomograph with 600 BGO crystals and 2.6 mm resolution , 1988 .

[11]  J. Colsher,et al.  Fully-three-dimensional positron emission tomography , 1980, Physics in medicine and biology.

[12]  L. G. Byars,et al.  A VMEBUS based, real time sorter design for positron emission tomography , 1986, IEEE Transactions on Nuclear Science.