Enanced video heart rate and respiratory rate evaluation: standard multiparameter monitor vs clinical confrontation in newborn patients

Vital signs in neonatal care are usually measured by a multiparameter monitor (MM). However, this kind of device is inaccurate on the assessment in newborns, which can present different respiratory patterns and superficial breathing, resulting hardly detectable by the MM. In some cases, the assessment of the physician is more effective on the evaluation of the patient. In this paper we present a non-contact solution to measure the respiration rate (RR) and heart rate (HR) using a commercially available web-camera (WeC) and a personal computer (PC). We use Eulerian Video Magnification (EVM) to amplify the color variations in the image sequence. By extracting the signal from the thorax portion of the patients we are able to measure the RR and HR by spectral analysis. The results are compared with the MM and the assessment of the physician. The measures of RR and HR correlate with the data from the MM and result even more accurate than the MM when compared with the physician’s evaluation. We collect data on 40 patients demonstrating the feasibility of this method. The measure of RR and HR shows a root mean square error of 6.8 bpm for the HR and 2.1 bpm for the RR.

[1]  H. C. Miller,et al.  The changing respiratory patterns in newborn infants. , 1953, A.M.A. American journal of diseases of children.

[2]  Laila Habiballah,et al.  Prevalence of neonate adhesive skin injuries in a Jordanian intensive care unit. , 2017, Nursing children and young people.

[3]  Lorenzo Scalise,et al.  Heart rate measurement in neonatal patients using a webcamera , 2012, 2012 IEEE International Symposium on Medical Measurements and Applications Proceedings.

[4]  Douglas G. Altman,et al.  Measurement in Medicine: The Analysis of Method Comparison Studies , 1983 .

[5]  B. Roth,et al.  Factors Influencing Apnea and Bradycardia of Prematurity – Implications for Neurodevelopment , 2006, Neonatology.

[6]  Kayvan Najarian,et al.  Motion Artifact Suppression in Impedance Pneumography Signal for Portable Monitoring of Respiration: An Adaptive Approach , 2017, IEEE Journal of Biomedical and Health Informatics.

[7]  Frédo Durand,et al.  Eulerian video magnification for revealing subtle changes in the world , 2012, ACM Trans. Graph..

[8]  Enrico Primo Tomasini,et al.  Evaluation of Heart Rate Variability by means of Laser Doppler Vibrometry measurements , 2015 .

[9]  D J Henderson-Smart,et al.  The effect of gestational age on the incidence and duration of recurrent apnoea in newborn babies , 1981, Australian paediatric journal.

[10]  Lorenzo Scalise,et al.  Measurement of respiration rate in preterm infants by laser Doppler vibrometry , 2011, 2011 IEEE International Symposium on Medical Measurements and Applications.

[11]  D. Altman,et al.  STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT , 1986, The Lancet.

[12]  E. P. Tomasini,et al.  The non-contact measure of the heart rate variability by laser Doppler vibrometry: comparison with electrocardiography , 2016 .

[13]  W. Walas,et al.  Iatrogenic skin injury in the neonatal intensive care unit , 2011, The journal of maternal-fetal & neonatal medicine : the official journal of the European Association of Perinatal Medicine, the Federation of Asia and Oceania Perinatal Societies, the International Society of Perinatal Obstetricians.

[14]  J M Bland,et al.  Statistical methods for assessing agreement between two methods of clinical measurement , 1986 .

[15]  U. Morbiducci,et al.  Non-contact cardiac monitoring from carotid artery using optical vibrocardiography. , 2008, Medical engineering & physics.

[16]  K. Barrington,et al.  Apnea Is Associated with Neurodevelopmental Impairment in Very Low Birth Weight Infants , 2004, Journal of Perinatology.