Technical Note: Extended field‐of‐view (FOV) MRI distortion determination through multi‐positional phantom imaging

Abstract Comprehensive characterization of geometric distortions for MRI simulators and MRI‐guided treatment delivery systems is typically performed with large phantoms that are costly and unwieldy to handle. Here we propose an easily implementable methodology for MR distortion determination of the entire imaging space of the scanner through the use of a compact commercially available distortion phantom. The MagphanRT phantom was scanned at several locations within a MR scanner. From each scan, an approximate location of the phantom was determined from a subset of the fiducial spheres. The fiducial displacements were determined, and a displacement field was fitted to the displacement data using the entire multi‐scan data set. An orthogonal polynomial expansion fitting function was used that had been augmented to include independent rigid‐body transformations for each scan. The rigid‐body portions of the displacement field were thereafter discarded, and the resultant fit then represented the distortion field. Multi‐positional scans of the phantom were used successfully to determine the distortion field with extended coverage. A single scan of the phantom covered 20 cm in its smallest dimension. By stitching together overlapping scans we extended the distortion measurements to 30 cm. No information about the absolute location or orientation of each scan was required. The method, termed the Multi‐Scan Expansion (MSE) method, can be easily applied for larger field‐of‐views (FOVs) by using a combination of larger phantom displacements and more scans. The implementation of the MSE method allows for distortion determination beyond the physical limitations of the phantom. The method is scalable to the user’s needs and does not require any specialized equipment. This approach could open up for easier determination of the distortion magnitude at distances further from the scanner’s isocenter. This is especially important in the newly proposed methodologies of MR‐only simulation in RT and in adaptive replanning in MR linac systems.

[1]  Rabih Hammoud,et al.  Characterization of 3D geometric distortion of magnetic resonance imaging scanners commissioned for radiation therapy planning. , 2016, Magnetic resonance imaging.

[2]  Pantelis Karaiskos,et al.  MRI-Related Geometric Distortions in Stereotactic Radiotherapy Treatment Planning: Evaluation and Dosimetric Impact , 2017, Technology in cancer research & treatment.

[3]  Dong-Hoon Lee,et al.  Characteristics of geometric distortion correction with increasing field-of-view in open-configuration MRI. , 2014, Magnetic resonance imaging.

[4]  Richard Bowtell,et al.  Novel gradient coils designed using a boundary element method , 2007 .

[5]  Slobodan Devic,et al.  MRI simulation for radiotherapy treatment planning. , 2012, Medical physics.

[6]  Eric S Paulson,et al.  Consensus opinion on MRI simulation for external beam radiation treatment planning. , 2016, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[7]  Paul J. Keall,et al.  MRI Linac Systems , 2019, MRI for Radiotherapy.

[8]  Rabih Hammoud,et al.  Development and validation of a novel large field of view phantom and a software module for the quality assurance of geometric distortion in magnetic resonance imaging. , 2015, Magnetic resonance imaging.

[9]  Ken-Pin Hwang,et al.  Spatial Precision in Magnetic Resonance Imaging-Guided Radiation Therapy: The Role of Geometric Distortion. , 2016, International journal of radiation oncology, biology, physics.

[10]  Lizette Warner,et al.  Clinical workflow for MR-only simulation and planning in prostate , 2017, Radiation oncology.

[11]  B. Zackrisson,et al.  Dedicated magnetic resonance imaging in the radiotherapy clinic. , 2009, International journal of radiation oncology, biology, physics.

[12]  Peter Metcalfe,et al.  Continuous table acquisition MRI for radiotherapy treatment planning: distortion assessment with a new extended 3D volumetric phantom. , 2015, Medical physics.

[13]  M. Barton,et al.  MRI-Linear Accelerator Radiotherapy Systems. , 2018, Clinical oncology (Royal College of Radiologists (Great Britain)).