Phosphene perceptions and safety of chronic visual cortex stimulation in a blind subject.

Stimulation of primary visual cortices has the potential to restore some degree of vision to blind individuals. Developing safe and reliable visual cortical prostheses requires assessment of the long-term stability, feasibility, and safety of generating stimulation-evoked perceptions.A NeuroPace responsive neurostimulation system was implanted in a blind individual with an 8-year history of bare light perception, and stimulation-evoked phosphenes were evaluated over 19 months (41 test sessions). Electrical stimulation was delivered via two four-contact subdural electrode strips implanted over the right medial occipital cortex. Current and charge thresholds for eliciting visual perception (phosphenes) were measured, as were the shape, size, location, and intensity of the phosphenes. Adverse events were also assessed.Stimulation of all contacts resulted in phosphene perception. Phosphenes appeared completely or partially in the left hemifield. Stimulation of the electrodes below the calcarine sulcus elicited phosphenes in the superior hemifield and vice versa. Changing the stimulation parameters of frequency, pulse width, and burst duration affected current thresholds for eliciting phosphenes, and increasing the amplitude or frequency of stimulation resulted in brighter perceptions. While stimulation thresholds decreased between an average of 5% and 12% after 19 months, spatial mapping of phosphenes remained consistent over time. Although no serious adverse events were observed, the subject experienced mild headaches and dizziness in three instances, symptoms that did not persist for more than a few hours and for which no clinical intervention was required.Using an off-the-shelf neurostimulator, the authors were able to reliably generate phosphenes in different areas of the visual field over 19 months with no serious adverse events, providing preliminary proof of feasibility and safety to proceed with visual epicortical prosthetic clinical trials. Moreover, they systematically explored the relationship between stimulation parameters and phosphene thresholds and discovered the direct relation of perception thresholds based on primary visual cortex (V1) neuronal population excitation thresholds.

[1]  J. Winawer,et al.  Linking Electrical Stimulation of Human Primary Visual Cortex, Size of Affected Cortical Area, Neuronal Responses, and Subjective Experience , 2016, Neuron.

[2]  W. Dobelle,et al.  Phosphenes produced by electrical stimulation of human occipital cortex, and their application to the development of a prosthesis for the blind , 1974, The Journal of physiology.

[3]  Daniel R. Merrill,et al.  Electrical stimulation of excitable tissue: design of efficacious and safe protocols , 2005, Journal of Neuroscience Methods.

[4]  Greg Worrell,et al.  Long-Term Measurement of Impedance in Chronically Implanted Depth and Subdural Electrodes During Responsive Neurostimulation in Humans , 2013, Brain Stimulation.

[5]  Ping Sun,et al.  Rules Governing Perception of Multiple Phosphenes by Human Observers , 2018, bioRxiv.

[6]  W. Bosking,et al.  Saturation in Phosphene Size with Increasing Current Levels Delivered to Human Visual Cortex , 2017, The Journal of Neuroscience.

[7]  G. Brindley,et al.  The sensations produced by electrical stimulation of the visual cortex , 1968, The Journal of physiology.

[8]  L. Geddes,et al.  The Strength-Duration Curve , 1985, IEEE Transactions on Biomedical Engineering.

[9]  R.V. Shannon,et al.  A model of safe levels for electrical stimulation , 1992, IEEE Transactions on Biomedical Engineering.

[10]  C. Kufta,et al.  Feasibility of a visual prosthesis for the blind based on intracortical microstimulation of the visual cortex. , 1996, Brain : a journal of neurology.

[11]  G. Loeb,et al.  Visual sensations produced by intracortical microstimulation of the human occipital cortex , 1990, Medical and Biological Engineering and Computing.

[12]  M. Mladejovsky,et al.  Artificial Vision for the Blind: Electrical Stimulation of Visual Cortex Offers Hope for a Functional Prosthesis , 1974, Science.

[13]  W. H. Dobelle,et al.  Artificial vision for the blind by electrical stimulation of the visual cortex. , 1979, Neurosurgery.

[14]  M. Mladejovsky,et al.  ‘Braille’ reading by a blind volunteer by visual cortex stimulation , 1976, Nature.

[15]  Daniel Yoshor,et al.  Receptive fields in human visual cortex mapped with surface electrodes. , 2007, Cerebral cortex.