A Comparison of Flow Field Structures of Two Tri-Leaflet Polymeric Heart Valves

Polymeric heart valves have the potential to reduce thrombogenic complications associated with current mechanical valves and overcome fatigue-related problems experienced by bioprosthetic valves. In this in vitro study, the velocity fields inside and downstream of two different prototype tri-lealfet polymeric heart valves were studied. Experiments were conducted on two 23 mm prototype polymeric valves, provided by AorTech Europe, having open or closed commissure designs and leaflet thickness of 120 and 80 μm, respectively. A two-dimensional LDV system was used to measure the velocity fields in the vicinity of the two valves under simulated physiological conditions. Both commissural design and leaflet thickness were found to affect the flow characteristics. In particular, very high levels of Reynolds shear stress of 13,000 dynes/cm2 were found in the leakage flow of the open commisure design. Maximum leakage velocities in the open and closed designs were 3.6 m/s and 0.5 m/s respectively; the peak forward flow velocities were 2.0 m/s and 2.6 m/s, respectively. In both valve designs, shear stress levels exceeding 4,000 dyne/cm2 were observed at the trailing edge of the leaflets and in the leakage and central orifice jets during peak systole. Additionally, regions of low velocity flow conducive to thrombus formation were observed in diastole. The flow structures measured in these experiments are consistent with the location of thrombus formation observed in preliminary animal experiments.

[1]  K. Chandran,et al.  In vitro comparison of velocity profiles and turbulent shear distal to polyurethane trileaflet and pericardial prosthetic valves. , 1989, Artificial organs.

[2]  N H Hwang,et al.  Human red blood cell hemolysis in a turbulent shear flow: contribution of Reynolds shear stresses. , 1984, Biorheology.

[3]  Bernd Klosterhalfen,et al.  Introduction of a Flexible Polymeric Heart Valve Prosthesis With Special Design for Mitral Position , 2003, Circulation.

[4]  D J Wheatley,et al.  In vitro function and durability of a polyurethane heart valve: material considerations. , 1996, The Journal of heart valve disease.

[5]  A P Yoganathan,et al.  A comparison of the hinge and near-hinge flow fields of the St Jude medical hemodynamic plus and regent bileaflet mechanical heart valves. , 2000, The Journal of thoracic and cardiovascular surgery.

[6]  P. Lu,et al.  A reevaluation and discussion on the threshold limit for hemolysis in a turbulent shear flow. , 2001, Journal of biomechanics.

[7]  Ajit P. Yoganathan,et al.  An automated method for analysis and visualization of laser doppler velocimetry data , 1997, Annals of Biomedical Engineering.

[8]  D J Wheatley,et al.  Artificial Heart and Cardiac Assist Devices Polyurethane heart valve durability : effects of leaflet thickness and material , 2018 .

[9]  S. Slack,et al.  The Effects of Flow on Blood Coagulation and Thrombosis , 1993, Thrombosis and Haemostasis.

[10]  S. Slack,et al.  Flow chambers and their standardization for use in studies of thrombosis. On behalf of the Subcommittee on Rheology of the Scientific and Standardization Committee of the ISTH. , 1994, Thrombosis and haemostasis.

[11]  D. Sandwell BIHARMONIC SPLINE INTERPOLATION OF GEOS-3 AND SEASAT ALTIMETER DATA , 1987 .

[12]  D J Wheatley,et al.  New polyurethane heart valve prosthesis: design, manufacture and evaluation. , 1996, Biomaterials.

[13]  D J Wheatley,et al.  Polyurethane: material for the next generation of heart valve prostheses? , 2000, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[14]  A P Yoganathan,et al.  In-vitro fluid dynamic characteristics of the abiomed trileaflet heart valve prosthesis. , 1983, Journal of biomechanical engineering.

[15]  Stuart L. Cooper,et al.  Polyurethanes In Medicine , 1986 .

[16]  Y. Nemerson,et al.  The Effect of Flow on Hemostasis and Thrombosis , 1991, Thrombosis and Haemostasis.

[17]  V T Turitto,et al.  Mechanical factors affecting hemostasis and thrombosis. , 1998, Thrombosis research.

[18]  R M Hochmuth,et al.  Shear-induced aggregation and lysis of platelets. , 1976, Transactions - American Society for Artificial Internal Organs.

[19]  Dan T Simionescu,et al.  Extracellular matrix degrading enzymes are active in porcine stentless aortic bioprosthetic heart valves. , 2003, Journal of biomedical materials research. Part A.

[20]  H Reul,et al.  Advances in design principle and fluid dynamics of a flexible polymeric heart valve. , 1991, ASAIO transactions.

[21]  S. Slack,et al.  Flow Chambers and their Standardization for Use in Studies of Thrombosis , 1994, Thrombosis and Haemostasis.

[22]  G. Gerosa,et al.  Isolation of intact aortic valve scaffolds for heart-valve bioprostheses: extracellular matrix structure, prevention from calcification, and cell repopulation features. , 2003, Journal of biomedical materials research. Part A.

[23]  Prosthesis-related complications: first-year annual rates. , 2002, The Journal of heart valve disease.

[24]  J Fisher,et al.  Hydrodynamic Function of a Biostable Polyurethane Flexible Heart Valve after Six Months in Sheep , 2001, The International journal of artificial organs.

[25]  H. T. Ellis,et al.  Lidar observations of a stratospheric dust cloud layer in the tropics , 1975 .

[26]  R. M. Privette,et al.  Cycle-to-cycle variation effects on turbulent shear stress measurements in pulsatile flows , 1988 .

[27]  G. M. Bernacca,et al.  Hydrodynamic function of polyurethane prosthetic heart valves: influences of Young's modulus and leaflet thickness. , 2002, Biomaterials.

[28]  Brandon R. Travis,et al.  An In Vitro Study of the Hinge and Near-Field Forward Flow Dynamics of the St. Jude Medical® Regent™ Bileaflet Mechanical Heart Valve , 2000, Annals of Biomedical Engineering.

[29]  A P Yoganathan,et al.  Velocity measurements and flow patterns within the hinge region of a Medtronic Parallel bileaflet mechanical valve with clear housing. , 1996, The Journal of heart valve disease.

[30]  L Zuckerman,et al.  Shear-induced activation of platelets. , 1979, Journal of biomechanics.

[31]  L. Jennings,et al.  Platelet size distribution measurements as indicators of shear stress-induced platelet aggregation , 1994, Annals of Biomedical Engineering.