Several studies have drawn attention to frequent user-error sources in self-monitoring of blood glucose (SMBG), for example, obtaining an inadequate blood volume, neglecting hand washing, applying blood incorrectly onto the test strip, or using expired test strips.1-3 However, there is no sound scientific evidence as to whether the time the patient requires to apply the blood onto the test strip affects measurement results. We investigated, with 3 different SMBG systems, if a time delay between the generation of a blood drop and the application onto the test strip affects glucose measurement results.
The investigation was performed by trained personnel at the Institut fur Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universitat Ulm, Ulm, Germany. The protocol was approved by the Ethics Committee, and the competent authority was notified.
Glucose measurements on capillary blood samples from fingertips of 89 subjects (49 male, 40 female; 26 with type 1 diabetes mellitus, 43 with type 2 diabetes mellitus, 20 without diabetes; mean age 59, ranging from 21 to 76 years) were evaluated. Measurements with each system were consecutively performed on 4 blood drops with a diameter of approximately 1 mm. The first and the fourth drops were measured immediately (0 s); the second and third drops remained on the skin surface for 10 and 25 s, respectively, before being measured. The drift between the first and the fourth measurement was checked to be ≤0.56 mmol/L (≤10%) for blood glucose concentrations ≤5.6 mmol/L (>5.6 mmol/L).
For each system, the relative deviations between the glucose value at 0 s (mean value of the first and the fourth measurement) and the glucose value after a delay of 10 and 25 s was calculated.
All 3 systems showed increased glucose values after a time delay between the sampling and the measurement procedure with mean relative deviations between 0.7% and 2.1% after 10 s and between 0.1% and 4.7% after 25 s when compared to immediate measurements (Table 1). An additional evaluation based on linear interpolation between the glucose concentrations at immediate measurement was used to exclude a possible influence of glucose instability effects (data not shown).
Table 1.
Relative Deviations Between the Glucose Value Measured Immediately After Sample Collection (Time Delay of 0 s) and the Glucose Value Measured After a Delay of 10 and 25 s
The amount of time that the blood drop is exposed to air can affect glucose measurements with some SMBG systems. The architecture and/or electrochemical composition of several test strips might also be prone to changes in the blood drop, for example, by evaporation, coagulation, or increase in pO2.4,5 Such changes in the blood drop over time might be particularly relevant in small drops with a higher surface-to-volume ratio (many systems require a sample volume less than 1 µL) or with long air-exposure times before the measurement, for example, when patients do not adequately prepare the measurement. Although further investigations are required to support the clinical relevance of our findings, patients should be advised to perform glucose measurements on capillary samples as quickly as possible.
[1]
G. Freckmann,et al.
Influence of Partial Pressure of Oxygen in Blood Samples on Measurement Performance in Glucose-Oxidase-Based Systems for Self-Monitoring of Blood Glucose
,
2013,
Journal of diabetes science and technology.
[2]
Ryuzo Kawamori,et al.
Glucose Monitoring After Fruit Peeling: Pseudohyperglycemia When Neglecting Hand Washing Before Fingertip Blood Sampling
,
2011,
Diabetes Care.
[3]
Dawn M. Bina,et al.
Identifying Variables Associated With Inaccurate Self-Monitoring of Blood Glucose: Proposed Guidelines to Improve Accuracy
,
2000,
The Diabetes educator.
[4]
G. Freckmann,et al.
Impact of partial pressure of oxygen in blood samples on the performance of systems for self-monitoring of blood glucose.
,
2014,
Diabetes technology & therapeutics.
[5]
G J Kost,et al.
Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing
,
2001,
Critical care medicine.
[6]
R. Bamberg,et al.
Effect of Adverse Storage Conditions on Performance of Glucometer Test Strips
,
2005,
American Society for Clinical Laboratory Science.