Hemodynamic evaluation of the CarboMedics prosthetic heart valve in the aortic position: comparison of noninvasive and invasive techniques.

Seventy-three patients with a CarboMedics aortic bileaflet valve prosthesis were examined by Doppler ultrasonography, and 27 of them were also assessed by transseptal catheterization. The ultrasonic mean systolic gradient was 17.1 +/- 5.6 mm Hg for valve size 19 mm, falling gradually with increasing valve size to 6.8 +/- 2.5 mm Hg for size 27 mm. The catheter mean systolic gradient was consistently smaller than the ultrasonic gradient (4.3 +/- 4.8 mm Hg), but Tobit regression analysis showed a significant association between the two methods. In all patients both methods revealed negligible to small amounts of retrograde leakage, which is assumed to be a normal finding for this valve. The effective flow areas of the valves calculated from the ultrasonic data were similar to the in vitro calculated flow areas. The hemodynamic potential of this valve is therefore completely utilized in vivo. The effective orifice area corrected for body surface area increased with increasing valve size, which demonstrates a moderate valve-patient mismatch.

[1]  A. Yoganathan,et al.  A 6 year experience with the St. Jude medical valve: hemodynamic performance, surgical results, biocompatibility and follow-up. , 1985, Journal of the American College of Cardiology.

[2]  R GORLIN,et al.  Hydraulic formula for calculation of the area of the stenotic mitral valve, other cardiac valves, and central circulatory shunts. I. , 1951, American heart journal.

[3]  R S Meltzer,et al.  Normal values of prosthetic valve Doppler echocardiographic parameters: a review. , 1988, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[4]  L S Wann,et al.  Doppler echocardiographic evaluation of Hancock and Björk-Shiley prosthetic values. , 1986, Journal of the American College of Cardiology.

[5]  E. C. Harrison,et al.  Flow characteristics of four commonly used mechanical heart valves. , 1986, The American journal of cardiology.

[6]  J. Dumesnil,et al.  Validation and applications of indexed aortic prosthetic valve areas calculated by Doppler echocardiography. , 1990, Journal of the American College of Cardiology.

[7]  S. Rahimtoola,et al.  The problem of valve prosthesis-patient mismatch. , 1978, Circulation.

[8]  H. Ihlen,et al.  Cardiac output measured by Doppler echocardiography in patients with aortic prosthetic valves. , 1990, European heart journal.

[9]  David W. Hosmer,et al.  Applied Logistic Regression , 1991 .

[10]  P. Stein,et al.  Color flow Doppler evaluation of St. Jude Medical prosthetic valves. , 1989, American Journal of Cardiology.

[11]  K. Hammermeister,et al.  Inadequacy of the Gorlin formula for predicting prosthetic valve area. , 1988, The American journal of cardiology.

[12]  H. Ihlen,et al.  Reproducibility of echocardiographic estimates of the area of stenosed aortic valves using the continuity equation. , 1990, International journal of cardiology.

[13]  T. Bowen,et al.  Hemodynamic evaluation of the St. Jude Medical valve prosthesis in the small aortic anulus. , 1981, The Journal of thoracic and cardiovascular surgery.

[14]  J Ross,et al.  Normal and abnormal prosthetic valve function as assessed by Doppler echocardiography. , 1986, Journal of the American College of Cardiology.

[15]  G Maurer,et al.  Discrepancies between Doppler and catheter gradients in aortic prosthetic valves in vitro. A manifestation of localized gradients and pressure recovery. , 1990, Circulation.

[16]  S Simonsen,et al.  Determination of cardiac output by Doppler echocardiography. , 1984, British heart journal.

[17]  S. Nitter‐Hauge,et al.  Aortic valve replacement. A randomized study comparing the Björk-Shiley and Lillehei-Kaster disc valves. Late haemodynamics related to clinical results. , 1979, Scandinavian journal of thoracic and cardiovascular surgery.

[18]  J B Seward,et al.  Continuous wave Doppler echocardiographic measurement of prosthetic valve gradients. A simultaneous Doppler-catheter correlative study. , 1989, Circulation.

[19]  L. Hatle,et al.  Doppler ultrasound in cardiology , 1985 .