Simultaneous intracranial EEG–fMRI in humans: Protocol considerations and data quality

We have recently performed simultaneous intracranial EEG and fMRI recordings (icEEG-fMRI) in patients with epilepsy. In this technical note, we examine limited thermometric data for potential electrode heating during our protocol and found that heating was ≤0.1 °C in-vitro at least 10 fold less than in-vivo limits. We quantify EEG quality, which can be degraded by MRI scanner-induced artefacts, and fMRI image (gradient echo echo-planar imaging: GE-EPI) signal quality around the electrodes, which can be degraded by electrode interactions with B1 (radiofrequency) and B0 (static) magnetic fields. We recorded EEG outside and within the MRI scanner with and without scanning. EEG quality was largely preserved during scanning and in particular heartbeat-related artefacts were small compared to epileptic events. To assess the GE-EPI signal reduction around the electrodes, we compared image signal intensity along paths into the brain normal to its surface originating from the individual platinum-iridium electrode contacts. GE-EPI images were obtained at 1.5 T with an echo time (TE) of 40 ms and repetition time (TR) of 3000 ms and a slice thickness of 2.5 mm. We found that GE-EPI signal intensity reduction was confined to a 10 mm radius and that it was reduced on average by less than 50% at 5mm from the electrode contacts. The GE-EPI image signal reduction also varied with electrode orientation relative to the MRI scanner axes; in particular, cortical grid contacts with a normal along the scanner's main magnetic field (B(0)) axis have higher artefact levels relative to those with a normal perpendicular to the z-axis. This suggests that the artefacts were predominantly susceptibility-related rather than due to B1 interactions. This information can be used to guide interpretation of results of icEEG-fMRI experiments proximal to the electrodes, and to optimise artefact reduction strategies.

[1]  Nikolaus Weiskopf,et al.  Optimal EPI parameters for reduction of susceptibility-induced BOLD sensitivity losses: A whole-brain analysis at 3 T and 1.5 T , 2006, NeuroImage.

[2]  M. Donahue,et al.  Artifacts in magnetic resonance imaging from metals , 1996 .

[3]  Robert Turner,et al.  The Effect of Slice Order and Thickness on fMRI Activation Data Using Multislice Echo-Planar Imaging , 1999, NeuroImage.

[4]  P. J. Hoopes,et al.  Basic principles of thermal dosimetry and thermal thresholds for tissue damage from hyperthermia , 2003, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[5]  J. Ebersole,et al.  Intracranial EEG Substrates of Scalp EEG Interictal Spikes , 2005, Epilepsia.

[6]  R. Deichmann,et al.  Compensation of Susceptibility-Induced BOLD Sensitivity Losses in Echo-Planar fMRI Imaging , 2001, NeuroImage.

[7]  P Bartolini,et al.  Temperature and SAR measurement errors in the evaluation of metallic linear structures heating during MRI using fluoroptic® probes , 2007, Physics in medicine and biology.

[8]  Louis Lemieux,et al.  Identification of EEG Events in the MR Scanner: The Problem of Pulse Artifact and a Method for Its Subtraction , 1998, NeuroImage.

[9]  Robert Turner,et al.  A Method for Removing Imaging Artifact from Continuous EEG Recorded during Functional MRI , 2000, NeuroImage.

[10]  Louis Lemieux,et al.  Safety of localizing epilepsy monitoring intracranial electroencephalograph electrodes using MRI: Radiofrequency-induced heating , 2008, Journal of magnetic resonance imaging : JMRI.

[11]  John S. Thornton,et al.  Functional MRI with active, fully implanted, deep brain stimulation systems: Safety and experimental confounds , 2007, NeuroImage.

[12]  David W. Carmichael,et al.  Image quality issues , 2009 .

[13]  John S. Thornton,et al.  Feasibility of simultaneous intracranial EEG-fMRI in humans: A safety study , 2010, NeuroImage.

[14]  S. Debener,et al.  Effects of simultaneous EEG recording on MRI data quality at 1.5, 3 and 7 tesla. , 2008, International journal of psychophysiology : official journal of the International Organization of Psychophysiology.

[15]  Louis Lemieux,et al.  Simultaneous intracranial EEG and fMRI of interictal epileptic discharges in humans , 2011, NeuroImage.