Gas-exchange function of a preprimed pediatric oxygenator stored for one year for emergency cardiopulmonary bypass.

To save priming time and perform more rapid initiation of emergency cardiopulmonary bypass for acute cardiopulmonary failure, an extracorporeal circuit with a hollow-fiber oxygenator (EL-2000 for pediatric use; Kurary Co. Ltd., Osaka, Japan) was preprimed, and the gas-exchange function was evaluated after 1 year of storage. EL-2000 has a dense polyolefin membrane with a surface area of 0.3 m2. When the bypass flow rates were 250, 500, 1,000, and 1,500 ml/min with 100% oxygen at the same flow rate as the bypass blood flow (namely, V/Q = 1) to the oxygenator, oxygen transport rates of the stored oxygenator were 19.6 +/- 0.3, 38.3 +/- 0.41, 64.4 +/- 0.9, and 76.4 +/- 2.7 ml/min (n = 5, mean +/- SD), respectively. PCO2 differences between pre- and postoxygenator blood (delta PCO2) were 18.6 +/- 1.4, 12.0 +/- 1.6, and 4.4 +/- 1.2 mm Hg at V/Q = 1 and the same bypass blood flow rates, respectively, excluding 1,500 ml/min, the data for which were excluded because of preparatory failure. PCO2 removal indices (defined as the ratio of delta PCO2 to PCO2 in preoxygenator blood) were 0.45 +/- 0.03, 0.29 +/- 0.12, and 0.10 +/- 0.03, respectively. Though the evaluation was done using only a single oxygenator, we feel strongly that the gas-exchange function of the preprimed dense-membrane hollow fiber oxygenator will be preserved even after 1 year of storage.