Investigating the performance of a wrist stabilization device for image-guided percutaneous scaphoid fixation

Purpose   Conventional navigated surgery relies on placement of a reference marker on the anatomy of interest. However, placement of such a marker is not readily feasible in small anatomic regions such as the scaphoid bone of the wrist. This study aimed to develop an alternative mechanism for patient tracking that could be used to perform navigated percutaneous scaphoid fixation.Methods   A prototype wrist stabilization device was developed to immobilize the scaphoid relative to a reference marker attached to the device. A position measurement system and 3D fluoroscopy were used to study the accuracy and limitations of wrist stabilization during simulated clinical usage with a cadaver specimen. Reference markers mounted on the device were used to measure intra-device motion. Radiometallic beads implanted in the scaphoid were used to measure patient-device motion. Navigated planning and guidance of scaphoid fixation were performed in five cadaver and eight “ideally immobilized” plastic specimens. Postoperative 3D fluoroscopy was used to assess the accuracy of navigated drilling.Results   The average intra-device motion was 1.9 mm during load application, which was elastically recovered upon release of the load. Scaphoid motion relative to the reference marker was predominately rotational with an average displacement of 1.25 mm and $$2.0^{\circ }$$. There was no significant difference in the accuracy of navigated drilling between the cadaver specimens and the ideally immobilized group.Conclusions   The prototype wrist stabilization device meets the criteria for effective wrist stabilization. This study provides insight concerning proper use of the device to minimize scaphoid displacement and design recommendations to improve immobilization.

[1]  Randy E. Ellis,et al.  Calibration and Use of Intraoperative Cone-Beam Computed Tomography: An In-Vitro Study for Wrist Fracture , 2010, MICCAI.

[2]  Gerlig Widmann,et al.  Kirschner wire placement in scaphoid bones using intraoperative CT-guided stereotaxy , 2013, Minimally invasive therapy & allied technologies : MITAT : official journal of the Society for Minimally Invasive Therapy.

[3]  Randy E. Ellis,et al.  Volume slicing of cone-beam computed tomography images for navigation of percutaneous scaphoid fixation , 2012, International Journal of Computer Assisted Radiology and Surgery.

[4]  Musa Citak,et al.  Computer-assisted navigation of volar percutaneous scaphoid placement. , 2010, The Journal of hand surgery.

[5]  Berthold K. P. Horn,et al.  Closed-form solution of absolute orientation using unit quaternions , 1987 .

[6]  Randall Smith,et al.  Estimating Uncertain Spatial Relationships in Robotics , 1987, Autonomous Robot Vehicles.

[7]  Philippe A Liverneaux,et al.  Kirschner wire placement in scaphoid bones using fluoroscopic navigation: a cadaver study comparing conventional techniques with navigation , 2008, The international journal of medical robotics + computer assisted surgery : MRCAS.

[8]  Musa Citak,et al.  Navigated scaphoid screw placement using customized scaphoid splint: an anatomical study , 2010, Archives of Orthopaedic and Trauma Surgery.

[9]  Andrew D. Wiles,et al.  Accuracy assessment and interpretation for optical tracking systems , 2004, Medical Imaging: Image-Guided Procedures.

[10]  Frank Langlotz,et al.  Technical Approaches to Computer-Assisted Orthopedic Surgery , 2004, European Journal of Trauma.