Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation

We present a case report of a 74‐year‐old male patient with an implantable cardioverter defibrillator who suffered an inappropriate defibrillation shock while bathing in the tub. Insight in the ICD stored electrogram episodes revealed electromagnetic interferences, with a typical 50 Hz electrical artifact mimicking fast ventricular tachycardia as a device misinterpreted. After this event, the maintenance workers investigated the electrical installation in the bathroom and revealed that there was voltage leaking between electrical installation and metal pipes. After the repair was completed without any additional programming, the patient has had no subsequent shocks.

[1]  T. Niendorf,et al.  Magnetic Field-Induced Interactions between Phones Containing Magnets and Cardiovascular Implantable Electronic Devices - Flip it to be Safe? , 2021, Heart rhythm.

[2]  M. Shenasa,et al.  Inappropriate ICD shock due to hot tub-induced external electrical interference. , 2018, Journal of electrocardiology.

[3]  N. Marx,et al.  Are patients with cardiac implants protected against electromagnetic interference in daily life and occupational environment? , 2015, European heart journal.

[4]  Kenneth A Ellenbogen,et al.  Troubleshooting implanted cardioverter defibrillator sensing problems I. , 2014, Circulation. Arrhythmia and electrophysiology.

[5]  M. Hours,et al.  Interference between Active Implanted Medical Devices and Electromagnetic Field Emitting Devices is Rare but Real: Results of an Incidence Study in a Population of Physicians in France , 2014, Pacing and clinical electrophysiology : PACE.

[6]  Nikolaus Marx,et al.  Electromagnetic Interference With Implantable Cardioverter-Defibrillators at Power Frequency: An In Vivo Study , 2014, Circulation.

[7]  Saman Nazarian,et al.  Effects of external electrical and magnetic fields on pacemakers and defibrillators: from engineering principles to clinical practice. , 2013, Circulation.

[8]  A. Moss,et al.  Impact of carvedilol and metoprolol on inappropriate implantable cardioverter-defibrillator therapy: the MADIT-CRT trial (Multicenter Automatic Defibrillator Implantation With Cardiac Resynchronization Therapy). , 2013, Journal of the American College of Cardiology.

[9]  J. Hartikainen,et al.  Electromagnetic interference with cardiac pacemakers and implantable cardioverter-defibrillators from low-frequency electromagnetic fields in vivo. , 2013, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.

[10]  K. Jokela,et al.  ICNIRP Guidelines GUIDELINES FOR LIMITING EXPOSURE TO TIME-VARYING , 1998 .

[11]  Wojciech Zareba,et al.  Inappropriate implantable cardioverter-defibrillator shocks in MADIT II: frequency, mechanisms, predictors, and survival impact. , 2008, Journal of the American College of Cardiology.

[12]  A. Bahl,et al.  Inappropriate shock delivery by implantable cardioverter defibrillator due to electrical interference with washing machine. , 2007, International journal of cardiology.

[13]  A. A. Khadra,et al.  Detection of refrigerator-associated 60 Hz alternating current as ventricular fibrillation by an implantable defibrillator , 2006 .

[14]  M. Sweesy Understanding electromagnetic interference. , 2004, Heart rhythm.

[15]  Ayman S Al Khadra,et al.  Detection of refrigerator-associated 60 Hz alternating current as ventricular fibrillation by an implantable defibrillator. , 2006, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.