Photoplethysmograph Signal Reconstruction Based on a Novel Hybrid Motion Artifact Detection–Reduction Approach. Part I: Motion and Noise Artifact Detection

Motion and noise artifacts (MNA) are a serious obstacle in utilizing photoplethysmogram (PPG) signals for real-time monitoring of vital signs. We present a MNA detection method which can provide a clean vs. corrupted decision on each successive PPG segment. For motion artifact detection, we compute four time-domain parameters: (1) standard deviation of peak-to-peak intervals (2) standard deviation of peak-to-peak amplitudes (3) standard deviation of systolic and diastolic interval ratios, and (4) mean standard deviation of pulse shape. We have adopted a support vector machine (SVM) which takes these parameters from clean and corrupted PPG signals and builds a decision boundary to classify them. We apply several distinct features of the PPG data to enhance classification performance. The algorithm we developed was verified on PPG data segments recorded by simulation, laboratory-controlled and walking/stair-climbing experiments, respectively, and we compared several well-established MNA detection methods to our proposed algorithm. All compared detection algorithms were evaluated in terms of motion artifact detection accuracy, heart rate (HR) error, and oxygen saturation (SpO2) error. For laboratory controlled finger, forehead recorded PPG data and daily-activity movement data, our proposed algorithm gives 94.4, 93.4, and 93.7% accuracies, respectively. Significant reductions in HR and SpO2 errors (2.3 bpm and 2.7%) were noted when the artifacts that were identified by SVM-MNA were removed from the original signal than without (17.3 bpm and 5.4%). The accuracy and error values of our proposed method were significantly higher and lower, respectively, than all other detection methods. Another advantage of our method is its ability to provide highly accurate onset and offset detection times of MNAs. This capability is important for an automated approach to signal reconstruction of only those data points that need to be reconstructed, which is the subject of the companion paper to this article. Finally, our MNA detection algorithm is real-time realizable as the computational speed on the 7-s PPG data segment was found to be only 7 ms with a Matlab code.

[1]  Tomás Ward,et al.  Artifact Removal in Physiological Signals—Practices and Possibilities , 2012, IEEE Transactions on Information Technology in Biomedicine.

[2]  John M. Graybeal,et al.  The Effect of Motion on Pulse Oximetry and Its Clinical Significance , 2007, Anesthesia and analgesia.

[3]  Ronald M. Aarts,et al.  Reducing motion artifacts in photoplethysmograms by using relative sensor motion: phantom study , 2012, Journal of biomedical optics.

[4]  Stephen James Wilson,et al.  A computational system to optimise noise rejection in photoplethysmography signals during motion or poor perfusion states , 2006, Medical and Biological Engineering and Computing.

[5]  Kang-Ming Chang,et al.  Pulse Rate Derivation and Its Correlation with Heart Rate , 2009 .

[6]  Y. Mendelson,et al.  A Comparative Evaluation of Adaptive Noise Cancellation Algorithms for Minimizing Motion Artifacts in a Forehead-Mounted Wearable Pulse Oximeter , 2007, 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[7]  Sun K. Yoo,et al.  Motion artifact reduction in photoplethysmography using independent component analysis , 2006, IEEE Transactions on Biomedical Engineering.

[8]  T. L. Rusch,et al.  Signal processing methods for pulse oximetry , 1996, Comput. Biol. Medicine.

[9]  Youngil Kim,et al.  Design of filter to reject motion artifact of pulse oximetry , 2004, Comput. Stand. Interfaces.

[10]  Sheng Lu,et al.  Automatic Real Time Detection of Atrial Fibrillation , 2009, Annals of Biomedical Engineering.

[11]  B. Natarajan,et al.  Onboard Tagging for Real-Time Quality Assessment of Photoplethysmograms Acquired by a Wireless Reflectance Pulse Oximeter , 2012, IEEE Transactions on Biomedical Circuits and Systems.

[12]  K. Wasserman,et al.  Effects of hypoxic hypoxia on O2 uptake and heart rate kinetics during heavy exercise. , 1996, Journal of applied physiology.

[13]  R. Sahni,et al.  Motion resistant pulse oximetry in neonates , 2003, Archives of disease in childhood. Fetal and neonatal edition.

[14]  C. Vosa,et al.  Coronary perfusion: impact of flow dynamics and geometric design of 2 different aortic prostheses of similar size. , 2012, The Journal of thoracic and cardiovascular surgery.

[15]  Manan Bawa,et al.  Comparison of different signal processing methods for reducing artifacts from photoplethysmograph signal , 2011, 2011 IEEE INTERNATIONAL CONFERENCE ON ELECTRO/INFORMATION TECHNOLOGY.

[16]  John Guttag,et al.  Real time reconstruction of quasiperiodic multi parameter physiological signals , 2012, EURASIP J. Adv. Signal Process..

[17]  A. Samorodov Application of a fuzzy integral for weak classifiers boosting , 2011, Pattern Recognition and Image Analysis.

[18]  S J Barker,et al.  Effects of motion, ambient light, and hypoperfusion on pulse oximeter function. , 1998, Journal of clinical anesthesia.

[19]  R G Mark,et al.  Robust heart rate estimation from multiple asynchronous noisy sources using signal quality indices and a Kalman filter , 2008, Physiological measurement.

[20]  S J Barker,et al.  The Effects of Motion on the Performance of Pulse Oximeters in Volunteers (Revised publication) , 1997, Anesthesiology.

[21]  Kejia Li,et al.  A Wireless Reflectance Pulse Oximeter With Digital Baseline Control for Unfiltered Photoplethysmograms , 2012, IEEE Transactions on Biomedical Circuits and Systems.

[22]  Eduardo Gil,et al.  Detection of decreases in the amplitude fluctuation of pulse photoplethysmography signal as indication of obstructive sleep apnea syndrome in children , 2008, Biomed. Signal Process. Control..

[23]  W Karlen,et al.  Photoplethysmogram signal quality estimation using repeated Gaussian filters and cross-correlation , 2012, Physiological measurement.

[24]  Jaques Reifman,et al.  Application of Information Technology: A Method for Automatic Identification of Reliable Heart Rates Calculated from ECG and PPG Waveforms , 2006, J. Am. Medical Informatics Assoc..

[25]  B. Saltin,et al.  Maximal perfusion of skeletal muscle in man. , 1985, The Journal of physiology.

[26]  Kirk H. Shelley,et al.  Statistical approach for the detection of motion/noise artifacts in Photoplethysmogram , 2011, 2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[27]  Paul B. Batchelder,et al.  A characterization of motion affecting pulse oximetry in 350 patients. , 2002, Anesthesia and analgesia.

[28]  Steve Warren,et al.  Two-Stage Approach for Detection and Reduction of Motion Artifacts in Photoplethysmographic Data , 2010, IEEE Transactions on Biomedical Engineering.

[29]  E. Hari Krishna,et al.  Use of spectral estimation methods for computation of SpO2 from artifact reduced PPG signals , 2011, 2011 IEEE Recent Advances in Intelligent Computational Systems.

[30]  Rolando Hong Enríquez,et al.  Analysis of the photoplethysmographic signal by means of the decomposition in principal components. , 2002, Physiological measurement.

[31]  Yuan-ting Zhang,et al.  Reduction of motion artifact in pulse oximetry by smoothed pseudo Wigner-Ville distribution , 2005, Journal of NeuroEngineering and Rehabilitation.

[32]  B. Hjorth EEG analysis based on time domain properties. , 1970, Electroencephalography and clinical neurophysiology.

[33]  B. Hjorth The physical significance of time domain descriptors in EEG analysis. , 1973, Electroencephalography and clinical neurophysiology.

[34]  R. Olfati-Saber,et al.  Consensus Filters for Sensor Networks and Distributed Sensor Fusion , 2005, Proceedings of the 44th IEEE Conference on Decision and Control.

[35]  Guang-Zhong Yang,et al.  Ratiometric Artifact Reduction in Low Power Reflective Photoplethysmography , 2011, IEEE Transactions on Biomedical Circuits and Systems.

[36]  Boreom Lee,et al.  Improved elimination of motion artifacts from a photoplethysmographic signal using a Kalman smoother with simultaneous accelerometry , 2010, Physiological measurement.

[37]  Chih-Jen Lin,et al.  A Practical Guide to Support Vector Classication , 2008 .

[38]  K. Nakajima,et al.  Monitoring of heart and respiratory rates by photoplethysmography using a digital filtering technique. , 1996, Medical engineering & physics.

[39]  Ju-Won Lee,et al.  The Periodic Moving Average Filter for Removing Motion Artifacts from PPG Signals , 2007 .

[40]  N H Lovell,et al.  Signal quality measures for pulse oximetry through waveform morphology analysis , 2011, Physiological measurement.

[41]  E. Hari Krishna,et al.  A Novel Approach for Motion Artifact Reduction in PPG Signals Based on AS-LMS Adaptive Filter , 2012, IEEE Transactions on Instrumentation and Measurement.