A feasibility study for using ABS plastic and a low-cost 3D printer for patient-specific brachytherapy mould design

This feasibility study aims to determine if a low-cost 3D printer (BitsFromBytes 3D Touch) with ABS plastic can print custom mould structures and catheter channels defined in a brachytherapy treatment planning system (Nucletron Oncentra) for patient-specific treatment. Printer accuracy was evaluated through physical measurement, and print quality was investigated by adjusting print parameters (print speed, layer thickness, percentage infill). Catheter positioning and reproducibility were measured over repeated insertions. ABS plastic water equivalency was investigated by comparing Ir-192 HDR source dose distributions, measured with radiochromic film, in ABS plastic and in water. Structures and catheter channels were printed accurately to within 0.5 mm laterally and 1 mm in the vertical print direction. Adjusting print parameters could reduce print time, albeit with reduced print quality. 3.5 mm channel diameters allowed for easy catheter insertion. Catheter positioning was reproducible to within 0.5 mm but, because of catheter flex within the channel, was on average 1 mm offset from defined TPS positions. This offset could be accounted for by repeating the treatment planning CT scan with the printed mould positioned on the patient. Dose attenuation in ABS plastic and in water was equivalent to within the measurement limitations. While clinical uses for this particular low-cost printer and ABS plastic are limited by print size restrictions and non-certification for biocompatibility, it has been demonstrated that a low-cost 3D printer set-up can accurately create custom moulds and catheter channels potentially acceptable for clinical use.

[1]  Jeroen Belien,et al.  Dose optimization in HDR brachytherapy: A literature review of quantitative models , 2011 .

[2]  Jeroen Belien,et al.  Dose optimization in high-dose-rate brachytherapy: A literature review of quantitative models from 1990 to 2010 , 2014 .

[3]  A. Gerbaulet,et al.  The GEC ESTRO handbook of brachytherapy , 2002 .

[4]  Benjamin Guix,et al.  Treatment of skin carcinomas of the face by high-dose-rate brachytherapy and custom-made surface molds , 2000 .

[5]  J Cunha,et al.  WE-F-16A-01: Commissioning and Clinical Use of PC-ISO for Customized, 3D Printed, Gynecological Brachytherapy Applicators. , 2014, Medical physics.

[6]  J. Adam M. Cunha,et al.  Evaluation of PC-ISO for customized, 3D printed, gynecologic 192Ir HDR brachytherapy applicators. , 2015, Journal of applied clinical medical physics.

[7]  Aradhana Katke,et al.  Dosimetric comparison of various optimization techniques for high dose rate brachytherapy of interstitial cervix implants , 2010, Journal of applied clinical medical physics.

[8]  J. Adam M. Cunha,et al.  An algorithm for computing customized 3D printed implants with curvature constrained channels for enhancing intracavitary brachytherapy radiation delivery , 2013, 2013 IEEE International Conference on Automation Science and Engineering (CASE).

[9]  L John Schreiner,et al.  SU-E-T-04: 3D Printed Patient-Specific Surface Mould Applicators for Brachytherapy Treatment of Superficial Lesions. , 2014, Medical physics.

[10]  David W. Rosen,et al.  Additive Manufacturing Technologies: Rapid Prototyping to Direct Digital Manufacturing , 2009 .

[11]  Aaron Fenster,et al.  A Novel Approach for Real-Time, Personalized Breast HDR Brachytherapy Treatment Using 3D Printing Technology , 2014 .

[12]  Rajesh Kumar,et al.  Acrylonitrile Butadiene Styrene (ABS) plastic‐based low cost tissue equivalent phantom for verification dosimetry in IMRT , 2009, Journal of applied clinical medical physics.

[13]  J. Williamson,et al.  Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations. , 2003, Medical physics.