Comparison of methods for the measurement of radiation dose distributions in high dose rate (HDR) brachytherapy: Ge-doped optical fiber, EBT3 Gafchromic film, and PRESAGE® radiochromic plastic.

PURPOSE Dose distribution measurement in clinical high dose rate (HDR) brachytherapy is challenging, because of the high dose gradients, large dose variations, and small scale, but it is essential to verify accurate treatment planning and treatment equipment performance. The authors compare and evaluate three dosimetry systems for potential use in brachytherapy dose distribution measurement: Ge-doped optical fibers, EBT3 Gafchromic film with multichannel analysis, and the radiochromic material PRESAGE® with optical-CT readout. METHODS Ge-doped SiO2 fibers with 6 μm active core and 5.0 mm length were sensitivity-batched and their thermoluminescent properties used via conventional heating and annealing cycles. EBT3 Gafchromic film of 30 μm active thickness was calibrated in three color channels using a nominal 6 MV linear accelerator. A 48-bit transmission scanner and advanced multichannel analysis method were utilized to derive dose measurements. Samples of the solid radiochromic polymer PRESAGE® , 60 mm diameter and 100 mm height, were analyzed with a parallel beam optical CT scanner. Each dosimetry system was used to measure the dose as a function of radial distance from a Co-60 HDR source, with results compared to Monte Carlo TG-43 model data. Each system was then used to measure the dose distribution along one or more lines through typical clinical dose distributions for cervix brachytherapy, with results compared to treatment planning system (TPS) calculations. Purpose-designed test objects constructed of Solid Water and held within a full-scatter water tank were utilized. RESULTS All three dosimetry systems reproduced the general shape of the isolated source radial dose function and the TPS dose distribution. However, the dynamic range of EBT3 exceeded those of doped optical fibers and PRESAGE® , and the latter two suffered from unacceptable noise and artifact. For the experimental conditions used in this study, the useful range from an isolated HDR source was 5-40 mm for fibers, 3-50 mm for EBT3, and 4-21 mm for PRESAGE® . Fibers demonstrated some over-response at very low dose levels, suffered from volume averaging effects in the dose distribution measurement, and exhibited up to 9% repeatability variation over three repeated measurements. EBT3 demonstrated excellent agreement with Monte Carlo and TPS dose distributions, with up to 3% repeatability over three measurements. PRESAGE® gave promising results, being the only true 3D dosimeter, but artifacts and noise were apparent. CONCLUSIONS The comparative response of three emerging dosimetry systems for clinical brachytherapy dose distribution measurement has been investigated. Ge-doped optical fibers have excellent spatial resolution for single-direction measurement but are currently too large for complex dose distribution assessment. The use of PRESAGE® with optical-CT readout gave promising results in the measurement of true 3D dose distributions but further development work is required to reduce noise and improve dynamic range for brachytherapy dose distribution measurements. EBT3 Gafchromic film with multichannel analysis demonstrated accurate and reproducible measurement of dose distributions in HDR brachytherapy. Calibrated dose measurements were possible with agreement within 1.5% of TPS dose calculations. The suitability of EBT3 as a dosimeter for 2D quality control or commissioning work has been demonstrated.

[1]  Andrew Nisbet,et al.  Verification of high dose rate brachytherapy dose distributions with EBT3 Gafchromic film quality control techniques , 2013, Physics in medicine and biology.

[2]  Larry A DeWerd,et al.  A dosimetric uncertainty analysis for photon-emitting brachytherapy sources: report of AAPM Task Group No. 138 and GEC-ESTRO. , 2011, Medical physics.

[3]  Zdenka Kuncic,et al.  Water and tissue equivalence of a new PRESAGE(®) formulation for 3D proton beam dosimetry: a Monte Carlo study. , 2012, Medical physics.

[4]  J J Wilkens,et al.  Comparison of Gafchromic EBT2 and EBT3 films for clinical photon and proton beams. , 2012, Medical physics.

[5]  I. Kawrakow,et al.  On the characterization and uncertainty analysis of radiochromic film dosimetry. , 2009, Medical physics.

[6]  M. Perucha,et al.  A comparison between radiochromic EBT2 film model and its predecessor EBT film model. , 2013, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[7]  M. Oldham,et al.  Polymer gel dosimetry. , 2010, Physics in medicine and biology.

[8]  J Adamovics,et al.  Dosimetry of the microSelectron-HDR Ir-192 source using PRESAGE and optical CT. , 2009, Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine.

[9]  Peter E Metcalfe,et al.  Radiochromic film for medical radiation dosimetry , 2003 .

[10]  D. Bradley,et al.  Physics-aspects of dose accuracy in high dose rate (HDR) brachytherapy: source dosimetry, treatment planning, equipment performance and in vivo verification techniques , 2012, Journal of contemporary brachytherapy.

[11]  S. Chiu‐Tsao,et al.  Evaluation of material heterogeneity dosimetric effects using radiochromic film for COMS eye plaques loaded with (125)I seeds (model I25.S16). , 2012, Medical physics.

[12]  J. Williamson,et al.  Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations. , 2004 .

[13]  K. Johansson,et al.  Quality assurance in brachytherapy: the displacement effect in the vicinity of 60Co and 192Ir brachytherapy sources. , 1993, Physics in medicine and biology.

[14]  N. Patel,et al.  The non-uniformity correction factor for the cylindrical ionization chambers in dosimetry of an HDR 192Ir brachytherapy source , 2006, Journal of medical physics.

[15]  Commissioning a CT-compatible LDR tandem and ovoid applicator using Monte Carlo calculation and 3D dosimetry. , 2012, Medical physics.

[16]  Simon J. Doran,et al.  Characterization of a parallel-beam CCD optical-CT apparatus for 3D radiation dosimetry , 2006, Physics in medicine and biology.

[17]  J. Galvin,et al.  Radiochromic film dosimetry: recommendations of AAPM Radiation Therapy Committee Task Group 55. American Association of Physicists in Medicine. , 1998, Medical physics.

[18]  N. Krstajić,et al.  Focusing optics of a parallel beam CCD optical tomography apparatus for 3D radiation gel dosimetry , 2006, Physics in medicine and biology.

[19]  D. Medich,et al.  Dependence of Yb-169 absorbed dose energy correction factors on self-attenuation in source material and photon buildup in water. , 2010, Medical physics.

[20]  C. De Wagter,et al.  On the accuracy of monomer/polymer gel dosimetry in the proximity of a high-dose-rate 192Ir source. , 2001, Physics in medicine and biology.

[21]  Andrea Molineu,et al.  Credentialing results from IMRT irradiations of an anthropomorphic head and neck phantom. , 2013, Medical physics.

[22]  Jan Seuntjens,et al.  Radiochromic film dosimetry of HDR (192)Ir source radiation fields. , 2011, Medical physics.

[23]  J. Williamson,et al.  Report of the Task Group 186 on model-based dose calculation methods in brachytherapy beyond the TG-43 formalism: current status and recommendations for clinical implementation. , 2012, Medical physics.

[24]  S. Webb,et al.  Experimental 3D dosimetry around a high-dose-rate clinical 192Ir source using a polyacrylamide gel (PAG) dosimeter. , 1999, Physics in medicine and biology.

[25]  R. Nath,et al.  Influence of the variation of energy spectra with depth in the dosimetry of 192Ir using LiF TLD. , 1988, Physics in medicine and biology.

[26]  F Maes,et al.  Calibrating page sized Gafchromic EBT3 films. , 2013, Medical physics.

[27]  Nikola Krstajić,et al.  Fast laser scanning optical-CT apparatus for 3D radiation dosimetry , 2007, Physics in medicine and biology.

[28]  J. Lambert,et al.  In vivo dosimeters for HDR brachytherapy: a comparison of a diamond detector, MOSFET, TLD, and scintillation detector. , 2007, Medical physics.

[29]  K. Jordan Review of recent advances in radiochromic materials for 3D dosimetry , 2010 .

[30]  M. Mitch,et al.  Measurement of the absorbed dose distribution near an 192Ir intravascular brachytherapy seed using a high-spatial-resolution gel dosimetry system , 2012, Physics in medicine and biology.

[31]  F Ballester,et al.  Technical note: Dosimetric study of a new Co-60 source used in brachytherapy. , 2007, Medical physics.

[32]  K. Hogstrom,et al.  Dose-response curve of EBT, EBT2, and EBT3 radiochromic films to synchrotron-produced monochromatic x-ray beams. , 2012, Medical physics.

[33]  A. Micke,et al.  Multichannel film dosimetry with nonuniformity correction. , 2011, Medical physics.

[34]  J. Williamson,et al.  High dose-rate brachytherapy treatment delivery: report of the AAPM Radiation Therapy Committee Task Group No. 59. , 1998, Medical physics.

[35]  Simon J Doran,et al.  The history and principles of chemical dosimetry for 3-D radiation fields: gels, polymers and plastics. , 2009, Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine.

[36]  D. Bradley,et al.  Establishment of Ge-doped optical fibres as thermoluminescence dosimeters for brachytherapy. , 2012, Applied Radiation and Isotopes.

[37]  D. Bradley,et al.  Review of doped silica glass optical fibre: their TL properties and potential applications in radiation therapy dosimetry. , 2012, Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine.

[38]  K. Nithya,et al.  Evaluation of dosimetric functions for Ir-192 source using radiochromic film , 2009 .

[39]  M. Bidmead,et al.  A survey of quality control practices for high dose rate (HDR) and pulsed dose rate (PDR) brachytherapy in the United Kingdom , 2012, Journal of contemporary brachytherapy.

[40]  Maria F Chan,et al.  An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scan. , 2012, Medical physics.

[41]  Bruce Thomadsen,et al.  American Brachytherapy Society consensus guidelines for locally advanced carcinoma of the cervix. Part I: general principles. , 2012, Brachytherapy.

[42]  C. Fallai,et al.  Solid state TL detectors for in vivo dosimetry in brachytherapy. , 2012, Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine.

[43]  M. Oldham,et al.  On the Feasibility of Verification of 3D Dosimetry Near Brachytherapy Sources Using PRESAGE/Optical-CT , 2010, Journal of physics. Conference series.

[44]  J F Williamson,et al.  Code of practice for brachytherapy physics: report of the AAPM Radiation Therapy Committee Task Group No. 56. American Association of Physicists in Medicine. , 1997, Medical physics.

[45]  S. Muscat,et al.  Treatment planning study of the 3D dosimetric differences between Co-60 and Ir-192 sources in high dose rate (HDR) brachytherapy for cervix cancer , 2012, Journal of contemporary brachytherapy.

[46]  S. Chiu‐Tsao,et al.  Energy Dependence of the New Gafchromic EBT3 Film: Dose Response Curves for 50 kV, 6 and 15 MV X-Ray Beams * , 2012 .

[47]  Kari Tanderup,et al.  Identifying afterloading PDR and HDR brachytherapy errors using real-time fiber-coupled Al(2)O(3):C dosimetry and a novel statistical error decision criterion. , 2011, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[48]  C Baldock,et al.  A basic study of some normoxic polymer gel dosimeters. , 2002, Physics in medicine and biology.

[49]  M. Oldham,et al.  Characterization of a new radiochromic three-dimensional dosimeter. , 2006, Medical physics.

[50]  Jose Perez-Calatayud,et al.  Dose calculation for photon-emitting brachytherapy sources with average energy higher than 50 keV: report of the AAPM and ESTRO. , 2012, Medical physics.