Anterior cruciate ligament replacement. Knee stability and the effects of implants.
暂无分享,去创建一个
The anteroposterior stability of cadaveric knees was investigated. There was a wide range of normal laxity; knees were more stable at 90 degrees than at 20 degrees flexion. Anterior cruciate ligament implants with different stiffnesses were inserted; normal stability could always be restored, and the stiffness or extensibility of implants did not affect knee behaviour significantly. The tightness of implants was critical--small tensioning errors caused subluxation, inhibited knee extension and allowed damagingly high implant tensions. It is concluded that the tension of ligament implants could not be adjusted simply with a pre-set instrument; the procedure will remain critically dependent on the judgment of the operating surgeon.