Computational analysis of the importance of flow synchrony for cardiac ventricular assist devices

This paper presents a patient customised fluid-solid mechanics model of the left ventricle (LV) supported by a left ventricular assist device (LVAD). Six simulations were conducted across a range of LVAD flow protocols (constant flow, sinusoidal in-sync and sinusoidal counter-sync with respect to the cardiac cycle) at two different LVAD flow rates selected so that the aortic valve would either open (60mLs(-1)) or remain shut (80mLs(-1)). The simulation results indicate that varying LVAD flow in-sync with the cardiac cycle improves both myocardial unloading and the residence times of blood in the left ventricle. In the simulations, increasing LVAD flow during myocardial contraction and decreasing it during diastole improved the mixing of blood in the LV cavity. Additionally, this flow protocol had the effect of partly homogenising work across the myocardium when the aortic valve did not open, reducing myocardial stress and thereby improving unloading.

[1]  Pablo Lamata,et al.  An accurate, fast and robust method to generate patient-specific cubic Hermite meshes , 2011, Medical Image Anal..

[2]  L. Miller,et al.  Partial aortic valve fusion induced by left ventricular assist device. , 2000, The Annals of thoracic surgery.

[3]  Toshiaki Hisada,et al.  Multiphysics simulation of left ventricular filling dynamics using fluid-structure interaction finite element method. , 2004, Biophysical journal.

[4]  Marshall I Hertz,et al.  The registry of the International Society for Heart and Lung Transplantation: twentieth official adult heart transplant report--2003. , 2003, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[5]  David Nordsletten,et al.  Simulating left ventricular fluid-solid mechanics through the cardiac cycle under LVAD support , 2013, J. Comput. Phys..

[6]  Theodosios Korakianitis,et al.  Numerical simulation of cardiovascular dynamics with left heart failure and in-series pulsatile ventricular assist device. , 2006, Artificial organs.

[7]  H. Oertel,et al.  Fluid-Structure Coupled CFD Simulation of the Left Ventricular Flow During Filling Phase , 2005, Annals of Biomedical Engineering.

[8]  J Michael Gaziano,et al.  Relation between modifiable lifestyle factors and lifetime risk of heart failure. , 2009, JAMA.

[9]  P. Hunter,et al.  Mathematical model of geometry and fibrous structure of the heart. , 1991, The American journal of physiology.

[10]  Daniel Burkhoff,et al.  A computational method of prediction of the end-diastolic pressure–volume relationship by single beat , 2007, Nature Protocols.

[11]  I. LeGrice,et al.  Shear properties of passive ventricular myocardium. , 2002, American journal of physiology. Heart and circulatory physiology.

[12]  E L Yellin,et al.  Mechanisms of mitral valve motion during diastole. , 1981, The American journal of physiology.

[13]  Patrick Patrick Anderson,et al.  Fluid-solid interactions: modeling, simulation, bio-mechanical applications A three-dimensional fluid-structure interaction method for heart valve modelling , 2005 .

[14]  M Flannery,et al.  Low incidence of myocardial recovery after left ventricular assist device implantation in patients with chronic heart failure. , 1998, Circulation.

[15]  David Nordsletten,et al.  A non-conforming monolithic finite element method for problems of coupled mechanics , 2010, J. Comput. Phys..

[16]  David Nordsletten,et al.  Modelling left ventricular function under assist device support , 2011 .

[17]  Ranjit John,et al.  Orthotopic Heart Transplantation , 2010 .

[18]  H. Scheld,et al.  Left ventricular pressure and volume unloading during pulsatile versus nonpulsatile left ventricular assist device support. , 2004, The Annals of thoracic surgery.

[19]  C. Peskin,et al.  A three-dimensional computational method for blood flow in the heart. II. contractile fibers , 1989 .

[20]  A. McCulloch,et al.  Modelling cardiac mechanical properties in three dimensions , 2001, Philosophical Transactions of the Royal Society of London. Series A: Mathematical, Physical and Engineering Sciences.

[21]  G. Plank,et al.  Length-dependent tension in the failing heart and the efficacy of cardiac resynchronization therapy. , 2011, Cardiovascular research.

[22]  P. Raviart,et al.  Conforming and nonconforming finite element methods for solving the stationary Stokes equations I , 1973 .

[23]  P. Hunter,et al.  Fluid–solid coupling for the investigation of diastolic and systolic human left ventricular function , 2011 .

[24]  Y. Hirota,et al.  A Clinical Study of Left Ventricular Relaxation , 1980, Circulation.

[25]  P. Hunter,et al.  Laminar structure of the heart: ventricular myocyte arrangement and connective tissue architecture in the dog. , 1995, The American journal of physiology.

[26]  J. Copeland,et al.  Success rates of long-term circulatory assist devices used currently for bridge to heart transplantation. , 1996, ASAIO journal.

[27]  M. Quiñones,et al.  Accurate noninvasive quantification of stenotic aortic valve area by Doppler echocardiography. , 1986, Circulation.

[28]  Daniel Burkhoff,et al.  LVAD-induced reverse remodeling: basic and clinical implications for myocardial recovery. , 2006, Journal of cardiac failure.

[29]  N P Smith,et al.  Coupling multi-physics models to cardiac mechanics. , 2011, Progress in biophysics and molecular biology.

[30]  Patrick D. Anderson,et al.  A fluid-structure interaction method with solid-rigid contact for heart valve dynamics , 2006, J. Comput. Phys..

[31]  C. Peskin Numerical analysis of blood flow in the heart , 1977 .

[32]  Toshiaki Hisada,et al.  Computer Simulation of Blood Flow, Left Ventricular Wall Motion and Their Interrelationship by Fluid-Structure Interaction Finite Element Method , 2002 .

[33]  T. Böhlke,et al.  Partitioned Fluid–Solid Coupling for Cardiovascular Blood Flow , 2010, Annals of Biomedical Engineering.

[34]  R De Paulis,et al.  Opening and closing characteristics of the aortic valve after valve-sparing procedures using a new aortic root conduit. , 2001, The Annals of thoracic surgery.