Towards diagnostic flow measurement from a catheter tip pressure sensor

This paper reports the investigation of the additional ability of a catheter tip pressure sensor to measure flow rate. The sensor's ability to measure flow rate builds on an earlier development of temperature measurement from the pressure sensor. The flow rates of interest range from 0 to 30 mL/hour, those expected through a hydrocephalus shunt. The feasibility of making flow measurements from the MEMS piezoresistive pressure sensor was assessed in terms of the sensor's repeatability, its potential resolution and its temporal response. Within the flow range 0 to 30 mL/hour the sensor has demonstrated the ability to resolve differences in flow rate of less than 5 mL/hour. The sensor demonstrates high degree of repeatability with coefficient of variation of less than 5 % across all flow rates in the measurement range.

[1]  B. Iskandar,et al.  Pitfalls in the diagnosis of ventricular shunt dysfunction: radiology reports and ventricular size. , 1998, Pediatrics.

[2]  Wensheng Guo,et al.  Have we made progress in preventing shunt failure? A critical analysis. , 2008, Journal of neurosurgery. Pediatrics.

[3]  H. Rekate,et al.  Pediatric hydrocephalus outcomes: a review , 2012, Fluids and Barriers of the CNS.

[4]  R M Henkelman,et al.  Cerebrospinal fluid shunts: flow measurements with MR imaging. , 1989, Radiology.

[5]  J. Mathews,et al.  Cancer risk in 680 000 people exposed to computed tomography scans in childhood or adolescence: data linkage study of 11 million Australians , 2013, BMJ.

[6]  Y. Kajimoto,et al.  New concept for the pressure setting of a programmable pressure valve and measurement of in vivo shunt flow performed using a microflowmeter. Technical note. , 2000, Journal of neurosurgery.

[7]  S J Schiff,et al.  Randomized trial of cerebrospinal fluid shunt valve design in pediatric hydrocephalus. , 1998, Neurosurgery.

[8]  J. Madsen,et al.  Evaluation of the ShuntCheck Noninvasive Thermal Technique for Shunt Flow Detection in Hydrocephalic Patients , 2011, Neurosurgery.

[9]  A. Sivaganesan,et al.  Neuroimaging of ventriculoperitoneal shunt complications in children , 2012, Pediatric Radiology.

[10]  Y. Kanda,et al.  A graphical representation of the piezoresistance coefficients in silicon , 1982, IEEE Transactions on Electron Devices.

[11]  D. Thompson,et al.  Antibiotic-impregnated ventriculoperitoneal shunts—a multi-centre British paediatric neurosurgery group (BPNG) study using historical controls , 2011, Child's Nervous System.

[12]  Yozo Kanda,et al.  Piezoresistance effect of silicon , 1991 .

[13]  Simon Malpas,et al.  Intracranial temperature and pressure measurement: In vitro temperature sensing characteristics of the dual sensing element , 2013, 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC).