Closed‐loop double‐vasopressor automated system vs manual bolus vasopressor to treat hypotension during spinal anaesthesia for caesarean section: a randomised controlled trial

Hypotension necessitating vasopressor administration occurs commonly during caesarean section under spinal anaesthesia. We developed a novel vasopressor delivery system that automatically administers phenylephrine or ephedrine based on continuous non‐invasive arterial pressure monitoring. A phenylephrine bolus of 50 μg was given at 30‐s intervals when systolic blood pressure fell < 90% of baseline; an ephedrine bolus of 4 mg was given instead if systolic pressure fell < 90% of baseline together with a heart rate < 60 beats.min−1. The control group used manual boluses of either phenylephrine 100 μg or ephedrine 8 mg, administered at 1‐min intervals based on the same thresholds for systolic pressure and heart rate. This randomised, controlled, double‐blinded trial involved 213 healthy women who underwent elective caesarean delivery under spinal anaesthesia using 11 mg hyperbaric bupivacaine with 15 μg fentanyl and 100 μg morphine. The automated vasopressor group had better systolic pressure control, with 37/106 (34.9%) having any beat‐to‐beat systolic pressure reading < 80% of baseline compared with 63/107 (58.9%) in the control group (p < 0.001). There was no difference in the incidence of reactive hypertension, defined as systolic pressure > 120% of baseline, with 8/106 (7.5%) in the automated vasopressor group vs 14/107 (13.1%) in the control group, or total dose of vasopressors. The automated vasopressor group had lower median absolute performance error of 8.5% vs control of 9.8% (p = 0.013), and reduced incidence of nausea (1/106 (0.9%) vs 11/107 (10.3%), p = 0.005). Neonatal umbilical cord pH, umbilical lactate and Apgar scores were similar. Hence, our system afforded better control of maternal blood pressure and reduced nausea with no increase in reactive hypertension when compared with manual boluses.

[1]  J Schüttler,et al.  Precision and accuracy of a new device (CNAPTM) for continuous non-invasive arterial pressure monitoring: assessment during general anaesthesia. , 2010, British journal of anaesthesia.

[2]  S. Riskin-Mashiah,et al.  Autonomic Imbalance in Preeclampsia: Evidence for Increased Sympathetic Tone in Response to the Supine‐Pressor Test , 1998, Obstetrics and gynecology.

[3]  D. Chestnut The Dose-Dependent Effects of Phenylephrine for Elective Cesarean Delivery Under Spinal Anesthesia , 2011 .

[4]  Warwick D Ngan Kee,et al.  Prophylactic Phenylephrine Infusion for Preventing Hypotension During Spinal Anesthesia for Cesarean Delivery , 2004, Anesthesia and analgesia.

[5]  Kim S. Khaw,et al.  Prevention of Hypotension during Spinal Anesthesia for Cesarean Delivery: An Effective Technique Using Combination Phenylephrine Infusion and Crystalloid Cohydration , 2005, Anesthesiology.

[6]  A. Sia,et al.  Closed‐loop double‐vasopressor automated system to treat hypotension during spinal anaesthesia for caesarean section: a preliminary study , 2012, Anaesthesia.

[7]  K. Downey,et al.  Phenylephrine Infusion Versus Bolus Regimens During Cesarean Delivery Under Spinal Anesthesia: A Double-Blind Randomized Clinical Trial to Assess Hemodynamic Changes , 2012, Anesthesia and analgesia.

[8]  W. White,et al.  A Double-Blind, Placebo-Controlled Trial of Four Fixed Rate Infusion Regimens of Phenylephrine for Hemodynamic Support During Spinal Anesthesia for Cesarean Delivery , 2010, Anesthesia and analgesia.

[9]  J. Eisenach,et al.  Pregnancy and Ephedrine Increase the Release of Nitric Oxide in Ovine Uterine Arteries , 1996, Anesthesia and analgesia.

[10]  W D Ngan Kee,et al.  Randomized comparison of closed-loop feedback computer-controlled with manual-controlled infusion of phenylephrine for maintaining arterial pressure during spinal anaesthesia for caesarean delivery. , 2013, British journal of anaesthesia.

[11]  F. C. Greiss Pressure-flow relationship in the gravid uterine vascular bed. , 1966, American journal of obstetrics and gynecology.

[12]  P Middleton,et al.  Techniques for preventing hypotension during spinal anaesthesia for caesarean section. , 2006, The Cochrane database of systematic reviews.

[13]  Lee Hp,et al.  Sample size estimation for clinicians. , 1995 .

[14]  L. Leduc,et al.  Baroreflex function in normal pregnancy. , 1991, American journal of obstetrics and gynecology.

[15]  W. N. Kee Prevention of maternal hypotension after regional anaesthesia for caesarean section. , 2010 .

[16]  David W. Cooper,et al.  Fetal and Maternal Effects of Phenylephrine and Ephedrine during Spinal Anesthesia for Cesarean Delivery , 2002, Anesthesiology.

[17]  W. N. Ngan Kee Prevention of maternal hypotension after regional anaesthesia for caesarean section , 2010, Current opinion in anaesthesiology.

[18]  Steven L. Shafer,et al.  Measuring the predictive performance of computer-controlled infusion pumps , 1992, Journal of Pharmacokinetics and Biopharmaceutics.

[19]  W D Ngan Kee,et al.  Closed‐loop feedback computer‐controlled infusion of phenylephrine for maintaining blood pressure during spinal anaesthesia for caesarean section: a preliminary descriptive study * , 2007, Anaesthesia.

[20]  Roshan Fernando,et al.  Faculty Opinions recommendation of Closed-loop feedback computer-controlled infusion of phenylephrine for maintaining blood pressure during spinal anaesthesia for caesarean section: a preliminary descriptive study. , 2008 .

[21]  W. N. Ngan Kee,et al.  A Quantitative, Systematic Review of Randomized Controlled Trials of Ephedrine Versus Phenylephrine for the Management of Hypotension During Spinal Anesthesia for Cesarean Delivery , 2002, Anesthesia and analgesia.

[22]  B. Bein,et al.  Investigation of the agreement of a continuous non-invasive arterial pressure device in comparison with invasive radial artery measurement. , 2012, British journal of anaesthesia.

[23]  C. Chooi,et al.  Techniques for preventing hypotension during spinal anaesthesia for caesarean section. , 2017, The Cochrane database of systematic reviews.