Hybrid analytical method for calibrating a standard NaI(Tl) gamma-ray scintillation detector using a lateral hexagonal radioactive source

In various fields of radiation and nuclear physics, research and its applications help in the management and control of human activities, especially in the analysis of nuclear waste and measurement of low activity environmental samples. The current study uses a hybrid analytical technique and a 152Eu lateral hexagonal radioactive source to calibrate a standard cylindrical NaI(Tl) γ-ray scintillation detector. The technique is a combination of experimental measurements and theoretical calculations because for each environmental and/or radioactive waste sample, a separate reference calibration source equivalent to samples of different compositions is required, which needs additional funding. This method is suitable for any source–detector geometry. It can be used to calculate the geometric, total, and full-energy peak efficiency and the P/T ratio using the model probability of detecting γ-radiation photons, in particular from a lateral hexagonal radioactive source. The detection probability can be mainly influenced by the direction of the γ-quantum motion. This can help build more realistic and complex models than the typical point source case, where simple integration is used to numerically calculate the efficiency of the detection system. Using this method, the self-attenuation and coincidence-summing effect can be calculated for any gamma source matrices. The results obtained can show how useful this technique can be for calibrating complex gamma-ray detectors with any type of gamma-ray source.

[1]  M. Badawi,et al.  Calibration of cylindrical NaI(Tl) gamma-ray detector intended for truncated conical radioactive source , 2021, Nuclear Engineering and Technology.

[2]  M. M. Gouda,et al.  Characterization of the efficiency of a cubic NaI detector with rectangular cavity for axially positioned sources , 2020, Journal of Instrumentation.

[3]  S. Ashrafi,et al.  A method to calculate the gamma ray detection efficiency of a cylindrical NaI (Tl) crystal , 2018 .

[4]  D. Todorovic,et al.  Semiempirical Efficiency Calibration in Semiconductor HPGe Gamma-Ray Spectroscopy , 2018 .

[5]  M. M. Gouda,et al.  Full-Energy peak efficiency of an NaI(Tl) detector with coincidence summing correction showing the effect of the source-to-detector distance , 2017 .

[6]  M. M. Gouda,et al.  Calibration of 4π NaI(Tl) detectors with coincidence summing correction using new numerical procedure and ANGLE4 software , 2017 .

[7]  M. M. Gouda,et al.  Well-type NaI(Tl) detector efficiency using analytical technique and ANGLE 4 software based on radioactive point sources located out the well cavity , 2016 .

[8]  C. Alonso-Hernández,et al.  Validation of an efficiency calibration procedure for a coaxial n-type and a well-type HPGe detector used for the measurement of environmental radioactivity , 2016 .

[9]  M. M. Gouda,et al.  Mathematical method to calculate full-energy peak efficiency of detectors based on transfer technique , 2016 .

[10]  M. M. Gouda,et al.  Calibration of а single hexagonal NaI(Tl) detector using a new numerical method based on the efficiency transfer method , 2015 .

[11]  D. Todorovic,et al.  Calculation of HPGe efficiency for environmental samples: comparison of EFFTRAN and GEANT4 , 2014 .

[12]  A. El-Khatib,et al.  STUDY ON THE EFFECT OF THE SELF-ATTENUATION COEFFICIENT ON γ-RAY DETECTOR EFFICIENCY CALCULATED AT LOW AND HIGH ENERGY REGIONS , 2014 .

[13]  M. Hult,et al.  Testing efficiency transfer codes for equivalence. , 2010, Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine.

[14]  D. Arnold,et al.  On the Monte Carlo simulation of HPGe gamma-spectrometry systems. , 2009, Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine.

[15]  M. Roteta,et al.  Standardization of 67Ga by 4πγ (NaI) and 4πβ–γ coincidence methods , 2004 .

[16]  B. Quintana,et al.  Efficiency calibration of a HPGe detector in the [46.54-2000] keV energy range for the measurement of environmental samples , 2001 .

[17]  Xiaoling Zhang,et al.  The determination of 125I activity using sum-peak method with a well-type HPGe-detector-based spectrometer , 2001 .

[18]  M. Badawi,et al.  Measurements correlated to efficiency studies of NaI(Tl) scintillator with vertical hexagonal radioactive source , 2021, Journal of Instrumentation.

[19]  M. M. Gouda,et al.  Experimental verification of gamma-efficiency calculations for scintillation detectors in ANGLE 4 software , 2015 .

[20]  A. Dell'Acqua,et al.  Geant4—a simulation toolkit , 2003 .