A fast, digitally controlled flow proportional gas injection system for studies in lung function

The aim of this paper is to describe a device for flow proportional injection of tracer gas in the lungs of mechanically ventilated patients. This device may then be used for the study of the multiple breath indicator gas washout technique to determine the end-expiratory lung volume. Such a tracer gas injection device may also be used in the study of other techniques that rely on uptake and elimination of tracer gas by the lungs. In this paper, an injector is described which enables injection of indicator gas at a predetermined concentration in a breathing circuit independent of the type of breathing. The presented setup uses a control computer to produce steering signals to a multivalve array in proportion to the input breathing signals. The multivalve array consists of ten circular valves, each with a different diameter, which can be opened or closed individually according to the input signal of the array. By opening of a certain combination of valves an amount of sulphur hexafluoride gas proportional to the inspiratory breathing signal is released. The rate of transmission between the components of the injection system was 80 Hz. The injector has a full flow range between 0-10 L/min. The delay time between the breathing signal and the flow response was 70 ms. The aimed washin gas concentration of 1% SF/sub 6/ was achieved after 0.5 s. The study describes the results of tests to determine valve-flow ratios, step response and dynamic response of the injector. The flow output response of the injector system was shown to increase in input frequencies above 3 Hz. The valve flow ratios showed the largest relative deviation in the two smallest valves of the 10 valve array, respectively 0.005 L/min (25%) and 0.002 L/min (20%). We conclude that the injector can achieve a stable concentration of indicator gas in a breathing system with an accuracy of 0.005 L/min to execute the multiple breath indicator washout test in human subjects. The results of the study indicate that the injector may be of use in other application fields in respiratory physiology in which breathing circuit injection of indicator gas is required.

[1]  T Yamamura,et al.  [Measurement of functional residual capacity by sulfur hexafluoride washout]. , 1992, Masui. The Japanese journal of anesthesiology.

[2]  B Jonson,et al.  Measurement of Functional Residual Capacity by Sulfur Hexafluoride Washout , 1985, Anesthesiology.

[3]  Jan M. Bogaard,et al.  A non-invasive determination of lung perfusion compared with the direct Fick method , 1978, Pflügers Archiv.

[4]  M Paiva,et al.  Dual tracer single breath studies of gas transport in the lung. , 1979, Respiration physiology.

[5]  B Jonson,et al.  Respiratory mechanics in patients ventilated for critical lung disease. , 1996, The European respiratory journal.

[6]  C E Hahn,et al.  Assessment of cardiorespiratory function using oscillating inert gas forcing signals. , 1994, Journal of applied physiology.

[7]  G D Swanson,et al.  A fast gas-mixing system for breath-to-breath respiratory control studies. , 1982, Journal of applied physiology: respiratory, environmental and exercise physiology.

[8]  C E Hahn,et al.  Measurement of cardio-respiratory function using single frequency inspiratory gas concentration forcing signals. , 1994, Advances in experimental medicine and biology.

[9]  C. Hahn,et al.  Gas exchange in a three-compartment lung model analyzed by forcing sinusoids of N2O. , 1993, Journal of applied physiology.

[10]  C E Hahn Oxygen respiratory gas analysis by sine-wave measurement: a theoretical model. , 1996, Journal of applied physiology.

[11]  C E Hahn,et al.  A digital system for generating dynamic sinusoidal gas concentration signals. , 1988, Journal of applied physiology.

[12]  B Jonson,et al.  Measurement of lung volume by sulfur hexafluoride washout during spontaneous and controlled ventilation: further development of a method. , 1987, Anesthesiology.

[13]  O Werner,et al.  An analyzer for in-line measurement of expiratory sulfur hexafluoride concentration. , 1985, Anesthesiology.

[14]  CAN INCE,et al.  A new ventilation inhomogeneity index from multiple breath indicator gas washout tests in mechanically ventilated patients , 1993, Critical care medicine.

[15]  O Werner,et al.  Functional residual capacity and ventilation homogeneity in mechanically ventilated small neonates. , 1992, Journal of applied physiology.

[16]  B. Jonson,et al.  Device for flow-proportional admixture of tracer gas in lung-function studies , 2006, Medical and Biological Engineering and Computing.

[17]  S.A. Barton,et al.  Dynamic models and solutions for evaluating ventilation, perfusion and mass transfer in the lung. II. Analog solutions , 1988, IEEE Transactions on Biomedical Engineering.

[18]  Andreas Schulze,et al.  Measurement of Functional Residual Capacity by Sulfur Hexafluoride in Small-Volume Lungs during Spontaneous Breathing and Mechanical Ventilation , 1994, Pediatric Research.

[19]  C. Hahn,et al.  Oxygen transport with oscillations of inspired oxygen concentration. , 1997, Respiration physiology.

[20]  T. Clemmer,et al.  Automated sulfur hexafluoride washout functional residual capacity measurement system for any mode of mechanical ventilation as well as spontaneous respiration , 1990, Critical care medicine.

[21]  P E Huygen,et al.  PDPS: a pulmonary data processing system for assessment of gas exchange properties by multiple gas wash-out. , 1991, Computer methods and programs in biomedicine.

[22]  D J Gavaghan,et al.  A tidal breathing model of the forced inspired inert gas sinewave technique. , 1996, Respiration physiology.

[23]  J R Jansen,et al.  A computerized respiratory system including test functions of lung and circulation. , 1989, Journal of applied physiology.

[24]  C. Hahn,et al.  Dynamic models and solutions for evaluating ventilation, perfusion and mass transfer in the lung. I. The models , 1988, IEEE Transactions on Biomedical Engineering.

[25]  Thomas D. East,et al.  Automated measurement of functional residual capacity by sulfur hexafluoride washout , 2004, Journal of Clinical Monitoring.

[26]  C. Hahn,et al.  Measurement of respiratory parameters by using inspired oxygen sinusoidal forcing signals. , 1996, Journal of applied physiology.

[27]  G. Puppels,et al.  Noninvasive monitoring of nonshunted pulmonary capillary blood flow in the acute respiratory distress syndrome , 2000, Critical care medicine.

[28]  J Jacky,et al.  Programmable gas mixer for controlling concentration as a function of time. , 1980, Journal of applied physiology: respiratory, environmental and exercise physiology.

[29]  M Paiva,et al.  Gas mixing in the lungs of dogs and pigs. , 1982, Respiration physiology.

[30]  H. Stam,et al.  Design and validation of an indicator gas injector for multiple gas washout tests in mechanically ventilated patients , 1990, Critical care medicine.

[31]  D J Gavaghan,et al.  Measurement of dead–space in a model lung using an oscillating inspired argon signal , 1994, Acta anaesthesiologica Scandinavica.

[32]  S J Allen,et al.  Flow-proportional administration of nitric oxide with a new delivery system: inspiratory nitric oxide concentration fluctuation during different flow conditions. , 1997, Chest.