Clinical accuracy of three-dimensional fluoroscopy-based computer-assisted cervical pedicle screw placement: a retrospective comparative study of conventional versus computer-assisted cervical pedicle screw placement.

OBJECT The authors performed a retrospective clinical study to evaluate the feasibility and accuracy of cervical pedicle screw (CPS) placement using 3D fluoroscopy-based navigation (3D FN). METHODS The study involved 62 consecutive patients undergoing posterior stabilization of the cervical spine between 2003 and 2008. Thirty patients (126 screws) were treated using conventional techniques (CVTs) with a lateral fluoroscopic view, whereas 32 patients (150 screws) were treated using 3D FN. Screw positions were classified into 4 grades based on the pedicle wall perforations observed on postoperative CT. RESULTS The prevalence of perforations in the CVT group was 27% (34 screws): 92 (73.0%), 12 (9.5%), 6 (4.8%), and 16 (12.7%) for Grade 0 (no perforation), Grade 1 (perforation < 1 mm), Grade 2 (perforation ≥ 1 and < 2 mm), and Grade 3 (perforation ≥ 2 mm), respectively. In the 3D FN group, the prevalence of perforations was 18.7% (28 screws): 122 (81.3%), 17 (11.3%), 6 (4%), and 5 (3.3%) for Grades 0, 1, 2, and 3, respectively. Statistical analysis showed no significant difference in the prevalence of Grade 1 or higher perforations between the CVT and 3D FN groups. A higher prevalence of malpositioned CPSs was seen in Grade 2 or higher (17.5% vs 7.3%, p < 0.05) in the 3D FN group and Grade 3 (12.7% vs 7.3%, p < 0.05) perforations in the CVT group. The ORs for CPS malpositioning in the CVT group were 2.72 (95% CI 1.16-6.39) in Grade 2 or higher perforations and 3.89 (95% CI 1.26-12.02) in Grade 3 perforations. CONCLUSIONS Three-dimensional fluoroscopy-based navigation can improve the accuracy of CPS insertion; however, severe CPS malpositioning that causes injury to the vertebral artery or neurological complications can occur even with 3D FN. Advanced techniques for the insertion of CPSs and the use of modified insertion devices can reduce the risk of a malpositioned CPS and provide increased safety.

[1]  Phillip M Young,et al.  Placement of thoracolumbar pedicle screws using three-dimensional image guidance: experience in a large patient cohort. , 2009, Journal of neurosurgery. Spine.

[2]  E. Schneider,et al.  Biomechanical Analysis of Transpedicular Screw Fixation in the Subaxial Cervical Spine , 2004, Spine.

[3]  K. Takaoka,et al.  Cervical pedicle screw insertion: assessment of safety and accuracy with computer-assisted image guidance. , 2000, Journal of spinal disorders.

[4]  H. Kato,et al.  Computer-assisted screw insertion for cervical disorders in rheumatoid arthritis , 2007, European Spine Journal.

[5]  Manabu Ito,et al.  Complications of Pedicle Screw Fixation in Reconstructive Surgery of the Cervical Spine , 2000, Spine.

[6]  M M Panjabi,et al.  Internal morphology of human cervical pedicles. , 2000, Spine.

[7]  G. Lekovic,et al.  A comparison of two techniques in image-guided thoracic pedicle screw placement: a retrospective study of 37 patients and 277 pedicle screws. , 2007, Journal of neurosurgery. Spine.

[8]  R. Gaines,et al.  Accuracy of Cervical Pedicle Screw Placement Using the Funnel Technique , 2001, Spine.

[9]  Yasuo Ito,et al.  Clinical accuracy of 3D fluoroscopy-assisted cervical pedicle screw insertion. , 2008, Journal of neurosurgery. Spine.

[10]  K. Kaneda,et al.  Transpedicular screw fixation for traumatic lesions of the middle and lower cervical spine: description of the techniques and preliminary report. , 1994, Journal of spinal disorders.

[11]  S. Rajasekaran,et al.  Iso-C3D Fluoroscopy-based Navigation in Direct Pedicle Screw Fixation of Hangman Fracture: A Case Report , 2007, Journal of spinal disorders & techniques.

[12]  A. Minami,et al.  Improved accuracy of computer-assisted cervical pedicle screw insertion. , 2003, Journal of neurosurgery.

[13]  René Schmidt,et al.  Cervical Pedicle Screws: Conventional Versus Computer-Assisted Placement of Cannulated Screws , 2005, Spine.

[14]  M. Richter,et al.  Computer-assisted posterior instrumentation of the cervical and cervico-thoracic spine , 2004, European Spine Journal.

[15]  Kevin T Foley,et al.  Intraoperative Spinal Navigation , 2003, Spine.

[16]  M. Neo,et al.  Efficacy of computer-assisted pedicle screw insertion for cervical instability in RA patients , 2007, Rheumatology International.

[17]  K. Kaneda,et al.  Pedicle Screw Fixation for Nontraumatic Lesions of the Cervical Spine , 1997, Spine.

[18]  S Vidyadhara,et al.  Randomized Clinical Study to Compare the Accuracy of Navigated and Non-Navigated Thoracic Pedicle Screws in Deformity Correction Surgeries , 2007, Spine.

[19]  R. Gaines,et al.  Morphologic Characteristics of Human Cervical Pedicles , 1997, Spine.

[20]  Robert F Spetzler,et al.  Intraoperative Iso-C C-Arm Navigation in Craniospinal Surgery: The First 60 Cases , 2004, Neurosurgery.

[21]  T. Albert,et al.  Placement of pedicle screws in the human cadaveric cervical spine: comparative accuracy of three techniques. , 2000, Spine.

[22]  Huang Yu-liange Transpedicular screw placement evaluated by axial computed tomography of the cervical pedicle , 2008 .

[23]  T. Hirano,et al.  Indications for Cervical Pedicle Screw Instrumentation in Nontraumatic Lesions , 2008, Spine.

[24]  Y. Rampersaud,et al.  Clinical Accuracy of Fluoroscopic Computer-Assisted Pedicle Screw Fixation: A CT Analysis , 2005, Spine.

[25]  L. Nolte,et al.  Improved Accuracy of Pedicle Screw Insertion With Computer-Assisted Surgery: A Prospective Clinical Trial of 30 Patients , 1997, Spine.

[26]  L. Claes,et al.  Pedicle Screws Enhance Primary Stability in Multilevel Cervical Corpectomies: Biomechanical In Vitro Comparison of Different Implants Including Constrained and Nonconstrained Posterior Instumentations , 2003, Spine.

[27]  D. Silcox,et al.  Cervical Pedicle Screws Versus Lateral Mass Screws: Anatomic Feasibility and Biomechanical Comparison , 1997, Spine.

[28]  M. Neo,et al.  The Clinical Risk of Vertebral Artery Injury From Cervical Pedicle Screws Inserted in Degenerative Vertebrae , 2005, Spine.