Hemodynamic effects of partial ventricular support in chronic heart failure: results of simulation validated with in vivo data.

OBJECTIVE Current left ventricular assist devices are designed to provide full hemodynamic support for patients with end-stage failing hearts, but their use has been limited by operative risks, low reliability, and device-related morbidity. Such concerns have resulted in minimum use of left ventricular assist devices for destination therapy. We hypothesize that partial circulatory support, which could be achieved with small pumps implanted with less-invasive procedures, might expand the role of circulatory support devices for treatment of heart failure. METHODS We examine the hemodynamic effects of partial left ventricular support using a previously described computational model of the cardiovascular system. Results from simulations were validated by comparison with an in vivo hemodynamic study. RESULTS Simulations demonstrated that partial support (2-3 L/min) increased total cardiac output (left ventricular assist device output plus native heart output) by more than 1 L/min and decreased left ventricular end-diastolic pressure by 7 to 10 mm Hg with moderate-to-severe heart failure. Analyses showed that the hemodynamic benefits of increased cardiac output and decreased left ventricular end-diastolic pressure are greater in less-dilated and less-dysfunctional hearts. Both the relationships between ventricular assist device flow and cardiac output and ventricular assist device flow and left atrial pressure predicted by the model closely approximated the same relationships obtained during hemodynamic study in a bovine heart failure model. CONCLUSIONS Results suggest that a pump with a flow rate of 2 to 3 L/min could meaningfully affect cardiac output and blood pressure in patients with advanced compensated heart failure. The development of small devices capable of high reliability and minimal complications that can be implanted with less-invasive techniques is supported by these findings.

[1]  D. Burkhoff,et al.  A theoretic analysis of the effect of pulmonary vasodilation on pulmonary venous pressure: implications for inhaled nitric oxide therapy. , 1996, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[2]  K. Litwak,et al.  Monensin toxicosis in the domestic bovine calf: a large animal model of cardiac dysfunction. , 2005, Contemporary topics in laboratory animal science.

[3]  H. Schima,et al.  Weaning of rotary blood pump recipients after myocardial recovery: a computer study of changes in cardiac energetics. , 2004, The Journal of thoracic and cardiovascular surgery.

[4]  D. Burkhoff,et al.  Heart reduction surgery: an analysis of the impact on cardiac function. , 1997, The Journal of thoracic and cardiovascular surgery.

[5]  J. Cohn,et al.  The effect of carvedilol on morbidity and mortality in patients with chronic heart failure. U.S. Carvedilol Heart Failure Study Group. , 1996, The New England journal of medicine.

[6]  K. Sagawa,et al.  Instantaneous pressure-volume relation of the ejecting canine left atrium. , 1987, Circulation research.

[7]  R. Applegate,et al.  Cardiac Contraction and the Pressure-Volume Relationship , 1990 .

[8]  Milton Packer,et al.  Cardiac resynchronization in chronic heart failure. , 2002, The New England journal of medicine.

[9]  References , 1971 .

[10]  W. Santamore,et al.  Hemodynamic consequences of ventricular interaction as assessed by model analysis. , 1991, The American journal of physiology.

[11]  D. DeMets,et al.  Effect of carvedilol on survival in severe chronic heart failure. , 2001, The New England journal of medicine.

[12]  D. Delurgio,et al.  Cardiac resynchronization in chronic heart failure. , 2002, The New England journal of medicine.

[13]  Daniel Burkhoff,et al.  Mechanical Device–Based Methods of Managing and Treating Heart Failure , 2005, Circulation.

[14]  D. Burkhoff,et al.  Why does pulmonary venous pressure rise after onset of LV dysfunction: a theoretical analysis. , 1993, The American journal of physiology.

[15]  M C Oz,et al.  Left ventricular assist device-induced reverse ventricular remodeling. , 2000, Progress in cardiovascular diseases.

[16]  D. Burkhoff,et al.  Left ventricular volume reduction surgery for heart failure: a physiologic perspective. , 2001, The Journal of thoracic and cardiovascular surgery.

[17]  A. Feldman,et al.  Heart failure management using implantable devices for ventricular resynchronization: Comparison of Medical Therapy, Pacing, and Defibrillation in Chronic Heart Failure (COMPANION) trial. COMPANION Steering Committee and COMPANION Clinical Investigators. , 2000, Journal of cardiac failure.